SAMPLE SCREEN TEST ..........................22 TROUBLESHOOTING ........................23 SYMPTOMS ..............................23 MAINTENANCE, ERROR AND OTHER MESSAGES .................. 23 CUSTOMER SERVICE INFORMATION ..................24 APPENDIX A: SAMPLE PREPARATION METHOD ................25 FluoroCheck II Revision 3.0 P a g e 3 | 25...
IEC 61326-1:2012 Power Supply Safety cULus, CE Specification: NOTE: The above specification and protection impairment may be affected if the equipment is not used as specified by the manufacturer. FluoroCheck II Revision 3.0 P a g e 4 | 25...
Indicates a potentially or imminently hazardous situation which, if not DANGER avoided, will result in death or serious injury. Indicates a situation which, if not avoided, may cause damage to the NOTICE instrument. Information that requires special emphasis. FluoroCheck II Revision 3.0 P a g e 5 | 25...
LCD display in PPM or NTU or FLR value DESCRIPTION The FluoroCheck II is a bench top Oil in Water Monitor with a fixed excitation bandpass filter (354 nm) and an emission bandpass filter (450 nm). It is designed specifically for the quantification of low ppm concentration measurements of hydrocarbons (oil) in water.
If this equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired. • Only accessories and parts approved or supplied by Arjay Engineering may be used for operating, maintaining, and servicing this product. •...
The FluoroCheck can store up to 10 different calibrations: • Location 0 (DIESEL1) is factory calibrated for solvent extraction testing of refined oils. • Location 1 (CRUDE1) is factory calibrated for solvent extraction testing of crude oils. FluoroCheck II Revision 3.0 P a g e 8 | 25...
Page 9
Crude Oil: REF A Refined Oils (e.g. Diesel): REF B Periodic re-calibrations using actual prepared or analyzed samples may be desired to maintain a good correlation against local laboratory results. FluoroCheck II Revision 3.0 P a g e 9 | 25...
POWER UP AND MAIN DISPLAY Plug in the power supply to an AC source. Plug in the DC jack into the back of the Fluorocheck II and turn the power switch to on (I). If already powered on, turn the switch off then on again to start this tutorial.
Page 11
The unit does not mathematically convert from one type of unit to another after calibration. Choose your desired display mode prior to calibration. Select the desired units now to continue this tutorial. The SETTINGS menu will return. FluoroCheck II Revision 3.0 P a g e 11 | 25...
Page 12
S o l v e n t E x t r a c t i o n S a m p l e S c r e e n i n g FluoroCheck II Revision 3.0 P a g e 12 | 25...
Operations can be monitored from the RS232 port if a computer or printer is connected to the unit. This may be desirable if logging or printed records are required. FluoroCheck II Revision 3.0 P a g e 13 | 25...
Page 14
The RS232 serial port of FluoroCheck II is a DB9-pin female connector. Connect the serial port of FluoroCheck II to computer serial port. The FluoroCheck II requires these settings in the device receiving the data: Baud rate 9600 Data bits...
2. Add 10 ml of solvent to your calibration jar using the Solvent dispenser. Do not add solvent to the jar tagged for the lab. 3. Close the lid on the jar and shake the sample for 2 minutes. FluoroCheck II Revision 3.0 P a g e 15 | 25...
8. Wipe the cuvet with clean tissue prior to testing. LED INTENSITY SET-UP The Fluorocheck II uses a UV LED that is driven by current. The more current, the more intense the LED. Different oil types yield different fluorescence energy. Refined and light oils yield lower fluorescence energy and need more UV light.
Select a Calibration Location from # 4 to # 9 to store your calibration, then press <ENTER>. Locations 0-3 are reserved for Factory Calibrations and cannot be changed. The screen will display the following: FluoroCheck II Revision 3.0 P a g e 17 | 25...
Page 18
Fill the appropriate cuvet to about 3/4 full with the sample and insert into the FluoroCheck. The raw fluorescence (FLR value) is displayed (The FLR value should be greater than Cal point 1) Wait a few seconds for this reading to stabilize and then Press <ENTER>. FluoroCheck II Revision 3.0 P a g e 18 | 25...
Press the <DISPLAY> key to exit the Calibration Menu. Go to Section 5.5. to complete the calibration. SETTING THE REFERENCE CUVET You can now assign a value to your Reference Cuvet. Crude Oil: REF A Refined Oils (e.g. Diesel): REF B FluoroCheck II Revision 3.0 P a g e 19 | 25...
Page 20
It is not indicative of other hydrocarbon types or sites. Periodic re-calibrations using actual prepared or analyzed samples may be desired to maintain a good correlation against local laboratory results. FluoroCheck II Revision 3.0 P a g e 20 | 25...
8. Use the cuvet holder for the 4mm cuvet. Insert the cuvet or holder into the FluoroCheck and check that the cuvet holder dots are matched. Close the lid. FluoroCheck II Revision 3.0 P a g e 21 | 25...
FluoroCheck to remove any contaminants from handling. 2. Insert into the FluoroCheck and close the lid. 3. The reading will be displayed automatically. 4. Dispose of the cuvet and water sample according to local regulations. FluoroCheck II Revision 3.0 P a g e 22 | 25...
different ppm values were attempted to be entered for the same sample concentration. Overrng: Lower LED mA The mV reading is too high, you need to lower LED current and recalibrate the instrument. FluoroCheck II Revision 3.0 P a g e 23 | 25...
CUSTOMER SERVICE INFORMATION Arjay Engineering offers complete technical support for all our products. If you have any questions about how to use this product, or would like to make an arrangement to repair it, please call or e-mail Arjay or your local Arjay representative.
This is not a precise method. Standing time, the volatility of the oil, operator measurements and equipment will all contribute to errors in the blend. This NOTICE should be used quickly and only when actual process conditions cannot be used for calibration. FluoroCheck II Revision 3.0 P a g e 25 | 25...
Page 26
EU Declaration of Conformity Product: Oil In Water Monitor (Bench Top) Model(s): Fluorocheck II (FCII) Manufacturer: Arjay Engineering Ltd. 2851 Brighton Road, Oakville, Ontario, Canada L6H6C9 This declaration of conformity is issued under the sole responsibility of the manufacturer. The object of the...
Page 27
BY: ARJAY ENGINEERING LTD. For service, call ARJAY Engineering Ltd. directly: Canada (905) 829-2418 North American Toll Free: 1-800-387-9487 Fax: (905) 829-4701 Internet: www.arjayeng.com...