34
DC drift:
CMRR (differential):
Input crosstalk:
Input noise:
Pod connectors:
Supply voltage:
Maximum current:
Communications:
Signal input:
Connector type:
Sampling
ADC resolution:
Linearity error:
Maximum sampling rates:
Available sampling rates:
Output Amplifier
Output configuration:
Output resolution:
Maximum output current:
Output impedance:
Slew rate:
Settling time:
Linearity error:
Software-corrected
96 dB @ 50 Hz (typical)
–110 dB typical
<2.4 µV
referred to input, DC to 20 kHz
rms
2
Combine power, I
C and single-ended or
differential analog input signals on one
connector, support pods, particular
transducers, etc.
±5 V regulated
50 mA per pod port
2
2-wire I
C
Positive and negative analog inputs
8-pin DIN.
16 bits (150 µV resolution at 10 V range)
±2 LSB (from 0 °C to 70 °C)
100 kHz on one channel;
40 kHz on two channels.
100 kHz down to 0.003 Hz using Chart
100 kHz down to 2 Hz using Scope
100 kHz to 1 Hz using Chart for Windows
Single-ended or differential (complementary)
16 bits (150 µV resolution at 10 V range)
±15 mA typical
0.1 Ω typical
6 V/µs
2 µs (to 0.01% of FSR for LSB change)
±1 LSB (from 0 °C to 70 °C)