- BS EN 60721-3-4: 1995 Part 3: Classification of Groups of Environmental Parameters and their severities, Section 3.4 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
6.2 General Operation .......................... 26 6.2.1 Physical Checks ........................26 6.2.2 Front Panel Operation ......................26 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 4
11.4 ISO4406:1999 Cleanliness Code System ..................73 11.5 NAS1638 Cleanliness Code System (now defunct, replaced by SAE AS4059E Table 1) .... 74 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 5
11.9 New ISO Medium Test Dust and its effect on ISO Contamination Control Standards ....77 11.10 Clean Working Practices ......................81 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Personnel involved in slinging and hoisting operations are aware of the issues SLINGING AND regarding the safe transfer of machinery or parts of it and, therefore, uses As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
The machinery is not suitable for outdoor use and all the electrical devices have a protection class starting from IP 55 upwards. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
PPEs to be used according to the activities on the machinery are listed in the following table: Activity PPEs Ordinary operation Shoes, gloves, goggles, overall Planned maintenance Shoes, gloves, goggles, overall As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
This location should not impede any other production or personnel. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Damage has been caused by the attempts of individuals, other than Stauff staff to repair or service the product;...
Page 11
Stauff product return policy has not been followed. b) Product failure is caused by any of the exclusions described above, when the customer will be liable for the full cost of the repair (parts and labour) plus all carriage and insurance costs to and from Stauff premises.
Nevertheless experience has shown the temperature measured is within a few degrees of that of the hydraulic system, in most applications. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Any subsequent use other than that intended may cause damage to the device As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Keypad & LCD 6 keys, 128x64 pixels, back-lit graphical display Communication RS485 Options MODBUS CANBUS 4-20mA As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
M/G Version – Viton (contact Stauff for any fluids that are incompatible with Viton Seal Material seals E Version - Perfluoroelastomer As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Stainless Steel Stainless Steel Sapphire (Al Sapphire (Al Sapphire (Al Viton Perfluoroelastomer Viton PTFE PTFE PTFE EPDM As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
If the flow is very low this can sweep the same particle backwards and forwards through the sensing volume multiple times, confusing the results. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
The right hand port (remote connector) should be used if temporary connection is required. NOTE: Start signal and relay outputs only apply to this connector. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 19
All the other signals are optional. Item Minimum Maximum Voltage 9V DC 36V DC 200mA Current As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 20
100 Ohm termination resistors should be fitted as shown for long cables, for example over 10m. Twisted pair wiring should be used for any length over 2m. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 21
It can control and monitor a remote LPM, as well as allowing an external controller to be connected to it (for data download, for example). As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 22
It will continue to show and report the previous result. This new operation mode applies whether or not continuous testing is enabled. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 23
0V. Similarly they can be converted to 0-10V outputs by connecting 500 ohm resistors. For details of how the test results are represented by the 4-20mA signals, see section 6.4.3. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
1 High or Low Pressure Parallel Connection Figure 6.8 LPM working pressure generated by hydraulic component As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 25
E.g. at 30cSt and 200ml/minute, this is 0.4 Bar. The maximum and minimum allowed differential pressures can also be determined using the 400ml/min and 20ml/min lines, respectively. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 26
20:1 variation in either viscosity or differential pressure during operation. Figure 6.11 Differential Pressure vs Fluid Viscosity, for various flow rates As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
LPM-D versions also have a screen that changes colour. The alarm thresholds can be set from LASPAC-View via the serial interface. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
(current cleanliness level, with water content and temperature if applicable). The graphical format allows a full display of all codes of the standards supported. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 29
The progress of a test is denoted by the horizontal line; it grows from left to right as the test progresses. When it reaches the right hand side a new result is generated. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
− Disconnected Operation As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
The first time that this is done, the correct communications port (COM port) on the computer has to be selected, as detailed below. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
The LPM incorporates a data logger, so previous test results can be downloaded into the test database using the Download New and Download All buttons. The difference between these is that Download New As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Actual operation is straightforward even when advanced settings are used during initial configuration. NOTE: Some items may be missing depending on the options fitted to the LPM. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
This setting can be used when there is no flow available but communications need to be tested. This is the box ticked in figure 6.22 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 35
It may be necessary that the low flow indicator is turned off if filtration is below 10um (ISO 14/12/10 (NAS Class 4)), see figure 6.22 for location. 6.3.3.7 Continuous Testing Figure 6.23 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 36
Typically only the final "clean" result requires logging. 6.3.3.8 Changing Communication Protocols Figure 6.24 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 37
80% RH, or the temperature is greater than 65°C. The lower alarm is never triggered since all the settings are empty. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 38
AS4059E Table 2 uses letters instead of numbers to indicate the particle size range, so the settings are labelled appropriately. The standard specifies ways to represent a cleanliness level using only a subset As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 39
Alarm Mode 0: Warning-Alarm Output 1 Output 2 Turns on When >Lower >Upper Intended Function Warning Alarm As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 40
Output 2 Turns on When Test Complete >Lower “Not Clean” Signal Intended Function Test Complete Signal As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
The LPM can be controlled via commands on its serial (RS485) interface, using the Modbus RTU protocol. It is possible to control every aspect and setting of the LPM, as is done by the Stauff LASPAC- View control software. All results and counts are available in all supported formats. We suggest using LASPAC-View to initially configure the LPM and check it is performing correctly, and then the customer- written software only has to read the test results.
