Chapter 5 Run the instrument
Add reinjections to a plate record
Status
Autosampler
Processing
Current runs
Collecting
Validated
Add reinjections to a plate record
66
Data is being collected from the plate.
Injection is in process.
Injection has been validated and will be run.
Table 4 Current Run fields
Field
Run ID
Application
Run protocol
2. Monitor the System status indicator in bottom left of screen.
•
—Instrument is running
•
flashing—Check the Event Log.
See "View the Event Log" on page 68 or "Display the Event Log" on
page 135.
•
—Instrument has stopped
To add a reinjection, edit the plate record, then run the plate again.
1. In the navigation pane of the Data Collection Software, double-click
Instruments4
ga37304
2. Select a plate record, then click Edit.
3. Select Edit4Add Sample Run.
Columns for Results Group 2 and Instrument Protocol 2 are added to the plate
record.
4. Specify the settings for the new injection.
3730xl DNA Analyzer with 3730xl Data Collection Software 5 User Guide
Description
Description
The internal number assigned by the software
Instrument protocol type selected in the instrument
protocol
Instrument protocol name selected in the plate record
Plate Manager.