  • Page 1 USER MANUAL & QUICK START GUIDE...
  • Page 3: Table Of Contents

    BREASTSHIELD SIZING ....................21 SIGNS YOUR BREASTSHIELD IS TOO SMALL ............22 SIGNS YOUR BREASTSHIELD IS TOO LARGE ............22 FOR YOUR INFORMATION ..................23 ASSEMBLING THE MILK GENIE ................... 24 ASSEMBLY INSTRUCTIONS ..................24 POWER ADAPTER ....................27 PREPARING TO PUMP ....................29 OPERATING THE MILK GENIE ..................
  • Page 4 FUNCTIONS AND CONTROLS .................. 33 MASSAGE MODE & EXPRESSION MODE ..............34 VACUUM & CYCLE SPEED ..................35 MEMORY MODE ...................... 36 BATTERY LIFE ......................37 MILK STORAGE ......................38 IMPORTANT SAFEGUARDS ..................40 THAWING FROZEN MILK ..................... 40 FEEDING BREAST MILK ....................40 WARMING MILK ......................
  • Page 5: Product Information

    FEATURES  Quiet Motor: The Milk Genie features a quiet motor that reaches a maximum of 270 mmHg making it strong enough to maintain supply and ensure you can pump quickly and effectively - and it is super quiet! ...
  • Page 6 Genie’s backlight emits enough light to see in the dark.  Battery: With 2.5 hours of battery life, you don’t need to be tethered to the wall when pumping with the Milk Genie.  Stylish, Portable Design: Weighing in at just 265 grams (.58 pounds) and small enough to fit in the palm of your hand, your purse, or your pocket.
  • Page 7: Package Contents

    The Milk Genie comes complete with a Power Adapter, and to accommodate single or double ❶ pumping, two each of Tubing, Tubing Connectors, Backflow Protectors, Valves, Breastshields (21mm, 24mm, 27mm), Connectors, Bottles, Bottle Caps, Soft Flow Teats, Bottle Necks, and Bottle Sealing Discs.
  • Page 8 ❷ ❸ ❹...
  • Page 9 ❺ ❼ ❻ ❽...
  • Page 10: Quick Start Guide

    See also: Operating the Milk Genie (page 30).
  • Page 11 See also: Assembling the Milk Genie (page 24).
  • Page 12 SINGLE-PUMPING DOUBLE-PUMPING POWER ADAPTER ASSEMBLY POWER ADAPTER USE See also: Assembling the Milk Genie (page 24).
  • Page 13 See also: Operating the Milk Genie (page 30).
  • Page 14 BREASTSHIELD SIZING  Pump in Expression Mode for 5 Minutes  Measure Diameter of Swollen Nipple  Select Shield 2-3mm Wider than Nipple Signs of a good fit are no pain, rubbing, redness, or whiteness of the nipple; and adequate milk expression.
  • Page 15 FOR MORE INFORMATION See the following pages in this manual for more comprehensive information on the topics covered in the Quick Start Guide. Or, visit us at www.pumpables.co Or, contact our expert customer service team: facebook.com/pumpables twitter.com/pumpables hello@pumpables.co We're pumping mamas too and we understand. Every one of our customer service team can assemble a breast pump blindfolded, and we love to help other pumping mamas.
  • Page 16: Important Safeguards

    HYGIENE  The Milk Genie is a single user breast pump. It is not intended for multiple users.  The Milk Genie breast pump is a closed system with the use of the backflow protector, preventing milk from entering the system tubing.
  • Page 17 Unplug and store out of reach of children when not in use. If you are pregnant, please consult your healthcare provider prior to using the Milk Genie. Pumping can cause contractions, and in extreme cases induce labor. If you are pregnant while nursing another child, please consult your healthcare provider prior to using this product.
  • Page 18: Cleaning & Care

    Note: These guidelines are for use with a healthy, full-term baby and assume that you have access to potable water. If your baby is premature or has health challenges your care provider may supply alternative advice. If you live in an area without potable water, you should always sterilize parts before use.
  • Page 19: Cleaning & Care Tips

    STERILIZING BREASTPUMP PARTS You can use any method you like to sterilize the parts of your Milk Genie. Please remember that you do not need to sterilize the tubing or pump motor. In most cases, you do not need to sterilize after every use of your breast pump;...
  • Page 20: Boiling

    The Milk Genie is a single-user pump. Choose from any of the following instructions to sterilize pump parts: Boiling 1. Place all the equipment in a large sauce pan and cover with tap water.
  • Page 21: Sterilization Tips

