Page 1
4050B Series Function/Arbitrary Waveform Generator USER MANUAL 1.800.561.8187 information@itm.com www. .com...
Page 2
Safety Summary The following safety precautions apply to both operating and maintenance personnel and must be followed during all phases of operation, service, and repair of this instrument. Before applying power to this instrument: Read and understand the safety and operational information in this manual. ...
Page 3
Category IV (CAT IV): Measurement instruments whose measurement inputs are meant to be connected to the primary power entering a building or other outdoor wiring. Do not use this instrument in an electrical environment with a higher category rating than what is specified in this manual for this instrument.
Page 4
Unless otherwise stated, a ground connection on the instrument's front or rear panel is for a reference of potential only and is not to be used as a safety ground. Do not operate in an explosive or flammable atmosphere Do not operate the instrument in the presence of flammable gases or vapors, fumes, or finely- divided particulates.
Page 5
Measurements made by this instrument may be outside specifications if the instrument is used in non-office-type environments. Such environments may include rapid temperature or humidity changes, sunlight, vibration and/or mechanical shocks, acoustic noise, electrical noise, strong electric fields, or strong magnetic fields. Do not operate instrument if damaged If the instrument is damaged, appears to be damaged, or if any liquid, chemical, or other material gets on or inside the instrument, remove the instrument's power cord, remove the...
Page 6
Do not touch live circuits Instrument covers must not be removed by operating personnel. Component replacement and internal adjustments must be made by qualified service-trained maintenance personnel who are aware of the hazards involved when the instrument's covers and shields are removed. Under certain conditions, even with the power cord removed, dangerous voltages may exist when the covers are removed.
Page 7
Cooling fans This instrument contains one or more cooling fans. For continued safe operation of the instrument, the air inlet and exhaust openings for these fans must not be blocked nor must accumulated dust or other debris be allowed to reduce air flow. Maintain at least 25 mm clearance around the sides of the instrument that contain air inlet and exhaust ports.
Compliance Statements Disposal of Old Electrical & Electronic Equipment (Applicable in the European Union and other European countries with separate collection systems) This product is subject to Directive 2002/96/EC of the European Parliament and the Council of the European Union on waste electrical and electronic equipment (WEEE), and in jurisdictions adopting that Directive, is marked as being put on the market after August 13, 2005, and should not be disposed of as unsorted...
Page 9
CE Declaration of Conformity This instrument meets the requirements of 2006/95/EC Low Voltage Directive and 2004/108/EC Electromagnetic Compatibility Directive with the following standards. Low Voltage Directive o EN61010-1: 2001 EMC Directive o EN 61000-3-2: 2006 o EN 61000-3-3: 1995+A1: 2001+A2: 2005 o EN 61000-4-2 / -3 / -4 / -5 / -6 / -11 o EN 61326-1: 2006 viii...
Safety Symbols Refer to the user manual for warning information to avoid hazard or personal injury and prevent damage to instrument. Electric Shock hazard Alternating current (AC) Chassis (earth ground) symbol. Ground terminal On (Power). This is the In position of the power switch when instrument is ON.
Table of Contents SAFETY SUMMARY ..............................I ............................OMPLIANCE TATEMENTS ................................AFETY YMBOLS ................................. OTATIONS GENERAL INFORMATION .......................... 1 ............................1 RODUCT VERVIEW ............................. 1 ACKAGE ONTENTS ............................3 RONT ANEL VERVIEW Front Panel Description ........................... 3 ............................4 ANEL VERVIEW Rear Panel Description ............................
Page 12
Selecting Modulation Source ............................ 43 To Set Modulation Depth ............................44 To Set Modulation Frequency ........................... 44 DSB-AM Modulation ............................44 FM Modulation .............................. 46 To Set Frequency Deviation ............................47 PM (Phase Modulation) ..........................47 To Set Phase Deviation .............................. 48 FSK Modulation .............................
Page 14
Example 8: Generate a Burst Waveform ..................... 110 Example 9: Generate an AM Modulation Waveform .................. 112 Example 10: Generate an FM Modulation Waveform ................113 Example 11: Generate a PM Modulated Waveform ................... 114 Example 12: Generate an FSK Modulated Waveform ................. 115 Example 13: Generate an ASK Modulation Waveform ................
General Information 1.1 Product Overview BK Precision’s 4050B series are dual-channel function/arbitrary waveform generators, capable of generating sine and square waves of up to 10 MHz, 30 MHz or 60 MHz (depending on model). They have an informative easy-to-read color display, user-friendly controls and a numeric keypad which allows users to easily configure waveform properties.
