Configuration Entries - Balzers QMG421C Operating Instructions Manual

Quadrupole mass spectrometers
Table of Contents

Advertisement

EIALZggs
5.
Operation
The following explains
how to
operate
a
control unit with
a
front
panel console, but the functions described also
apply to control units with dummy
panels.
5.1
Initial start up
5
1.1
Mains power
connection
5.1.2 Configuration
entries
CAUTION:
Before
switching
on
the apparatus:
Proceed
as
follows to configure the control unit
for your
-
make
sure
that
is set
to the voltage
of
QMG system;
the
power
supply
-
Select the <general>
function
group and the
-
the unit must
be
properly grounded
<config>
function.
-
all modules, cables
and
periphery
components (compare section
3)
must
-
Select the «SYSTEM» parameter
by
pressing
the
be
properly installed
appropnate
softkey.
-
alt
the technical
data
(compare section
-
Enter the family to which your quadrupole analyzer
7)
must
be
met
belongs.
If
necessary, select the
«QMA»
submenu
by
pressing
the
appropriate
softkey
and
change the
DANGER:
Any
interruption
of
the
protective
parameter value using the
or
key,
conductor inside
or
outside
the
apparatus
or
disconnection
of
the
-
Enter the
mass range for
which your quadrupole
protective earth terminal
is
likely to
electronics
or
RF
generator
is
configured.
If
necessary,
make the apparatus dangerous,
select the «MASS-R» submenu
by
pressing the
Intentional interruption
is
prohibited,
appropriate
softkey,
and
change
the
parameter value
using
the
or
key.
The main
power
switch
is
-
Enter the
type
of ion
detector mounted
on your
located în the lower left-hand
powe
analyzer.
If
necessary,
sefect
the «DETECT» submenu
section
of
the
front
panel,
by
pressing the appropriate
softkey,
and
change the
When the unit.is powered
on,
it
parameter value using the
or
key
runs
through
a
self-diagnostics
e
1
/
-
Enter
the
type
of
ion
source
mounted
on your
analyzer.
routine and activates
the
default
/
If
necessary, select the «lS-TYP» submenu
by
pressing
parameters (default parameter
the appropriate
softkey,
and change the parameter
the parameters
that
were
active
o
,
value using the
L
or
L
i
key,.
when
the unit
was
turned
off
the
last
timet
-
If
you
are
using
an
option, select the «OPTION»
submenu
and
define the option using the
o
or
key.
-
Finalize these configuration entries
by
pressing the
«RETURN»
softkey.
OMS42tgg
opm
Operation
17