    Sterilization Tips  Tip: Frequent microwave sterilizing of breast pump parts and bottles may cause discoloration. This does not affect usability or safety of the parts.
  • Page 22: Breastshield Sizing

    To ensure you pump with maximum efficiency and comfort, the Milk Genie comes with 3 different sized breastshields - 21mm, 24mm, and 27mm - to help you get the best fit. Follow the breastshield sizing guide below to determine your recommended size.
  • Page 23: Signs Your Breastshield Is Too Small

    Note: The above image is intended for guideline purposes only and by no means is representative of all women. Breastshield sizing is trial and error and a good fit may look different for every woman. SIGNS YOUR BREASTSHIELD IS TOO SMALL ...
  • Page 24: For Your Information

    One breast may perform better with a different sized breastshield than the other. This is why Pumpables includes three sizes of breastshields with the Milk Genie; so, it’s less likely you’ll have to worry about finding the perfect size, and more likely you’ll start your pumping journey without a hitch! If you are experiencing friction, pain, or discomfort while pumping, contact us or a certified lactation consultant for help.
  • Page 25: Assembling The Milk Genie

    Before using your Milk Genie for the first time, you will need to sterilize all parts, except for the pump motor and tubing (sterilizing the pump motor and tubing can cause irreparable damage). For more information on sterilizing and washing the parts that come with your Milk Genie, please refer to the Cleaning &...
  • Page 26 1. Attach the silicone duck valve to the underside of the connector. 2. Attach the breastshield to the connector. For more information on selecting the correct breastshield, refer to the Breastshield Sizing section of this manual (page 21). 3. Screw the milk collection bottle into the connector. 4.
  • Page 27 7. Connect the tubing to the backflow protector. 8. Connect the tubing to the Milk Genie. For double-pumping, remove the cap on the airport (see below, left). For single-pumping, make sure this cap remains in place (see below, right). 9. Insert power adapter, if necessary.
  • Page 28: Power Adapter

    POWER ADAPTER The power adapter comes with interchangeable parts for localization to your region (US, AU, UK, EU and Asia). Connect the power charger and the appropriate adapter by sliding the adapter down into the charger (see below). You will hear a loud click once they are fully assembled.
  • Page 29 To remove the adapter, press down on the button labeled “Push” and slide it back up and out (see below). Note: The power adapter is for internal use only. The adapter is FDA compliant and adheres to US and Canadian component requirements.
  • Page 30: Preparing To Pump

     To maintain suction and overall functionality of the pump, silicone parts need to be replaced regularly. Replacements are available on the Pumpables website (pumpables.co); contact us if you need assistance. Replacement Frequency Replacement Frequency...
  • Page 31: Operating The Milk Genie

    The Milk Genie is designed to operate as either a single pump or double pump. The pump can be operated either on battery power, or while plugged in. The Milk Genie comes with an integrated timer so you can easily track how long you have been pumping. After 30 minutes of operation, the Milk Genie will automatically turn off.
  • Page 32: Icons

    Increases Vacuum Intensity  Stops the stored program in the Memory Function (when held down for 2 seconds) ICONS The following icons display on the LED screen of the Milk Genie. Function Overview  Memory Displays when there is a stored Mode program in the Memory Mode function.
  • Page 33 Function Overview  Massage Displays when Milk Genie is is in Mode Massage Mode.  Flashes when Massage Mode is being played back as part of a stored program in the Memory Mode function.  Expression Displays when Milk Genie is is in Mode Expression Mode.
  • Page 34: Functions And Controls

    FUNCTIONS AND CONTROLS...
  • Page 35: Massage Mode & Expression Mode

    To utilize these features, observe the following instructions: 1. Press the power button to turn on your Pumpables Milk Genie. 2. Press the Mode button to switch to Massage Mode. The screen will display the icon.
  • Page 36: Vacuum & Cycle Speed

    VACUUM & CYCLE SPEED To best mimic your baby’s suckling, the Milk Genie’s vacuum intensity and cycle speeds are codependent. You will notice that as you increase the vacuum level, the cycle speed will lengthen. This mimics how your baby nurses.
  • Page 37: Memory Mode

    MEMORY MODE The Milk Genie comes with an integrated Memory feature to make pumping easier by storing your favorite pumping program. To utilize the Memory feature, observe the following instructions: 1. While the pump is powered on, hold down the button for 2 seconds to activate the Memory function.
  • Page 38: Battery Life

    BATTERY LIFE The Milk Genie comes with approximately 2.5 hours of battery life, enabling you to pump on the go. The Milk Genie performs best when the battery is fully charged. You may experience a decrease in vacuum intensity as the battery is nearly depleted; this means it’s time to plug in your Milk Genie to a power outlet.
  • Page 39: Milk Storage

