Adjustment Analogue MAX

A
12 mA (5 V) Analog Min is successful adjusted.
OUT
Fill the container with the material to be detected up to the desired level Analogue Max.
Connect the grey wire (ETW) with supply voltage (+) and release the adjustment connection (ETW)
within 8...13 seconds. The LED is flashing during the adjustment (15 Hz) A
Adjustment analogue Max. successful done, A
ETW
H
L
T
0
(LED flashing 15 Hz)
A
OUT
Adjustment
20
16
12
8
4
0
Initialisation situation: MIN adjusted
With an adjustment failure (ERROR) the probe returns to the previous adjusted state.
With unsuccessful adjustment Analogue Max., the adjustment Analogue Max. can be repeated.
MAX
8s
13s
MAX - OK