Page 16
1 x 4050B series function/arbitrary waveform generator 1 x Getting started manual (printed) 1 x AC power cord 1 x USB type A to Type B cable 1 x Certificate of calibration Verify that all items above are included in the shipping container. If anything is missing, please contact B&K Precision.
1.3 Front Panel Overview Front Panel Description Power On/Off switch Menu softkeys Menu buttons Channel 1 output terminal Channel 2 output terminal Channel 1 and 2 Output On/Off buttons Arrow keys Rotary dial knob Numeric keypad TFT LCD color display USB host port/*USB-to-GPIB adapter interface Accepts USB flash drive to save/recall instrument settings and waveforms.
Getting Started Before connecting and powering up the instrument, please review and go through the instructions in this chapter. 2.1 Input Power Requirements Input Power The supply has a universal AC input that accepts line voltage and frequency input within: 100 –...
1. Use good quality 50 Ω coaxial cable and connectors. 2. Make all connections tight and as short as possible. 3. Use good quality attenuators, if it is necessary to reduce waveform amplitudes applied to sensitive circuits. 4. Use termination or impedance-matching devices to avoid reflections. 5.
Figure 1 - Preliminary Self-Test Figure 2 - Self-Test Options Note: Refer to TEST/CAL section for more information about the Self-Test function. Check Model and Firmware Version The model and firmware version can be verified from within the menu system. Press Utility, press System select and press Page 1/2 from the menu to enter the second menu page, and select System Info option.
Output Check Follow the steps below to do a quick check of the settings and waveform output. 1. Turn on the instrument and set the instrument to default settings. To set to default, press Utility → System → Set To Default to set the system to the default setting. The instrument will set both channels with the following parameters: Output Default...
Page 24
Start Phase 0° Cycles 1 Cycle Trig Out Delay 521 ns Trigger Default Source Internal Table 1 - Default Settings 1. Connect the BNC output of CH1 (yellow) into an oscilloscope. 2. Press the Output button on top of CH1 output BNC to turn on the output and observe a sine wave with the parameters above.
3 Front Panel Operation 3.1 Menu Options All settings and parameters can be configured from the menu system of the instrument. The channel specific menu options are the same for both channel 1 and channel 2. Use the Ch1/Ch2 button keys to toggle the channel selection. The selected option will be highlighted in blue.
Page 26
Offset/LowLevel Configures the DC offset or the low level of the waveform. Phase Configures the phase relative to the other channel. DutyCycle Configures the duty cycle of the waveform. Ramp Frequency/Period Configures the frequency or period of the waveform. Amplitude/HighLevel Configures the amplitude or the high level of the waveform.
Page 27
Source Selects modulating source. AM Depth Set the modulation depth. Shape Configures the modulating waveform shape. AM Freq Set the modulating waveform frequency. Frequency range: 1 mHz to 20 kHz (internal source only). Sweep Sweep Time Configures the sweep time for sweep output. StartFreq/CenterFReq Configures the sweep start frequency or the center frequency.
Copy channel settings between channels. 3.2 Selecting a Channel The 4050B series function/arbitrary waveform generators have dual channel outputs. They can be operated independently or in sync with each other. To select between channel 1 and 2 and view/change their parameters, press the Ch1/Ch2 key.
Figure 4 - Channel 2 Selected 3.3 Configure Waveform Output Configure Waveform Shape The instrument can generate many standard as well as arbitrary waveforms. There is a dedicated waveform key on the front panel that will allow the user to select between different waveform shapes to output, as listed in Table 4.
Page 30
Press Waveforms to select a waveform. Menu options relevant to the selected waveform shape will display at the bottom of the screen. The screenshots below illustrates the menu options for each of the waveform types. Sine Waveform Square Waveform Ramp Waveform Pulse Waveform Noise Waveform DC Waveform...
Table 5 - Waveforms Configure Frequency This section does not apply to noise or DC waveforms. Press the Frequency/Period softkey to toggle between the Frequency and Period settings. The option selected will be highlighted in blue. The current value for the waveform’s frequency or period is now highlighted on the Waveform Parameters display section of the screen.
4. Use the menu function keys to select the unit. Available units are: MHz, kHz, Hz, mHz, and uHz. Note: When using the numeric keyboard to enter the value, the left arrow key can be used to move the cursor backward and delete the previous digit. Configure Amplitude This section does not apply for noise or DC waveform.
User-Defined High and Low Level The user has the option to adjust the high and low level of the waveform. To do this, toggle the menu function key from Amplitude to HighLevel and the LowLevel option will be available as well.
Figure 8 - Offset 6. Use the menu function keys to select the unit. Available units are: Vdc and mVdc. Configure Phase This section does not apply for pulse, noise, or DC waveforms. The Phase setting is useful for adjusting the phase relationship between channels 1 and 2, between a channel and its sync signal, and synchronizing multiple instruments.
