GETTING STARTED ....................1
1-
Introduction ......................1
GETTING STARTED ....................3
2-
Quick Start Guide ....................3
Using the THM1176/TFM1186 with a desktop computer ................. 3 2-1-1 PC: Software Installation ....................... 3 ...
Page 4
TRIGger ..............................42 UNIT ............................... 43 5-10 Write and Read ............................43 PROGRAMMING THE THM1176/TFM1186 ............44
6-
USB Interface ....................44
General ..............................44 SCPI Instrument Model .......................... 45 ...
The recommended calibration interval is 18 months. You can benefit from a discounted price for the calibration if you return your instrument to Metrolab at a time that corresponds to our batch calibrations; please see Section 7-10 for details.
Page 7
We hope the THM1176/TFM1186 will help you perform your magnetic field measurements easily and accurately. If you have problems and your reseller cannot help you further, the Metrolab team is ready to help. Even if you don’t have problems, we are always interested in knowing more about how our instruments are used.
Overview. See Chapter 4-Software User Interface for details on using the software. • (Does not apply to TFM1186 – see Section 2-4!) Check the offset before each series of measurements by placing the probe in the Zero Gauss Chamber. If the offset is higher than desired, leave the probe in the Zero Gauss Chamber and perform the User Offset Correction procedure using the Offset / ZERO button on the Setup page.
• Be sure to stop the measurement software using the Stop button before turning off the PDA or unplugging the THM1176/TFM1186 – see the note below. • Manipulate the probe and check the offset as described in Section 2-1-3.
Page 13
The disadvantages are that the sensor becomes thicker, and that it cannot be recalibrated. Because of this trade-off, Metrolab have decided to leave it to the customer to decide whether or not they want to use the ceramic plates.
Page 15
• Some PDA USB-Host adaptor cables may not have a power connector. Plug the AC adaptor into the DC-in connector on the PDA instead. • If you plug the THM1176/TFM1186 into the PDA before the PDA completes its initial boot procedure, you will see an “Unidentified USB Device” dialog.
The active field volume of the THM1176-LF and TFM1186 is much larger, on the order of several mm, but this is usually sufficient for weak fields.
Page 17
• User offset correction (not for TFM1186 – see Section 2-4!): To guarantee the specified accuracy, the measurement offset should be checked before each measurement sequence, using the zero-gauss chamber supplied.
(see below). However, the THM1176-HFC lacks all such protection; please note the special handling precautions in Section 2-4, Precautions. The TFM1186 sensor is relatively large and mechanically very robust. • Small gaps: The THM1176-MF, -HF and -LF probes are 10 mm thick.
The desktop kit includes everything except the computer: the THM1176/TFM1186 instrument, calibration, a zero-Gauss chamber (not for TFM1186 – see Section 2-4), the National Instruments VISA runtime library and USBTMC driver, turn-key measurement software, all the LabVIEW source code, software licenses from Metrolab as well as National Instruments, and this manual.
PC software. • Includes the entire desktop kit: You can choose to connect the THM1176/TFM1186 either to the PDA included in the handheld kit, or to a PC. In the second scenario, you can do everything you can do with the desktop kit.
THM1176 family magnetometers will be calibrated. To avoid substantial extra charges, you should ship the unit back to Metrolab in order to coincide with one of these dates. Please see Section 7-10 for a list of these dates.
Page 23
To comply with the license associated with the LabVIEW Mobile Module, Metrolab distributes upgrades to the PDA software in an encoded file. You need to contact Metrolab for the password for this file. Please send an e- mail with your instrument’s serial number to contacts@metrolab.com. The serial number is printed on the label located on the instrument electronics.
USING THE THM1176/TFM1186 4- Software User Interface The THM1176/TFM1186 can be used as a laboratory instrument, plugged into a Personal Computer (PC), or, optionally, as a handheld instrument. The display of the handheld instrument will vary slightly from the illustrations below; however, the utilization is the same except when noted.
Both the LabVIEW API and User Interface (UIF) are included in source code format on the CD that came with your THM1176/TFM1186. If upgrades become available, you will be able to download them from the Metrolab website, download.metrolab.com.
