The technical specifications specifies the maximum required power at nominal operating values, e.g. 24 V supply voltage,
70°C ambient temperature, 600 Ω load, and 20 mA output current.
DIN rail solution - device daisy chain:
The units can be supplied with 24 VDC ±30%
via direct wiring and a loop between the
devices.
Protective fuse: 2.5 A.
Protective fuse: 0.4 A.
Power rail solution #1:
Alternately, you can connect 24 VDC to any
one JUMO dTRANS T/S08 device with power
rail connector which will then energize other
units on the rail.
External fuse characteristics:
The 2.5 A fuse must break after not more than 120 seconds at 6.4 A.
6
Flexible supply
Power rail solution #2:
The JUMO dTRANS T/S08 PCU power
connector unit allows easy connection of a
24 VDC / 2.5 A source to the power rail.
Protective fuse: 2.5 A.
V1.00/EN/00698691 - 70720300T90Z001K000