Figure 9 - Phase 6. Use the menu function keys to select the degree (°) unit. Note: When the IndependentMode is enabled, the phase parameter cannot be modified (i.e., the phase can only be set when there is a reference clock waveform provided on the back panel). Configure Duty Cycle: Square Waveform Figure 10 - Duty Cycle The duty cycle setting range is limited by the “Frequency/Period”...
Figure 11 - Duty Cycle of a Square Waveform 4. Use the menu function keys to select the degree (°) unit. Configure Width and Duty Cycle: Pulse Waveform The pulse width and duty cycle parameters are related to each other and both control the length of the “On Time”...
Follow the steps below to configure the pulse width or duty cycle. 1. Press the Waveforms button. 2. Press the Pulse softkey. 3. Select PulWidth softkey for pulse width adjustment or DutyCycle for duty cycle adjustment. 4. The cursor position will now highlight the first digit of the width or duty parameter display.
3. Toggle the Rise/Fall softkey to set the rise or fall edge. The Rise/Fall softkey lets the user toggle between the Rise and Fall Edge settings. The selected setting will be highlighted in blue. Use the numeric keyboard to input the parameter value directly and press the corresponding key to select the parameter unit.
Figure 16 - Pulse Delay 4. Use the rotary knob or the numeric keypad to change the pulse delay. After entering the numeric value, use the menu function keys to select s, ms, us, or ns. Configure Symmetry: Ramp Waveform A ramp waveform with 50% of symmetry is a triangular waveform.
Figure 18 - Ramp Symmetry 4. Use the rotary knob or the numeric keypad to change the symmetry. 5. Once the value has been entered, press the “%” softkey from the menu to set the value. Configure Standard Deviation and Mean: Noise Waveform There are two parameters that can be adjusted of the noise waveform: Standard deviation and mean.
Figure 19 - Noise Standard Deviation Figure 20 - Noise Mean 5. Use the rotary knob or the numeric keypad to change the two parameters. Both parameters can be specified in V or mV units. Configure Offset: DC Waveform The instrument can output a DC waveform output at a range of voltage levels (-12 Vdc – 12 Vdc).
2. Select and press the DC waveform softkey from the menu. 3. The cursor position will now be highlighted the first digit of the DC offset parameter display. Figure 21 - DC Offset 4. Use the rotary knob or the numeric keypad to change the two parameters. Both parameters can be specified in Vdc or mVdc units.
Page 43
Waveform Function Waveform Function StairUp Stair-up waveform Npulse Negative pulse StairDn Stair-down waveform UpRamp UpRamp waveform Stair-up and down StairUD DnRamp DnRamp waveform waveform Trapezia Trapezia waveform SineTra Sine-Tra waveform Ppulse Positive pulse SineVer Sine-Ver waveform Arbitrary Waveform: Math Waveform Function Waveform Function...
Page 44
Arbitrary Waveform: Engine Waveform Function Waveform Function Cardiac Cardiac signal SCR firing profile Quake Analog quake waveform TV signal Chirp Chirp signal Voice Voice signal TwoTone TwoTone signal Surge Surge signal SNR signal Radar Analog radar signal AmpALT Gain oscillation curve Ripple Ripple wave of battery Attenuation oscillation...
Page 46
Square waveform with 2% Square waveform with 38% SquareDuty02 SquareDuty38 duty duty Square waveform with 4% Square waveform with 40% SquareDuty04 SquareDuty40 duty duty Square waveform with 6% Square waveform with 42% SquareDuty06 SquareDuty42 duty duty Square waveform with 8% Square waveform with 44% SquareDuty08 SquareDuty44...
Page 47
Arbitrary Waveform: Square 2 Waveform Function Waveform Function Square waveform with 70% Square waveform with 86% SquareDuty70 SquareDuty86 duty duty Square waveform with 72% Square waveform with 88% SquareDuty72 SquareDuty88 duty duty Square waveform with 74% Square waveform with 90% SquareDuty74 SquareDuty90 duty...
Page 48
ECG3 Electrocardiogram ECG15 Electrocardiogram 15 Waveform of the low ECG4 Electrocardiogram LFPulse frequency pulse electrotherapy Waveform 1 of the nerve ECG5 Electrocardiogram Tens1 stimulation electrotherapy Waveform 2 of the nerve ECG6 Electrocardiogram Tens2 stimulation electrotherapy Waveform 3 of the nerve ECG7 Electrocardiogram Tens3...
