Page 1
65-2422RK-05 Hydrogen Sulfide Transmitter Operator’s Manual Part Number: 71-0114RK Revision: A Released: 2/16/11 www.rkiinstruments.com...
Page 2
Typical calibration frequencies for most applications are between 3 and 6 months, but can be required more often or less often based on your usage. 65-2422RK-05 H S Transmitter...
Page 3
Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
WARNING: When using the 65-2422RK-05, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65-2422RK-05 and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 65-2422RK-05 as described in this manual.
A pre- amplifier, located between the sockets and two interconnect wires, conditions the sensor’s signal before the signal reaches the amplifier. 2 • 65-2422RK-05 H S Transmitter...
The amplifier included with the H S transmitter is designed for use with RKI’s toxic gas and oxygen transmitters. The amplifier type selector determines for which transmitter the amplifier is intended. For H S transmitters, a jumper block is installed over the TOXIC selector (see Figure 1).
.25 Dia. Mounting 5.20 Hole, 2X 3.94 2.75 3/4 NPT Conduit Hub 5.46 6.10 8.5 max 1 1/2-20 For Calibration Cup 1.10 Figure 2: Mounting the H S Transmitter 4 • 65-2422RK-05 H S Transmitter...
Page 9
Connect the feedback wire to the terminal labeled 4/20 FB. CAUTION: If using shielded cable, leave the drain wire insulated and disconnected at the transmitter. You will connect the opposite end of the cable’s drain wire at the controller. 65-2422RK-05 H S Transmitter • 5...
S Transmitter to a Controller 11. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the controller. RKI controllers typically have a ground stud that can be used to ground the cable’s drain wire.
The procedure below describes applying zero emission air, usually called zero air, using a calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
It describes the test using a calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
The controller indicates a fail condition. Probable causes • The transmitter wiring is disconnected or misconnected. • The transmitter’s zero reading is low enough to cause a fail condition. • The transmitter is malfunctioning. 65-2422RK-05 H S Transmitter • 9...
• If the fail condition continues, replace the H S sensor. • If the fail condition continues, contact RKI for further instruction. Slow or No Response/Difficult or Unable to Calibrate Symptoms • The transmitter responds slowly or does not respond during the monthly response test.
Page 15
11. Reconnect the detector leads to the detector terminal strip as shown in Table 3 and Figure 3. Table 3: Reconnecting the H S Detector to the Amplifier Amplifier S Detector Lead Interconnect Terminal Strip Black TOXIC BK 65-2422RK-05 H S Transmitter • 11...
Page 16
10. Turn on the controller. CAUTION: Allow the replacement detector to warm up for 5 minutes before you continue with the next step. 11. Calibrate the replacement detector as described in the Calibration section of this 12 • 65-2422RK-05 H S Transmitter...
It describes the test using a calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
When the display reading falls below the alarm setpoints, return the controller to normal operation. Verify that the controller display reading decreases and stabilizes at 0 ppm. Store the components of the calibration kit in a safe and convenient place. 14 • 65-2422RK-05 H S Transmitter...
65-2422RK-05 S transmitter (includes detector and amplifier), CSA classified 65-2423RK-05 S replacement detector assembly (includes sensor), CSA classified 71-0113RK 65-2422RK-05 H S Transmitter Operator’s Manual (this document) 81-0151RK-04 Calibration cylinder (25 PPM H S in nitrogen; 34 liter aluminum) 81-0076RK-01...
Need help?
Do you have a question about the 65-2422RK-05 and is the answer not in the manual?
Questions and answers