THM1176/TFM1186. Reset [In]: Reset the instrument after opening a session to it. THM1176 [Out]: Instrument identified itself as THM1176/TFM1186. Only valid if ID Query is selected. THM1176 Description [Out]: Instrument description returned by *IDN command. Only valid when "ID Query" is selected.
Page 41
Reset; o Continuous trigger initiation; o Changing trigger parameters. Sleep [In]: If set, the THM1176/TFM1186 will switch off the analog power after executing the command. (The microcontroller also puts itself to sleep whenever it can, but this happens automatically.) The resulting power...
You may, however, want to write a LabVIEW program optimized for your application, bypassing or complementing the general-purpose API provided by Metrolab. This chapter will provide you with the information necessary to do so.
Systems Alliance, available from http://www.ivifoundation.org/Downloads/Specifications.htm. 6-2 SCPI INSTRUMENT MODEL The THM1176/TFM1186 complies with the Standard Commands for Programmable instruments (SCPI) standard. SCPI uses a standard instrument model to organize the command structure. The diagram below shows the SCPI Instrument Model SCPI Instrument Model subsystems concerning the signal flow.
Page 51
488.2 “Common Commands” (start with “*”) as well as SCPI commands. Many commands have additional options, or query forms to return the currently set value; see later sections for details. In addition to ASCII commands, the THM1176/TFM1186 also supports certain USBTMC-USB488 controls, also noted in this table. Functional...
Standard Event Status Register. SCPI adds the Operation Status Register, Questionable Status Register and Error/Event Queue. The diagram below, taken from the SCPI standard, provides a good summary. This section describes how the THM1176/TFM1186 uses these status registers. www.metrolab.com...
Page 55
Read-only register that is constantly updated to reflect the current state of the instrument. o Event Register Transitions in a Condition Register are latched in the corresponding Event Register. The THM1176/TFM1186 only latches transitions from 0 to 1. Event Registers are cleared when read. www.metrolab.com...
Note that National Instruments' "Measurement & Automation Explorer" (part of the full NI-VISA package on Windows systems) provides a very useful tool to explore the command set. Select the THM1176/TFM1186 under "System / Peripherals & Interfaces / USB Devices," and click the "Open VISA Test Panel" icon. This opens a window from which you can try all functions available through NI-VISA.
Page 71
THM1176 is designed to have very low offset and offset drift, but as with any Hall device, these remain a significant source of error. Note that this does not apply to the TFM1186 – See Section 2-4. After an INITiate command with TRIGger:SOURce = BUS, the instrument expects TRIGger:COUNt triggers before resuming normal operation.
- AC adaptor/charger (100-240 VAC 50/60 Hz, 500 mA) with wall socket adaptor plugs for Europe, UK, USA, Australia - USB-Host adaptor cable to connect PDA to THM1176/TFM1186 - USB-Device adaptor cable to connect PDA to PC - Pre-installed acquisition software, with licenses from Metrolab,...
Four weeks in the year are fixed to perform batch calibrations: Week 10, Week 24, Week 37, Week 50 To benefit from the discounted pricing for this batch operation, you must ensure that your instrument arrives at Metrolab the week before. Certification CE approved...
“Command Reference”, provide a generic description of all possible error codes. In general, the codes are between -800 and 300. This chapter describes only the error codes produced by the THM1176/TFM1186, and the circumstances that might produce each error. In addition to the SCPI errors, the THM1176 Application Programming Interface might produce additional errors, in the range of -1073807360 to -1073807192 and 1073676290 to 1073676457 (0xBFFF00xx and 3FFF00xx –...
-400: Query error Generic query error: • Device being cleared when query was received. On the THM1176/TFM1186, no other conditions should generate this error. If this is not the case, please report to Metrolab the circumstances under which you received this error.
• encountered something other than a Boolean or DEFault for a Boolean parameter; or • encountered an unknown parameter type. Please report to Metrolab the circumstances under which you received this error. 9-6-2 204: Data buffer was overrun In timed-trigger mode, data was lost because the read-out did not keep up with the acquisition.
Need help?
Do you have a question about the TFM1186 and is the answer not in the manual?
Questions and answers