Page 49
Follow the steps below to browse and select a predefined arbitrary waveform. 1. Press the button. Waveforms 2. Go to the second page of waveform options by pressing the 1/2 softkey. 3. Select and press the Arb softkey from the menu. Figure 22 - Arbitrary Waveform 4.
5. Once selected, press the knob to confirm and set the waveform or navigate to page 3/3 and press the Accept option from the menu. 6. The generator will return to the main Arb parameter menu and the waveform display area will show the waveform shape of the selected predefined arbitrary waveform.
Page 51
1. Setup Press the Waveforms button. 2. Go to the second page of waveform options by pressing the 1/2 softkey. Select and press the Arb softkey from the menu. 3. Press Arb Type and select Stored Waveforms to see a list of all arbitrary waveform files created and loaded from EasyWave software.
6. Use the knob to navigate to the file and either press the Recall softkey or press the knob to load the waveform. If the waveform is sent to the unit using EasyWave, the file will be saved to the internal memory.
Page 53
Figure 27 - Harmonic Generator Item Menu Function Type Set the harmonic type to “odd”, “ever” or “all”. Order Set the order of the harmonic. Harmonic Set the amplitude of the harmonic. Ampl Harmonic Set the phase of the harmonic. Phase Return Return to the sine parameters menu.
1. Press Order. The range is limited by the maximum output frequency of the instrument and current fundamental waveform frequency. Range: 2 to maximum output frequency of the instrument ÷ current fundamental waveform frequency The maximum is 10. 2.
source (internal/external), depth, modulating frequency, modulating waveform and carrier. DSB-AM source (internal/external), modulating frequency, modulating waveform and carrier source (internal/external), modulating frequency, frequency deviation, modulating waveform and carrier source (internal/external), phase deviation, modulating frequency, modulating waveform and carrier source (internal/external), key frequency and carrier source (internal/external), key frequency, hop frequency and carrier source (internal/external), key frequency, polarity and carrier source (internal/external), modulating frequency, width/duty cycle...
Page 56
Figure 28 - Setting Interface of AM Modulation Function Settings Description Menu Type Amplitude modulation Internal The source is internal Source The source is external. Use the [Aux In/Out] External connector at the rear panel. AM Depth Set the modulation depth. Sine Square Triangle...
Selecting Modulation Source The generator can accept a modulating signal from an internal or external modulation source. Press Mod → AM → Source to select “Internal” or “External” modulation source. The default is “Internal”. Internal Source When internal AM modulation source is selected, press Shape to select Sine, Square, Triangle, UpRamp, DnRamp, Noise or Arb as the modulating waveform.
cable. 2. Select CH1 and press Mod to select the desired modulation type as well as set the corresponding parameters, and then select external modulation source. 3. Select CH2 and select the desired modulating waveform and set the corresponding parameters. 4.
Page 59
Figure 29 - Setting Interface of DSB-AM Modulation Function Settings Description Menu Type DSB-AM DSB Amplitude modulation. Internal The source is internal. Source The source is external. Use the [Aux In/Out] External connector at the rear panel. Set the modulating waveform frequency. DSB Freq Frequency range: 1 mHz to 20 kHz (internal source) Sine...
FM Modulation The modulated waveform consists of two parts: the carrier and the modulating waveform. In FM, the frequency of the carrier varies with the instantaneous voltage of the modulating waveform. Press Mod → Type → FM; the parameters of FM modulation are shown in Figure 30. Figure 30 - Setting Interface of FM Modulation Function Settings...
Set the modulating waveform frequency. FM Freq Frequency range: 1 mHz to 20 kHz (internal source). Table 10 - FM Modulation Menu Description To Set Frequency Deviation Press FM Dev to highlight the parameter and use the numeric keyboard or arrow keys and knob to input the desired value.
Internal The source is internal Source The source is external. Use the [Aux In/Out] External connector at the rear panel. Phase Dev Phase deviation ranges from 0° to 360°. Sine Square Triangle Shape Choose the modulating waveform. UpRamp DnRamp Noise Set the modulating waveform frequency.
Figure 32 - Setting Interface of FSK Modulation Function Settings Description Menu Type Frequency shift keying modulation Internal The source is internal Source The source is external. Use the [Aux In/Out] External connector at the rear panel. Set the frequency at which the output frequency shifts between the carrier frequency and the hop Key Freq frequency (internal modulation only): 1 mHz to 50...
To Set Hop Frequency The range of the hop frequency depends on the carrier frequency currently selected. Press Hop Freq to highlight the parameter and use the numeric keyboard or knob to input the desired value. Sine: 1 uHz to max output depending on model. ...
