Calibration - Gram Precision RK-15 Series Operation Manual

Table of Contents

Advertisement

Available languages

Available languages

SET15
RESET
press key
when displaying OK.

CALIBRATION

Note: Before calibration, please set the capacity first. The unit used in
calibration is the one that has been set before. During the calibration
procedure, long press
saving.
Here we take 3kg/0.2g as an example
1. Press and hold
displays "CAL"
2. With no load on the weighing pan, press
3. Wait till the window displays the first calibration value.
Note: The first calibration value is default. If the value has been changed, the
default value is 1/3 of full load.
Note: If you need to change the value, do as the following: Press key
to enter the value setting. Press key
rightwards. Press key
Press key
to save.
4. Put the corresponding weight on the weighing pan, and then press
complete the first point calibration.
parameter initialization
then key
to return to normal weighing mode without
while powering on, not release it till the window
or
to modify the value (or via numeric keys).
to start initialization. It is finished
to start zero point calibration.
or
to move leftwards or
to
17

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rk-6 seriesRk-3 seriesRk-30 series

Table of Contents