REF. 7802.00 Electric Hi-Lo Bathtub COCOON, Key Hole tub shape, Watts Eurotherm thermostatic mixer REF. 7803.00 Electric Hi-Lo Bathtub COCOON, Key Hole tub shape, Rada Sense digital thermostatic mixer OPTIONS : REF. 7800.10 Disinfection shower handset with trigger, flow adjustment, suction device of the disinfectant product with stop valve, 2 litres tank, secured door with lock with key.
• Max. weight (water + patient) : 320kg • Weight of the Cocoon R’GO bathtub : 130kg • Weight of the Cocoon Key Hole bathtub : 140kg • Variable height from 75 to 110cm COCOON R’GO TECHNICAL DATAS • Ergonomical tub with « no-slip » concept and extra-height of the legs.
OPERATION Hi/Lo of the bathtub "Cocoon" is obtained thanks an electric jack and a gas jack, the first one is driven by an electric control. Filling of bathtub and mixed water feeding of cleaning shower are obtained thanks a thermostatic mixer.
Keep this manual with the tub and refer to it as required. Contents of this manual are subjected to change without prior notice to users. • The Cocoon bathing system must be installed by a contractor or a certified installation technician. • DO NOT try to lift the tub.
DIRECTIONS FOR USE HIGH / LOW FUNCTIONS In order to change the height of the tub, use the buttons located on the control panel : PLUG OPERATION Turn the button in a clockwise direction to close the drain plug and counter clockwise to open it 8/20...
DIRECTIONS FOR USE CONCEANED DIGITAL MIXING VALVE RADA SENSE (MODELS 7801.00 & 7803.00) To adjust the water temperature,, pass your finger in front of the cold water sensitive captor (blue triangle) or in front of the hot water (red triangle) according to the temperature desired. The max.
DIRECTIONS FOR USE BATH FILL (MODELS 7800.00 & 7803.00) Bath fill is operated by the valve located under the mixer. Filling and stop filling is made by turning the valve button. CLEANING SHOWER HANDSET Select a temperature by adjusting the thermostatic. It is possible to regulate the pressure of the shower by closing or slightly opening the valve.
DIRECTIONS FOR USE DISINFECTION SHOWER HANDSET The disinfection shower is identified thanks to a red label "disinfection shower" and a yellow flexible. The disinfection shower allows the disinfection of the bathtub after each use and is connected to a disinfectant tank (see: disinfection system).
The auto-clean function significantly reduces the risk and development of micro -biological organisms that can develop in humid and wet environments. Reval's unique air ducting system can be dismantled without the need for tools and can be termically cleaned during routine servicing in a matter of minutes.
DIRECTIONS FOR USE CHROMOTHERAPY (OPTION) The chromotherapy option consists of 4 LED lamps and 1 control button. The control button has 3 following functions : On/Off / Automatically alternate of 7 colours / Colour fixed. Push one time on the button to turn on the lamps. The colours of the lamps automatically alternate during 30 seconds for each colour.
AUTOMATIC STOP FILLING (1 LEVEL) FOR 7801.00 & 7803.00 OPERATION On the 7801.00 & 7803.00 Cocoon bathtubs provided with the Rada Sense thermostatic mixer, the automatic stop filling is made by the adjustment of the filling time. For the model 7801.00 (R’GO shape tub), the filling time is pre-set at 5 minutes for a 27cm height of water level.
The water level n°1 corresponds to a height of water of 20cm. The water level n°2 corresponds to a height of water of 27cm. ADJUSTMENT OF THE WATER LEVELS It is possible to adjust the heights of water levels by moving the probes. Contact Reval for more information. 15/20...
DISINFECTION SYSTEM (OPTION) COMPOSITION The bath is equipped with an automatic system to determine the quantity of disinfectant. The disinfection system includes : - 1x 2 litres tank receiving the disinfectant product. - 1X plunger with calibrated tube, located in the tank, and connected by a flexible going through the top of the venturi (white plastic part).
CLEANING THE FILTERING CARTRIDGE OF THE MIXER (WATTS MIXER) Dismount stage by stage elements of the thermostatic mixer starting with the brace (reference mark n°12) until the filter cartridge (reference mark n°2). In the event of too important limestone deposit, contact France Reval. N° REF. DESIGNATIONS N°...
CLEANING / MAINTENANCE CLEANING OF THE FILTER OF THE FILLING SPOUT The filling spout is equipped with an aerator with filter. The aerator can be dismantled and cleaned. A special key delivered with the bathtub allows unscrewing easily the aerator. It is advised to clean and to decalcify the aerator at least twice a year.
6) Defects and damages of importance to the safety of the bathtub that have occurred between inspections and have already lead to corrective actions should be entered in the logbook. 7) Defects and damages should be reported back to Reval for action. This feedback should be reported in the logbook.