Type Amplitude shift keying modulation Internal The source is internal Source The source is external. Use the [Aux In/Out] External connector at the rear panel. Set the frequency at which the output amplitude Key Freq shifts between the carrier amplitude and zero (internal modulation only): 1 mHz to 50 kHz.
Internal The source is internal Source The source is external. Use the [Aux In/Out] External connector at the rear panel. Set the frequency at which the output phase shifts Key Freq between the carrier phase and 180° (internal modulation only): 1 mHz to 20 kHz. Positive Polarity Set the modulating polarity...
Function Settings Description Menu Type Pulse width modulation. The carrier is pulse. Internal The source is internal Source The source is external. Use the [Aux In/Out] External connector at the rear panel. Width Set the width deviation. Duty Dev Set the duty deviation. Sine Square Triangle...
Page 68
Figure 36 - Width Deviation Setting Interface The width deviation cannot exceed the current pulse width. The width deviation is limited by the minimum pulse width and current edge time setting. Duty Deviation represents the variation (%) of the modulated waveform duty cycle relative to the original duty cycle.
Duty deviation and width deviation are dependent. Once one parameter is changed, the other will be automatically changed. Note: The methods of setting the parameters of PWM are similar to AM. 3.5 Configure Sweep Output In the sweep mode, the generator steps from the start frequency to the stop frequency in the sweep time specified by the user.
Manual Trigger a sweep manually. Disable trigger out. Trig Out Enable trigger out. Linear Set the sweep with linear profile. Type Set the sweep with logarithmic profile. Sweep upward. Direction Down Sweep downward. Table 16 - Sweep Function To select and configure the Sweep Function, follow the steps below: 1.
a. Press the Source softkey. b. Press the Internal softkey to select an internal trigger. The generator outputs a continuous sweep waveform when internal trigger is selected. The default is “Internal”. External Trigger a. Press the Source softkey. b. Press the External softkey to select an external trigger. The generator uses the trigger signal input to the [Aux In/Out] connector at the rear panel.
Sweep Type 10. In the Sweep Function menu, press the Page 1/2 softkey to access the second page of Sweep Options. “Linear” and “Log” sweep types are supported. The default type is “Linear”. A. LINEAR SWEEP In linear sweep, the output frequency of the instrument varies linearly with the sweep time.
Figure 41 - Log Sweep Interface Direction 1. In the Sweep Function menu, press the Page 1/2 softkey to access the second page of Sweep Options. 2. Press the Direction softkey to toggle between the two options: Up and Down. ...
Trig Out Enable trigger out. Cycles In N-Cycle mode, the generator will output a specified number N of waveform cycles (periods) after receiving the trigger signal. To configure Burst Cycles, follow these instructions: 1. Press the Burst button. 2. Press the NCycle softkey. 3.
This trigger softkey becomes available when the manual trigger is selected. Figure 43 – NCycle Burst Manual Trigger Gated In gated mode, the generator controls the waveform output according to the gated signal level. When the gated signal is ON, the generator outputs a continuous waveform. When the gated signal is OFF, the generator first completes the output of the current period and then stops.
Figure 44 - Negative Gated Burst To configure Burst Gated, follow the steps below: 1. Press the Burst button. 2. Press the Gated softkey. 3. Select a Polarity for the gate. This determines whether the gate will go from low to high (positive gate) or from high to low (negative gate).
Burst Trigger Source The burst trigger source can be internal, external or manual. The generator will generate a burst output when a trigger signal is received and wait for the next trigger source. To configure the Trigger Source, once in the Burst menu: Internal Trigger 1.
4 Utility Functions The utility menu can be accessed by pressing the Utility button from the front panel. This section describes all the configurable settings of this menu. Figure 45 - Utility Menu Item Description System Set the system configuration. Test/Cal Test and calibrate the instrument.
4.1 System Settings 1. Press the button. Utility 2. Pres the System softkey to enter the system options: Figure 46 - System Options Item Description Number format Set the numerical format. English Language Set the language. Chinese Default All the settings return to their defaults at power on. PowerOn All the settings return to the settings at the last power Last...
1hour 2hour 5hour Disables screensaver. System Info View system information Firmware Update Update firmware using a USB flash drive. Help View the help information. Accept Save the current settings and return to the Utility menu. Figure 47 - System Settings Numerical Format To change the numerical format of the generator, follow these instructions: a.
Press the System softkey. c. Press the Language softkey to select the desired language. Power On The 4050B series offers two options for power on settings: Last and Default. To set either option, follow these instructions: a. Press the Utility button b.
Set to Default Sets the unit to defaults settings: a. Press the Utility button. b. Press the System softkey. c. Press the Set to Default to set the system to the default setting. The default settings of the system are: Table 22 - Factory Default Settings Setting Default value...
