Foreword Thank you for purchasing a MultiTherm™ Shaker. This operations manual contains instructions for the care of this Instrument. In order to use the instrument properly, please read this manual carefully before operation. Please keep it available for future use. Prior to the initial operation: Please check the Instrument and the accessories against the packing list when you first open the shipping carton.
Page 3
Safety Warnings and Guidelines 1. Important operation information: Please read this manual carefully prior to any attempt to operate the Instrument. Read the manual prior to power up of the instrument. Read the guidelines and directions below and be familiar with all safety warnings and guidelines.
Page 4
Before connectivity to a power source, check that the voltage listed on the serial number label (on the bottom or back side of the unit) to ensure that it matches the voltage available in your area. Be sure to install the Instrument in a cool, dry location away from any direct sunlight or strong light sources.
H5000 Series MultiTherm Shakers Chapter 1: Introduction The MultiTherm shaker is an ideal instrument for intensive mixing of samples under temperature regulated conditions. Mixing and temperature control modes can be used both simultaneously and independently i.e, the device can work as shaker and/or as an incubator. The base instrument can be used with a variety of different blocks.
H5000 Series MultiTherm Shakers 2. Technical data, H5000-H Model H5000-H Parameter Mixing rate* 200~1500 rpm Orbit Temperature setting range 5°C to 100°C Temperature control range* Ambient +5°C to 100°C Timing range 1min to 99h59min Accuracy of the temperature ≤0.5°C Heating time ≤15min (from 20°C to 100°C)...
Page 8
H5000 Series MultiTherm Shakers Technical data, H5000-HC Model H5000-HC Parameter Mixing rate* 200~1500 rpm Orbit Temperature setting range 0° to 100°C 0° to 100° at Room temp. ≤20°C 4° to 100° at Room temp.≤25°C Temperature control range* 10° to 100° at Room temp.≤30°C Timing range 1min ~ 99h59min Accuracy of the temperature...
H5000 Series MultiTherm Shakers Chapter 3: Preparations This chapter provides a physical description, location of controls, function of individual keys, as well as some preparations before the initial power-on. You should be familiar with this chapter before the Thermo Shaker is first operated. 1.
H5000 Series MultiTherm Shakers 2. Keyboard and Display panel Display Panel: Remaining time Current program Current temperature Current speed P1 100.0 1200 10:30 P1 100.0 1200 15:00 Selected time Selected temperature Selected program Selected speed ─ ─...
H5000 Series MultiTherm Shakers 3. Key Functions Program Selection Keys: Use the up and down keys to Prog. choose the desired program (P1, P2, P3, P4 or P5). Temperature Selection Keys: Press“” or “” to select the Temp. desired temperature in increments of 1°C.
H5000 Series MultiTherm Shakers Chapter 4: Operation Guide 1. Single temperature, speed and time setting a) Following power up, the unit beeps once and System‐Testing the LCD will display “System Testing”. After .. approx. 5 seconds, the current and selected settings appear on the display and the instrument is ready for use.
H5000 Series MultiTherm Shakers NOTE: Press and hold the “” or “” adjustment keys quickly scroll through available temperature, speed and time range Star/Stop Press the Key to begin the cycle Sequence Link Program Setting (Available on H5000-HC only) By Pressing the “Seq. Link” Key, the user can connect up to five programs to run consecutively (P1-P2, P1-P2-P3, P1-P2-P3-P4, P1-P2-P3-P4-P5).
H5000 Series MultiTherm Shakers To deactivate the shaker speed, press and hold the “” Speed key until the set speed shows “OFF”. The shaker is now deactivated and the instrument can be used solely for temperature control. To deactivate the timer, press and hold the “” Time key until the set time shows “OFF”.
Page 15
To enter the temperature calibration mode, press ADJ P:20.5 the “” and “” Prog. Keys simultaneously. AdjTemp= 10.0 The instrument begins to cool to the low calibration point. - H5000-H P:10.0 ADJ Calibration points: 40°C, 100°C AdjTemp= 10.0 - H5000-HC Calibration points: 10°C, 40°C, 100°C ADJ ...
Page 16
H5000 Series MultiTherm Shakers ADJ P:40.1 d) Repeat the above step for the remaining AdjTemp= 40.0 temperature calibation points. Following the adjustment to the final calibration point, the temperature calibration of the instrument has now been completed and the displayed temperature will Adj ...
H5000 Series MultiTherm Shakers 6.Changing the block a) Open the transparent lid and use the supplied hex wrench (3mm) to remove the four screws that hold the block to the heating board. b) Close the lid and lift the block up and away from the base unit.
Chapter 5: Failure analysis and troubleshooting Failure analysis and processing procedures Symptom Possible Causes Action No power Check the power Exchange fuse(250V 3.0A Fuse failure display when 5X20mm) powered up. Broken switch Replace the switch Others Call for service actual displayed Temperature sensor failure Call for service...