Add Reinjections To A Plate Record - Applied Biosystems 3730xl User Manual

Dna analyzer
Table of Contents

Advertisement

5
Chapter 5 Run the instrument

Add reinjections to a plate record

Status
Autosampler
Processing
Current runs
Collecting
Validated

Add reinjections to a plate record

66
Data is being collected from the plate.
Injection is in process.
Injection has been validated and will be run.
Table 4 Current Run fields
Field
Run ID
Application
Run protocol
2. Monitor the System status indicator in bottom left of screen.
—Instrument is running
flashing—Check the Event Log.
See "View the Event Log" on page 68 or "Display the Event Log" on
page 135.
—Instrument has stopped
To add a reinjection, edit the plate record, then run the plate again.
1. In the navigation pane of the Data Collection Software, double-click
Instruments4
ga37304
2. Select a plate record, then click Edit.
3. Select Edit4Add Sample Run.
Columns for Results Group 2 and Instrument Protocol 2 are added to the plate
record.
4. Specify the settings for the new injection.
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
Description
Description
The internal number assigned by the software
Instrument protocol type selected in the instrument
protocol
Instrument protocol name selected in the plate record
Plate Manager.