Direction ↑ Burst Burst Period 10 ms Start Phase 0° Cycles 1 Cycle Trig Out Delay 521 ns Trigger Source Internal Key Sound (Beeper) Turns the instrument key press sound ON or OFF. When off, this setting will also disable the sound for any errors that may occur during front panel or remote operation.
Figure 49 - System Information Screen Firmware Update The software version and configuration file of the generator can be updated directly via USB flash drive. To update the firmware, follow the steps below: a. Insert USB flash drive with firmware update file (*.ADS) and configuration file (*.CFG) to USB host interface on the front panel of the generator.
Built-in Help The 4050B Series provides a built-in help system. There are 10 topics in the help list. Use the softkeys to navigate through the options and select a topic to read. To access the Built-in Help follow the below instructions: a.
4.2 Test/Cal This function allows the user to run self-diagnostics. a. Press the Utility button. b. Press the Test/Cal softkey. The following options will be available: Item Description ScrTest Run the screen test program. KeyTest Run the keyboard test program. LEDTest Run the key indicator lights test program.
Key Test a. Press the KeyTest softkey to enter the key test interface. The on-screen white rectangle shapes represent the front panel keys. The circle between two arrows represents the knob. b. Test all keys by pressing them and knob by rotating it. Also verify that all the backlight keys illuminate correctly.
Figure 54 - LED Test Board Test a. Press the BoardTest softkey to test the internal hardware of the generator. It will indicate either “pass” or “fail”. Figure 55 - Board Test 4.3 Frequency Counter The generator has a built-in frequency counter that can measure frequencies between 100 mHz to 200 MHz (DC Coupling) and 10 Hz to 200 MHz (AC Coupling).
Page 90
b. Press the Counter softkey. The following options will be available: Figure 56 - Frequency Counter Function Description Menu Turn the counter on. State Turn the counter off. Frequency Measure frequency. Period Measure period. PWidth Measure positive width. NWidth Measure negative width. Set the reference frequency.
MHz. 4.4 Output The 4050B series has output settings that can be configured to either adjust the display to match the output of the unit to high impedance or 50 Ω, change the polarity of the output, and set the output phase to be the same for both channels a.
Figure 57 - Output Setup Load For [CH1] and [CH2] connectors on the front panel, the generator has an output impedance of 50 Ω. If the load does not match the load impedance, the displayed voltage will not be the same as the output voltage setting on the generator.
Note: The Sync signal related to the waveform is not inverted when the waveform is inverted. EqPhase The unit can set both channel 1 and channel 2 to be in phase. Once the selected channel has the required phase follow the steps below to copy the phase to the other channel: 1.
Figure 58 - Wave Combine Menu Function Description Menu Output the CH1 waveform. CH1 Switch CH1+CH2 Output the CH1+CH2 waveform. Output the CH2 waveform. CH2 Switch CH1+CH2 Output the CH1+CH2 waveform. Save the current settings and exit the current Return interface.
Function Description Menu CH1=>CH2 Copy all parameters and states of CH1 to CH2. CH2=>CH1 Copy all parameters and states of CH2 to CH1. Perform the current selection and return to the Utility Accept menu. Give up the current selection and return to the Utility Cancel menu.
of the other channel will be changed automatically and always keeps the specified frequency deviation/ratio, amplitude deviation/ratio or phase deviation/ratio relative to the modified channel. To configure channel coupling settings, follow the instructions below: 1. Press the Utility button. 2. Press the CH Copy Coupling softkey. 3.
2. Press AmplMode to choose “Deviation” or “Ratio” and use the numeric keyboard or knob to input the desired value. Deviation: the amplitude deviation between CH1 and CH2. The resulting signal is represented by: AmplCH2-AmplCH1 = AmplDev. Ratio: the amplitude ratio of CH1 and CH2. The resulting signal is represented by: AmplCH2/AmplCH1 = AmplRatio.
Page 98
Figure 61 - Tracking 4. Press PhaseDev to enter the following menu: Figure 62 - Phase Deviation 5. Then use the numeric keyboard or knob and arrow keys to input the desired value for the phase deviation between CH1 and CH2. The resulting signal is represented by: PhaseCH2-PhaseCH1=PhaseDev.
4.6 Remote Interface The 4050B Series can be controlled remotely via USB, LAN and GPIB (optional) interfaces. 1. Press the Utility button. 2. Press the Page 1/2 softkey. 3. Press the Interface softkey to set the LAN parameters or GPIB address in the following...
