Detectors; The Detector Firmware Is Not Updated After A New Firmware Update File Is Transferred Via Usb Flash Drive; The Detector Time Is Incorrect, Or Is Not The Same As The Module Time; Daylight Savings Time Did Not Work - BW Technologies IntelliDoX User Manual

Docking module
Table of Contents

Advertisement

INTELLIDOX DOCKING MODULE

Detectors

The detector firmware is not updated after a new firmware update file is
transferred via USB flash drive.
The file name is incorrect.
The file is not in the System folder on the USB flash drive.
Update Detector Firmware is not enabled via Fleet Manager II software.
The firmware on the detector is the same version as the firmware update.

The detector time is incorrect, or is not the same as the module time.

Use software to enable the Synchronize detector time and date option.
The module time was not set via Fleet Manager II.
The time zone is incorrect.

Daylight Savings Time did not work

The module Daylight Savings Time was not set.
The time zone is incorrect.
The module time was not set via Fleet Manager II.
BW TECHNOLOGIES BY HONEYWELL
The firmware update file name is not the same as the downloaded file name.
Rename the file name to match the name of the downloaded firmware update
file, or download a new firmware update file. See Update Firmware.
Create a folder called System on the USB flash drive. Move the file to
the System folder. See Updating Module Firmware via USB Flash Drive.
Use software to enable detector firmware updates. Transfer the modified
configuration file to the appropriate modules.
Verify that the firmware file on the USB flash drive is the most recent firmware
version. See Update Firmware.
See Adjust Time and Date Settings via Fleet Manager II Software and in
the Automated Operations section, select Synchronize detector time and date.
Use software to set the module time. See Adjust Time and Date Settings via
Fleet Manager II Software.
Use software to select the time zone. See Adjust Time and Date Settings via
Fleet Manager II Software.
Use software to automatically adjust clock for Daylight Savings Time. See Adjust
Time and Date Settings via Fleet Manager II Software.
Use software to select the time zone. See Adjust Time and Date Settings via
Fleet Manager II Software.
Use software to set the module time. See Adjust Time and Date Settings via
Fleet Manager II Software.
USER MANUAL || TROUBLESHOOTING