Sensor Replacement - AIR SYSTEMS BB15-COPM Operating Instructions And Replacement Parts List Manual

Table of Contents

Advertisement

Replacement sensors are shipped with a metal spring installed between the electrodes. Do not remove the clip until the
sensor is to be installed into the monitor.
Step 1)
Disconnect all external connections.
Remove CO monitor from the unit.
Step 4)
Remove sensor from sensor cup and remove leads. Take the new
sensor and remove the metal spring. Reattach leads to the proper
colored terminals on the new sensor. Install new sensor into sensor
cup.

Sensor Replacement

Step 2)
Remove the four screws from the moni-
tor's left endplate.
UNUSED
CONTROL
(BLACK)
SENSING
(RED)
11
Step 3)
Remove endplate to gain access to the
sensor cup.
REFER-
ENCE
Step 5)
Reassemble monitor and reinstall in unit.
Connect all cables and air sample hose.
Allow monitor to stabilize 30 minutes to 1
hour and recalibrate.