Page 1
65-2396RK Transmitter Operator’s Manual Part Number: 71-0184RK Revision: A Released: 4/12/11 RKI Instruments, Inc. www.rkiinstruments.com...
Page 2
Typical calibration frequencies for most applications are between 6 and 12 months, but can be required more often or less often based on your usage. 65-2396RK CO Transmitter...
Page 3
Product Warranty RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials, workmanship, and performance for a period of one year from date of shipment from RKI Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our option, free of charge.
± 5% of reading or ± 2% of full scale (whichever is greater) WARNING: When using the 65-2396RK, you must follow the instructions and warnings in this manual to assure proper and safe operation of the 65-2396RK and to minimize the risk of personal injury. Be sure to maintain and periodically calibrate the 65-2396RK as described in this manual.
Four color coded leads, red, white, green, and black, extend from the other end of the detector. The leads allow you to connect the detector to the amplifier. 2 • 65-2396RK CO Transmitter...
• Mount the transmitter at least six feet from the floor to minimize the possibility of someone breathing on the unit. The exhaled CO may cause an alarm. 4 • 65-2396RK CO Transmitter...
To gain access to a plug-in terminal strip for wiring, pull it out of its socket by grasping the terminal strip and pulling. The detector strip is keyed so that the controller and detector terminal strips cannot be reversed inadvertently. 65-2396RK CO Transmitter • 5...
Page 10
CAUTION: Do not route power and transmitter wiring through the same controller conduit hub. The power cable may disrupt the transmission of the transmitter signal to the controller. 11. Connect the wires to the applicable detector/transmitter terminal strip at the controller as shown in Figure 3. 6 • 65-2396RK CO Transmitter...
Page 11
Transmitter to a Controller 12. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the controller. RKI controllers typically have a ground stud that can be used to ground the cable’s drain wire.
The procedure below describes applying 100% nitrogen using a calibration kit that includes a calibration cup, calibration gas, sample tubing, and a fixed flow regulator with an on/off knob. RKI Instruments, Inc. recommends using a 0.5 LPM (liters per minute) fixed flow regulator.
NOTE: The following procedure assumes the use of a calibration kit which includes a calibration gas cylinder, a 0.5 LPM fixed flow regulator with an on/off knob, a calibration cup for the detector, and a length of sample tubing. 65-2396RK CO Transmitter • 9...
Page 14
When the controller display reading falls below the alarm setpoints, return the controller to normal operation. 10. Store the components of the calibration kit in a safe place. Biannually Calibrate the CO transmitter as described in “Calibration” on page 14 of this manual. 10 • 65-2396RK CO Transmitter...
4. If the calibration/response difficulties Note: Under “normal” continue, replace the detector. circumstances, the transmitter requires 5. If the calibration/response difficulties calibration once every 6 continue, contact RKI for further months. instruction. Some applications may require a more frequent calibration schedule.
Page 16
4-20 mA output is stabilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication.
Page 17
4-20 mA output is stabilizing and may be above the controller alarm points or well below zero momentarily. RKI controllers have a one minute warmup period when the controller does not display any gas reading or give any alarm indication.
6 months is adequate for most infrared CO transmitter applications. Unless experience in a particular application dictates otherwise, RKI Instruments, Inc. recommends a calibration frequency of every 6 months. If an application is not very demanding, for example detection in a clean, temperature controlled environment where calibration adjustments are minimal at calibration, then a calibration frequency of every 9 to 12 months is adequate.
The 0-50 %volume and 0-100 %volume detectors will not display a reading in fresh air because their ranges are too large. Store the components of the calibration kit in a safe and convenient place. 65-2396RK CO Transmitter • 15...
Steel calibration cylinder, 100% nitrogen 34-liter 81-1050RK Regulator, 0.5 liter/minute; with pressure gauge and flow control knob, for 17 liter and 34 liter steel calibration cylinders 81-1103RK Calibration cup 82-0006RK Pot adjust screwdriver, for calibration 16 • 65-2396RK CO Transmitter...
Need help?
Do you have a question about the 65-2396RK and is the answer not in the manual?
Questions and answers