USB Interface The instrument has a USB interface (USBTMC) in the rear panel for remote communication with EasyWave software. Note: A type A to type B USB cable is required for PC connectivity. To connect with EasyWave software, you must install NI-VISA. NI-VISA is downloadable from LabVIEW™...
Page 101
Power indicator LED GPIB Generator AK40G Communication indicator LED Note: Once connected, do not unplug the adapter on either end before powering down the instrument and the computer first. Configure GPIB address 1. Press the Utility button. 2. Press the 1/2 softkey to enter the second page of the Utility menu. 3.
Note: EasyWave software does not support GPIB interface. It must be used with USB interface only. The 4050B SERIES can communicate with a PC through a LAN interface. Users can view and modify the LAN parameters. Connect the generator to your local area network using the network cable.
Press IP Address and use the arrow keys and numeric keyboard or knob to enter the IP address. The setting is stored in non-volatile memory and will be loaded automatically when the generator is powered on. To Set Subnet Mask The format of subnet mask is NNN NNN and each NNN ranges from 0 to 255.
4.7 Sync Output The generator provides Sync output through the [Aux In/Out] connector on the rear panel. When the synchronization is on, the port can output a CMOS signal with the same frequency as basic waveforms (except Noise and DC), arbitrary waveforms, and modulated waveforms (except external modulation).
Syncing Signals of Different Waveforms Basic Waveform and Arbitrary Waveform 1. When the frequency of the waveform is less than or equal to 10 MHz, the sync signal is a Pulse with approximately 50 ns pulse width and the same frequency as the waveform.
Then set the output frequencies of all the generators to the same value to achieve synchronization. 4.9 Mode The 4050B series has two modes of operation: Phase-Locked and Independent. To select a mode of operation, follow the steps below: 1. Press the Utility button.
Page 107
Figure 67 - Phase-Locked Mode Independent Mode When changing the frequency, both channels’ DDS don’t reset and the phase deviation between CH1 and CH2 changes at random. When the independent mode is enabled, the phase parameter cannot be modified, and the Phase menu will be hidden as shown below.
4.10 Over Voltage Protection 1. Press the Utility button. 2. Press the Page 1/2 softkey. 3. Press the OverVoltage Protection to turn this function on or off. When overvoltage protection occurs, a message will be displayed and the output will be disabled.
5 Store and Recall The 4050B series can store the current instrument state and user-defined arbitrary waveform data in internal or external memory and recall them when needed. Press Store/Recall to enter the following interface. Figure 69 - Store/Recall Interface (Page 1)
State Settings files. File Type Data Arbitrary waveform file. View the current directory. Browse Save the waveform to the specified path. Save Recall the waveform or setting information in the specific position of Recall the memory. Delete the selected file. Delete Copy the selected file.
Figure 72 - Storage System Note: The 4050B series can only identify files of which filenames consist of English letters, numbers and underscore. If other characters are used, the name may be displayed in the store and recall interface with errors or not found at all.
Data File The 4050B can recall the data files in “*.csv” or “*.dat” format from the external memory and transfer them into “*.bin” format then store them in the internal memory. When it is done, the generator will enter the arbitrary waveform interface automatically. In addition, users can edit arbitrary waveforms with PC software —...
Page 113
Figure 73 - Filename Input Interface Item Description Cursor upward to select. Cursor downward to select. Down Select the current character. Select Delete the current character. Delete Store the file with the current name. Save Return to the store/Recall interface. Cancel a.
Then press Accept, the generator will delete the currently selected file. 5.6 Copy and Paste File 4050B series supports the internal and external storage to copy files from each other. For example, copy an arbitrary wave file in the U-disk to the instrument, the procedure is as follows: 1.
Page 115
2. Choose the file to be copied. Rotate the knob to select USB Device (0:) and press the knob to open its directory. Then rotate the knob to select the file you want to copy and press Page 1/2 → Copy. 3.
6 Examples The examples below use the default settings of the instrument except in special cases. This chapter includes the following topics: Example 1: Generate a Sine Waveform Example 2: Generate a Square Waveform Example 3: Generate a Ramp Waveform ...
Example 1: Generate a Sine Waveform Generate a sine waveform with 1 MHz frequency, 5 Vpp amplitude and 1 V DC offset. Set the frequency. 1. Press Waveforms → Sine →Frequency/Period and choose Frequency, which will display in a blue color. 2.
Example 2: Generate a Square Waveform Generate a square waveform with 5kHz frequency, 2 Vpp amplitude, 1 V DC offset and 30% duty cycle. Set the frequency. 1. Press Waveforms → Square → Frequency/Period and choose Frequency, which will display in a blue color.
