RKI 65-2396RK Operator's Manual

Co2 transmitter

Advertisement

Quick Links

65-2396RK
CO
Transmitter
2
Operator's Manual
Part Number: 71-0184RK
Revision: A
Released: 4/12/11
RKI Instruments, Inc.
www.rkiinstruments.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 65-2396RK and is the answer not in the manual?

Questions and answers

Summary of Contents for RKI 65-2396RK

  • Page 1 65-2396RK Transmitter Operator’s Manual Part Number: 71-0184RK Revision: A Released: 4/12/11 RKI Instruments, Inc. www.rkiinstruments.com...
  • Page 2 Typical calibration frequencies for most applications are between 6 and 12 months, but can be required more often or less often based on your usage. 65-2396RK CO Transmitter...
  • Page 3 Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
  • Page 4: Table Of Contents

    Parts List ..............16 65-2396RK CO...
  • Page 5: Overview

    ± 5% of reading or ± 2% of full scale (whichever is greater) WARNING: When using the 65-2396RK, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65-2396RK and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 65-2396RK as described in this manual.
  • Page 6: Description

    Four color coded leads, red, white, green, and black, extend from the other end of the detector. The leads allow you to connect the detector to the amplifier. 2 • 65-2396RK CO Transmitter...
  • Page 7: Amplifier

    Turn the adjustment screw clockwise to increase the gas response and counterclockwise to decrease the gas response. 65-2396RK CO Transmitter • 3...
  • Page 8: Junction Box

    • Mount the transmitter at least six feet from the floor to minimize the possibility of someone breathing on the unit. The exhaled CO may cause an alarm. 4 • 65-2396RK CO Transmitter...
  • Page 9: Wiring The Co 2 Transmitter

    To gain access to a plug-in terminal strip for wiring, pull it out of its socket by grasping the terminal strip and pulling. The detector strip is keyed so that the controller and detector terminal strips cannot be reversed inadvertently. 65-2396RK CO Transmitter • 5...
  • Page 10 CAUTION: Do not route power and transmitter wiring through the same controller conduit hub. The power cable may disrupt the transmission of the transmitter signal to the controller. 11. Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown in Figure 3. 6 • 65-2396RK CO Transmitter...
  • Page 11 Transmitter to a Controller 12. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the controller. RKI controllers typically have a ground stud that can be used to ground the cable’s drain wire.
  • Page 12: Startup

    The procedure below describes applying 100% nitrogen using a calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
  • Page 13: Maintenance

    NOTE: The following procedure assumes the use of a calibration kit which includes a calibration gas cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a calibration cup for the detector, and a length of sample tubing. 65-2396RK CO Transmitter • 9...
  • Page 14 When the controller display reading falls below the alarm setpoints, return the controller to normal operation. 10. Store the components of the calibration kit in a safe place. Biannually Calibrate the CO transmitter as described in “Calibration” on page 14 of this manual. 10 • 65-2396RK CO Transmitter...
  • Page 15: Troubleshooting

    4. If the calibration/response difficulties Note: Under “normal” continue, replace the detector. circumstances, the transmitter requires 5. If the calibration/response difficulties calibration once every 6 continue, contact RKI for further months. instruction. Some applications may require a more frequent calibration schedule.
  • Page 16 4-20 mA output is stabilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication.
  • Page 17 4-20 mA output is stabilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication.
  • Page 18: Calibration Frequency

    6 months is adequate for most infrared CO transmitter applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 6 months. If an application is not very demanding, for example detection in a clean, temperature controlled environment where calibration adjustments are minimal at calibration, then a calibration frequency of every 9 to 12 months is adequate.
  • Page 19: Setting The Zero Reading

    The 0-50 %volume and 0-100 %volume detectors will not display a reading in fresh air because their ranges are too large. Store the components of the calibration kit in a safe and convenient place. 65-2396RK CO Transmitter • 15...
  • Page 20: Parts List

    Steel calibration cylinder, 100% nitrogen 34-liter 81-1050RK Regulator, 0.5 liter/minute; with pressure gauge and flow control knob, for 17 liter and 34 liter steel calibration cylinders 81-1103RK Calibration cup 82-0006RK Pot adjust screwdriver, for calibration 16 • 65-2396RK CO Transmitter...

Table of Contents