Controller - Balzers QMG421C Operating Instructions Manual

Quadrupole mass spectrometers
Table of Contents

Advertisement

BALZERS
OMS
Controller
Problem
Possible
cause
Correction
ERR
17: stack overflow
OC 421 quadrupole controller
defective
Contact
Balzers
Service
for
replacement
ERR
18: Reset
error
ERR
19: watchdog
errot
ERR
33: idle error
ERR
40: CAN error
ERR
20: EPROM defect
<config>
«TEST» «QMS»
:
«EPROM-T»
Contact
Balzers
Service
for
replacement
ERR
23:
NOVRAM
defect
<config>
«TEST» «QMS»
:
«RAM-T»
ERR
22: dual port RAM defect
<config>
«TEST» «OMS»
:
«RAM T»
ERR
23:
buffer
RAM defect
<config>
«TEST» «OMS»
:
«RAM-T»
\4cold
Console
Problem
Possible
cause
Correction
ERR
3:
stack
overflow
CS 421
console defective
Contact
Balzers
Service
for
replacement
ERR
4: idle
error
ERR 7:
watchdog
error
ERR
14: Display RAM defect
Contact
Balzers
Service
for
replacement
ERR
15: EPROM defect
<config>
«TEST»
«CS»
:
«EPROM-T»
ERR
16: RAM defect
<config>
«TEST» «CS»
:
«RAM-T»
EL
La]
Signal
processor
Problem
Possible
cause
Correction
ERR
24: EPROM defect
<config>
«TEST» «DSP»
:
«EPROM-T»
Contact
Balzers
Service
for
replacement
ERR
25: dual port RAM defect
<config>
«TEST» «DSP»
:
«RAM-T»
fon source
supply
Problem
Possible
cause
CorrectÍon
ERR
36:
lon source
error
The
IS
420 ion source supply
is
over-
Correct according to
the
OPM
for
the
loaded
or
defective
iS
420
[7]
102
TrLb
slooting
OMs421oa.opm

Advertisement

Table of Contents
loading

This manual is also suitable for:

Qmg421

Table of Contents