Example 3: Generate a Ramp Waveform Generate a ramp waveform with 10 μs period, 100 mVpp amplitude, 20 mV DC offset, 45° phase and 30% symmetry. Set the Period. 1. Press Waveforms → Ramp → Frequency/Period and choose Period which will display in a blue color.
Figure 76 - Generate a Ramp Waveform Example 4: Generate a Pulse Waveform Generate a pulse waveform with 5 kHz frequency, 5 V high level, -1 V low level, 40 μs pulse width and 20 ns delay. Set the Frequency. 1.
1. Press Delay to choose Delay which will display in a blue color. 2. Input ‘20’ from the keyboard and choose the unit ‘ns’. The delay is set to 20 ns. When the frequency, high level, low level, pulse width and delay are set, the waveform generated is shown in Figure 77.
Figure 78 - Generate a Noise Waveform Example 6: Generate a DC Waveform Generate a 3 V DC waveform. Choose the DC waveform. 1. Press Waveforms →Page 1/2→ DC to choose the DC waveform. Set the Offset. 1. Press Offset and choose Offset which will display in a blue color. 2.
Figure 79 - Generate a DC Waveform Example 7: Generate a Linear Sweep Waveform Generate a sine sweep waveform whose frequency starts at 100 Hz and sweeps to a frequency of 10 kHz. Use internal trigger mode, linear sweep, and a sweep time of 2 s. ...
Set the stop frequency. Press StopFreq, input ‘10’ from the keyboard and choose the unit ‘kHz’ to set the stop frequency to 10 kHz. Set the sweep profiles. Press Type and choose Linear. When all parameters above are set, the linear sweep waveform generated is shown in Figure Figure 80 - Generate a Linear Sweep Waveform Example 8: Generate a Burst Waveform Generate a burst sine waveform with 5 cycles.
Page 125
from the keyboard and choose the unit ‘Vpp’ to set the amplitude to 4 Vpp. 3. Press Offset/LowLevel to choose Offset which will display in blue color. Input ‘0’ from the keyboard and choose the unit ‘Vdc’ to set the DC offset to 0 V DC. ...
Example 9: Generate an AM Modulation Waveform Generate an amplitude-modulated waveform with 80% depth. The carrier is a sine wave with 10 kHz frequency and the modulating wave is a sine wave with 200 Hz frequency. Set the frequency, amplitude and offset of the carrier wave. 1.
Example 10: Generate an FM Modulation Waveform Generate an FM modulation waveform: the carrier is a sine wave with 10 kHz frequency, the modulating wave is a sine wave with 1 Hz frequency and 2 kHz frequency deviation. Set the frequency, amplitude and offset of the carrier wave. 1.
Example 11: Generate a PM Modulated Waveform Generate a PM modulated waveform, the carrier is a sine wave with 10 kHz frequency, and the modulating wave is a sine wave with 2 kHz frequency and 90° phase deviation. Set the frequency, amplitude and offset of the carrier wave. 1.
Example 12: Generate an FSK Modulated Waveform Generate an FSK modulated waveform with a 200 Hz key frequency. The carrier is a sine wave with a 10 kHz frequency and the hop frequency is 500 Hz. Set the frequency, amplitude and offset of the carrier wave. 1.
Example 13: Generate an ASK Modulation Waveform Generate an ASK modulation waveform with 500 Hz key frequency. The carrier is a sine wave with 5 kHz frequency. Set the frequency, amplitude and offset of the carrier wave. 1. Press Waveforms and choose the sine waveform as the carrier wave 2.
Example 14: Generate a PSK Modulated Waveform Generate a PSK modulated waveform with 200 Hz key frequency. The carrier is a sine wave with 1 kHz frequency. Set the frequency, amplitude and offset of the carrier wave. 1. Press Waveforms and choose the sine waveform as the carrier wave 2.
Example 15: Generate a PWM Modulated Waveform Generate a PWM modulated waveform with 200 Hz key frequency. The carrier is a pulse wave with 5 kHz frequency. Set the frequency, amplitude and offset of the carrier wave. 1. Press Waveforms and choose the Pulse waveform as the carrier wave 2.
Example 16: Generate a DSB-AM Modulated Waveform Generate a DSB-AM modulated waveform with 100 Hz modulating frequency. The carrier is a sine wave with 2 kHz frequency. Set the frequency, amplitude and offset of the carrier wave. 1. Press Waveforms and choose the sine waveform as the carrier wave. 2.
Troubleshooting Guide Below are some frequently asked questions and answers. Please check if any apply to your power supply before contacting B&K Precision. Q: I cannot power up the generator o Check that the power cord is securely connected to the AC input and there is live power from your electrical AC outlet.
Need help?
Do you have a question about the 4050B Series and is the answer not in the manual?
Questions and answers