LPM-II-plus
Particle Monitor
Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LPM-II-plus and is the answer not in the manual?

Questions and answers

Summary of Contents for STAUFF LPM-II-plus

  • Page 1 LPM-II-plus Particle Monitor Manual...
  • Page 2: Ec Declaration Of Conformity

    - BS EN 60721-3-4: 1995 Part 3: Classification of Groups of Environmental Parameters and their severities, Section 3.4 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 3: Table Of Contents

    6.2 General Operation .......................... 26 6.2.1 Physical Checks ........................26 6.2.2 Front Panel Operation ......................26 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 4 11.4 ISO4406:1999 Cleanliness Code System ..................73 11.5 NAS1638 Cleanliness Code System (now defunct, replaced by SAE AS4059E Table 1) .... 74 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 5 11.9 New ISO Medium Test Dust and its effect on ISO Contamination Control Standards ....77 11.10 Clean Working Practices ......................81 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 6: General Warnings And Information For The End User

    Personnel involved in slinging and hoisting operations are aware of the issues SLINGING AND regarding the safe transfer of machinery or parts of it and, therefore, uses As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 7: Operator Position And Dangerous Areas

    The machinery is not suitable for outdoor use and all the electrical devices have a protection class starting from IP 55 upwards. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 8: Dangers And Hazards That Cannot Be Eliminated

    PPEs to be used according to the activities on the machinery are listed in the following table: Activity PPEs Ordinary operation Shoes, gloves, goggles, overall Planned maintenance Shoes, gloves, goggles, overall As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 9: Transportation And Storage

    This location should not impede any other production or personnel. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 10: Warranty, Limitations And Disclaimers

    Damage has been caused by the attempts of individuals, other than Stauff staff to repair or service the product;...
  • Page 11 Stauff product return policy has not been followed. b) Product failure is caused by any of the exclusions described above, when the customer will be liable for the full cost of the repair (parts and labour) plus all carriage and insurance costs to and from Stauff premises.
  • Page 12: Product Presentation

    Nevertheless experience has shown the temperature measured is within a few degrees of that of the hydraulic system, in most applications. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 13: Data Logger

    Any subsequent use other than that intended may cause damage to the device As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 14: Technical Specification

    Keypad & LCD 6 keys, 128x64 pixels, back-lit graphical display Communication RS485 Options MODBUS CANBUS 4-20mA As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 15: Physical Attributes

    M/G Version – Viton (contact Stauff for any fluids that are incompatible with Viton Seal Material seals E Version - Perfluoroelastomer As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 16: Fluid Characteristics

    Stainless Steel Stainless Steel Sapphire (Al Sapphire (Al Sapphire (Al Viton Perfluoroelastomer Viton PTFE PTFE PTFE EPDM As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 17: Product Installation And General Operation

    If the flow is very low this can sweep the same particle backwards and forwards through the sensing volume multiple times, confusing the results. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 18: Electrical Interface

    The right hand port (remote connector) should be used if temporary connection is required. NOTE: Start signal and relay outputs only apply to this connector. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 19 All the other signals are optional. Item Minimum Maximum Voltage 9V DC 36V DC 200mA Current As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 20 100 Ohm termination resistors should be fitted as shown for long cables, for example over 10m. Twisted pair wiring should be used for any length over 2m. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 21 It can control and monitor a remote LPM, as well as allowing an external controller to be connected to it (for data download, for example). As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 22 It will continue to show and report the previous result.  This new operation mode applies whether or not continuous testing is enabled. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 23 0V. Similarly they can be converted to 0-10V outputs by connecting 500 ohm resistors. For details of how the test results are represented by the 4-20mA signals, see section 6.4.3. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 24: Hydraulic Connection

    1 High or Low Pressure Parallel Connection Figure 6.8 LPM working pressure generated by hydraulic component As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 25 E.g. at 30cSt and 200ml/minute, this is 0.4 Bar. The maximum and minimum allowed differential pressures can also be determined using the 400ml/min and 20ml/min lines, respectively. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 26 20:1 variation in either viscosity or differential pressure during operation. Figure 6.11 Differential Pressure vs Fluid Viscosity, for various flow rates As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 27: General Operation

    LPM-D versions also have a screen that changes colour. The alarm thresholds can be set from LASPAC-View via the serial interface. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 28: Date Of Issue: 06 November