Page 42
5-15 ≥21µ 15-25 ≥25µ 25-50 ≥38µ 50-100 ≥50µ 100+ ≥70µ Table 6B RESULT CODES Registers 56-63 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 43
The special value -32768 (0x8000 hex) is again used to indicate ``No result’’, as per the contamination result codes. This could be due to a sensor failure or to the unit still in the process of powering up. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 44
Differential counts can be derived by subtraction. E.g., the NAS 5-15μm count could be calculated by subtracting the ISO 6μm(c) count from the 14μm(c) count. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
This procedure is described earlier in section 6. You will need a suitable RS485 interface, such as the LPM-USB Interface. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 46
The LPM was designed to be as flexible as possible. There are large number of options for setting operating modes, test result formats, alarm settings, downloading stored data etc. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 47
For a CanOpen network, use a base address of 0x182 for example. This will result in message IDs corresponding to the CanOpen ``pre-defined connection set’’. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
PLC. The PLC would need to be programmed to read each parameter at the correct time. These modes are still under development; please contact Stauff for details. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 49
= 6 + / 2 = 2 × ( − 6) So 6mA = ISO 0, 20mA = ISO28 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
− Fluids used with the LPM should be fully drained and disposed of according to EU waste framework directive and ISO44001 Environmental Management. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
This warning can also come on when there are no particles whatsoever detected, i.e. the fluid is totally "clean’’. In this case the correct result e.g. 0/0/0 is still generated. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Stauff Bottle Sampling Unit. See flushing guidelines on supplied product USB. DO NOT USE ACETONE As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
"Normal" or "Dirty" Systems – Shorter test times or lower flows are acceptable. Figure 7.1 Test time needed for reliable indication by ISO code As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
-The full colour front panel LED indicator also reflects the test results, indicating if any set limits have been exceeded. See section 6.2.2. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Note: Computer should have mains power applied at all times. Detailed installation and usage instructions are provided in the separate product user guide. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
A pressure compensated flow control valve suitable for the LPM. This may be needed if the application produces an oil flow that varies outside the upper flow range of the unit. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
As a policy of continual improvement, Stauff reserves the right to alter the specification without prior notice. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Test Number 32 bit integer 10-17 Test Reference Array of 16 packed characters Test Duration Unsigned integer As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 59
NAS1638, where we have to represent the NAS “00” class as -1, and “000” as -2. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 60
24-25. It can be useful during development to be able to read the clock and see a continuously incrementing value of seconds. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 61
This is read-only register 30. It contains a number (an enumeration) indicating the status of the LPM. See table 7A As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 62
IO_OP1 Alarm output 1 IO_OP2 Alarm output 2 UNUSED Not currently used Table 11D Status Flags As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
This 255 byte response frame looks like: <0xCC> <0x04> <0xfa> <250 bytes of data> <2 bytes of checksum> As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 64
For PLC users the details will be dependent on their own programming environment and facilities. But hopefully the above can be used as a guide for their own implementation. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
PC with USB ports running Windows Special made-up CAN-bus cable detailed below 12 or 24 Volt DC power supply As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 66
Plug the LPM into the LPM-USBInterface Start LASPAC-View Select Tools/Remote Control to connect to the LPM. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 67
“base address”. This will define the start of the block of CAN message identifiers used by the software (using a value compatible with the J1939 standard). As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 68
Select a baud rate to match that being used on the LPM, for example 250k. Select the “Extended” message filter (so that 29 bit identifiers are used). Press OK to go to the main PCAN-View screen. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 69
(the display is in hexadecimal (base16) so the actual code is 23/21/19). Figure 11.5 PCAN-View Main Screen Figure 11.6 Reception of a Test Result Codes Message As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Transmit PDO 3 0x18FFB73F 0xFFB7 Commands 0x202 Receive PDO 1 0x18EF3F00 PDU1 Table 11G CAN-bus Messages As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 71
A number between 0 and 100 indicating the progress of the test. This will increase from 0 to 100 during the set test time. It can be used to drive a progress indicator. As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031...
Page 72
Set AS4059E Table 2 result format Format AS4059_E1 Set AS4059E Table 1 result format Format ISO11218 Set ISO11218 result format As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Examples: Hydraulic Oil @ 30°C = 200ppm=100% Saturation Hydraulic Oil @ 65°C = 500ppm=100% Saturation As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
These are at 5μm and 15μm equivalent to the 6μm(c) and 14μm(c) of the APCs (Figure 11.2) As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
1024000 182400 32400 5760 1024 Table 11.2 AS4059E Table 1 Cleanliness Classes for Differential Particle Counts As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
12-15 systems 15-25 General equipment of limited reliability 25-40 Low pressure equipment not in continuous service As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
The introduction of ISO MTD has necessitated changes to certain ISO standards. The standards affected include: As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 79
ACFTD calibration. Particle Size Obtained Using As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 80
21.2 22.0 22.7 23.5 24.2 24.9 25.7 26.4 27.1 27.9 28.5 29.2 29.9 30.5 31.1 31.7 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
Page 81
20/18/13 20/18/15 21/19/13 4400 6300F 21/19/16 22/20/13 6300 22/20/17 23/21/14 15000 23/21/18 24/22/15 21000 25/23/17 100000 As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...
- Don’t mix fluids into the same system. They can emulsify and eliminate any chance of a reliable particle count. - Don’t use unsuitable containers to take a fluid sample. [End of Manual] As a policy of continual improvement, STAUFF reserves the right to alter the specification without prior notice. 201.031 Date of Issue: 06 November 2018...