    (current cleanliness level, with water content and temperature if applicable). The graphical format allows a full display of all codes of the standards supported. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 29 The progress of a test is denoted by the horizontal line; it grows from left to right as the test progresses. When it reaches the right hand side a new result is generated. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 30: Lpm Removal And Product Maintenance

    − Disconnected Operation As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 31: Computer Connection

    The first time that this is done, the correct communications port (COM port) on the computer has to be selected, as detailed below. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 32: Pc Software Operation

    The LPM incorporates a data logger, so previous test results can be downloaded into the test database using the Download New and Download All buttons. The difference between these is that Download New As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 33: Settings

    Actual operation is straightforward even when advanced settings are used during initial configuration. NOTE: Some items may be missing depending on the options fitted to the LPM. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 34: 2 As A Policy Of Continual Improvement, Stauff Reserves The Right To Alter The Specification Without Prior Notice

    This setting can be used when there is no flow available but communications need to be tested. This is the box ticked in figure 6.22 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 35 It may be necessary that the low flow indicator is turned off if filtration is below 10um (ISO 14/12/10 (NAS Class 4)), see figure 6.22 for location. 6.3.3.7 Continuous Testing Figure 6.23 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 36 Typically only the final "clean" result requires logging. 6.3.3.8 Changing Communication Protocols Figure 6.24 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 37 80% RH, or the temperature is greater than 65°C. The lower alarm is never triggered since all the settings are empty. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 38 AS4059E Table 2 uses letters instead of numbers to indicate the particle size range, so the settings are labelled appropriately. The standard specifies ways to represent a cleanliness level using only a subset As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 39 Alarm Mode 0: Warning-Alarm Output 1 Output 2 Turns on When >Lower >Upper Intended Function Warning Alarm As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 40 Output 2 Turns on When Test Complete >Lower “Not Clean” Signal Intended Function Test Complete Signal As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 41: Standard Communication Protocols

    The LPM can be controlled via commands on its serial (RS485) interface, using the Modbus RTU protocol. It is possible to control every aspect and setting of the LPM, as is done by the Stauff LASPAC- View control software. All results and counts are available in all supported formats. We suggest using LASPAC-View to initially configure the LPM and check it is performing correctly, and then the customer- written software only has to read the test results.
  • Page 42 5-15 ≥21µ 15-25 ≥25µ 25-50 ≥38µ 50-100 ≥50µ 100+ ≥70µ Table 6B RESULT CODES Registers 56-63 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 43 The special value -32768 (0x8000 hex) is again used to indicate ``No result’’, as per the contamination result codes. This could be due to a sensor failure or to the unit still in the process of powering up. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 44 Differential counts can be derived by subtraction. E.g., the NAS 5-15μm count could be calculated by subtracting the ISO 6μm(c) count from the 14μm(c) count. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 45: Can-Bus

    This procedure is described earlier in section 6. You will need a suitable RS485 interface, such as the LPM-USB Interface. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 46 The LPM was designed to be as flexible as possible. There are large number of options for setting operating modes, test result formats, alarm settings, downloading stored data etc. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 47 For a CanOpen network, use a base address of 0x182 for example. This will result in message IDs corresponding to the CanOpen ``pre-defined connection set’’. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 48: Analogue 4-20Ma Modes

    PLC. The PLC would need to be programmed to read each parameter at the correct time. These modes are still under development; please contact Stauff for details. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 49 ���� = 6 + ������ / 2 ������ = 2 × (���� − 6) So 6mA = ISO 0, 20mA = ISO28 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 50: Disposal

    − Fluids used with the LPM should be fully drained and disposed of according to EU waste framework directive and ISO44001 Environmental Management. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 51: Troubleshooting / Faq

    This warning can also come on when there are no particles whatsoever detected, i.e. the fluid is totally "clean’’. In this case the correct result e.g. 0/0/0 is still generated. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 52: Test Status

    Stauff Bottle Sampling Unit. See flushing guidelines on supplied product USB. DO NOT USE ACETONE As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 53: Test Duration

    "Normal" or "Dirty" Systems – Shorter test times or lower flows are acceptable. Figure 7.1 Test time needed for reliable indication by ISO code As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 54: How To Order

    -The full colour front panel LED indicator also reflects the test results, indicating if any set limits have been exceeded. See section 6.2.2. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 55: Related Products

    Note: Computer should have mains power applied at all times. Detailed installation and usage instructions are provided in the separate product user guide. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 56: Lpm-Dav

    A pressure compensated flow control valve suitable for the LPM. This may be needed if the application produces an oil flow that varies outside the upper flow range of the unit. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 57: Service And Recalibration

    As a policy of continual improvement, Stauff reserves the right to alter the specification without prior notice. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 58: Reference

    Test Number 32 bit integer 10-17 Test Reference Array of 16 packed characters Test Duration Unsigned integer As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 59 NAS1638, where we have to represent the NAS “00” class as -1, and “000” as -2. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 60 24-25. It can be useful during development to be able to read the clock and see a continuously incrementing value of seconds. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 61 This is read-only register 30. It contains a number (an enumeration) indicating the status of the LPM. See table 7A As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 62 IO_OP1 Alarm output 1 IO_OP2 Alarm output 2 UNUSED Not currently used Table 11D Status Flags As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 63: Implementing Modbus

    This 255 byte response frame looks like: <0xCC> <0x04> <0xfa> <250 bytes of data> <2 bytes of checksum> As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 64 For PLC users the details will be dependent on their own programming environment and facilities. But hopefully the above can be used as a guide for their own implementation. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 65: Further Can-Bus Information

    PC with USB ports running Windows  Special made-up CAN-bus cable detailed below  12 or 24 Volt DC power supply As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 66 Plug the LPM into the LPM-USBInterface  Start LASPAC-View  Select Tools/Remote Control to connect to the LPM. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 67 “base address”. This will define the start of the block of CAN message identifiers used by the software (using a value compatible with the J1939 standard). As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 68 Select a baud rate to match that being used on the LPM, for example 250k. Select the “Extended” message filter (so that 29 bit identifiers are used). Press OK to go to the main PCAN-View screen. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 69 (the display is in hexadecimal (base16) so the actual code is 23/21/19). Figure 11.5 PCAN-View Main Screen Figure 11.6 Reception of a Test Result Codes Message As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 70: Messages

    Transmit PDO 3 0x18FFB73F 0xFFB7 Commands 0x202 Receive PDO 1 0x18EF3F00 PDU1 Table 11G CAN-bus Messages As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 71 A number between 0 and 100 indicating the progress of the test. This will increase from 0 to 100 during the set test time. It can be used to drive a progress indicator. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 72 Set AS4059E Table 2 result format Format AS4059_E1 Set AS4059E Table 1 result format Format ISO11218 Set ISO11218 result format As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 73: Measuring Water In Hydraulic And Lubricating Fluids

    Examples: Hydraulic Oil @ 30°C = 200ppm=100% Saturation Hydraulic Oil @ 65°C = 500ppm=100% Saturation As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 74: Iso4406:1999 Cleanliness Code System

    These are at 5μm and 15μm equivalent to the 6μm(c) and 14μm(c) of the APCs (Figure 11.2) As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 75: Nas1638 Cleanliness Code System (Now Defunct, Replaced By Sae As4059E Table 1)

    1024000 182400 32400 5760 1024 Table 11.2 AS4059E Table 1 Cleanliness Classes for Differential Particle Counts As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 76: Recommendations

    As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
  • Page 77: Hydraulic System Target Cleanliness Levels

    12-15 systems 15-25 General equipment of limited reliability 25-40 Low pressure equipment not in continuous service As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 78: New Iso Medium Test Dust And Its Effect On Iso Contamination Control Standards

    The introduction of ISO MTD has necessitated changes to certain ISO standards. The standards affected include: As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 79 ACFTD calibration. Particle Size Obtained Using As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 80 21.2 22.0 22.7 23.5 24.2 24.9 25.7 26.4 27.1 27.9 28.5 29.2 29.9 30.5 31.1 31.7 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 81 20/18/13 20/18/15 21/19/13 4400 6300F 21/19/16 22/20/13 6300 22/20/17 23/21/14 15000 23/21/18 24/22/15 21000 25/23/17 100000 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
  • Page 82: Clean Working Practices

    - Don’t mix fluids into the same system. They can emulsify and eliminate any chance of a reliable particle count. - Don’t use unsuitable containers to take a fluid sample. [End of Manual] As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...