    The following chart describes how and where to safely store your expressed breast milk, based from recommendations outlined by the United States Centers for Disease Control and Prevention (CDC). If your care provider has provided alternative instructions, always adhere to those. Preferred Acceptable Duration...
  • Page 40 Preferred Acceptable Duration Duration Fresh Breast Milk after Feeding At room temperature 1 hour 2 hours (up to 26°C / 78°F) In the cooler bag with an ice pack Up to 4 hours Up to 4 hours (up to 15°C / 59°F) In the back of the refrigerator Up to 24 hours...
  • Page 41: Important Safeguards

    IMPORTANT SAFEGUARDS  Do not refreeze previously frozen breast milk once thawed.  Do not overheat or microwave breast milk as much of its nutritive value will be lost and hot spots can occur and result in serious burns. There are three ways to thaw frozen breast milk safely. If your caregiver has given you alternative instructions, always adhere to those.
  • Page 42: Safely Bottle-Feed

     Do not heat milk in a microwave. Microwaving can cause severe burns from hot spots that develop in the milk during the heating process.  Do not heat in boiling water or water that is too hot to touch. ...
  • Page 43: Troubleshooting

    The following section contains common issues and resolutions. Should your issue not be resolved by following the contents of the following section, please contact our customer support team. POSSIBLE PROBLEM SOLUTION Pump will Not Turn On Out of batteries. Insert power adapter. Interchangeable parts on Assemble adapter as per power adapter not...
  • Page 44 POSSIBLE PROBLEM SOLUTION Low Suction (cont.) Silicone valve is not attached Reposition the valve and then correctly. reassemble. Backflow protectors are not Reassemble and reattach the assembled correctly. backflow protectors. Tubing is not fully attached. Firmly attach tubing to both the backflow protector and the pump nozzle.
  • Page 45: Need More Help

    NEED MORE HELP? Visit us at www.pumpables.co Or, contact our expert customer service team: facebook.com/pumpables twitter.com/pumpables hello@pumpables.co...
  • Page 46: Technical Specifications

    Vacuum Range: 20 - 270 mmHg Cycle Speed (Expression Mode): 25 - 60 RPM Cycle Speed (Massage Mode): 70 RPM Net Weight: 265 g (0.58 lbs) Warranty: 2 Years Battery Life: 2.5 Hours Time to Charge: 2.5 Hours Motor Life: 350 Hours Country of Origin: South Korea Certification: ISO 9001, 13485 Color: Teal and White...
  • Page 47: Performance Characteristics

    PERFORMANCE CHARACTERISTICS Cycles per Minute 24 28 32 38 41 47 49 53 57 60 63 70 Note: Massage Mode is always 70 Cycles per Minute Vacuum Level (mmHg) Massage Single Pump Expression Massage Double Pump Expression...
  • Page 48: Warranty

    This product is warranted by Pumpables to be free from defects in material and workmanship for a period of two years from the date of purchase. This warranty does not cover damage caused by accident, misuse, abuse, improper maintenance, unauthorized modification, or connection to an improper power supply.
  • Page 49: Disposal

    The Milk Genie must not be disposed of with your regular household waste. It must be taken to a designated electronics recycling location. For more information about where you can dispose of electronic waste, please contact your local authority. To comply with EU Directive 2002/96/EC on Waste Electrical...
  • Page 50: Electromagnetic Compatibility

    Genie while in operation. Do not use adjacent or stacked with any other electronic equipment. EMISSIONS The Milk Genie has been tested to comply within the limits of a Class B digital device, meaning it is safe for use at home and in public establishments; specifically, anywhere connected to a low-voltage power...
  • Page 51: Electromagnetic Emissions

    Electromagnetic Emissions The Milk Genie is intended for use in the electromagnetic environment specified below. The customer or the user of the pump should ensure that it is used in such an environment. Emissions Compliance Electromagnetic Environment Test Guidance RF Emissions...
  • Page 52: Electromagnetic Immunity

    Electromagnetic Immunity The Milk Genie is intended for use in the electromagnetic environment specified below. The customer or the user of the pump should ensure that it is used in such an environment. Immunity Test Level Compliance Electromagnetic Test Level...
  • Page 53 0.5 cycle on power supply for 0.5 cycle 40% UT input lines. 40% UT If the user of the Milk Genie (60% dip in UT) IEC61000-4-11 requires continued (60% dip in UT) for 5 cycle, 6 operation during power...
  • Page 54 Immunity Test Level Compliance Electromagnetic Test Level Environment Guidance Power Frequency 30 A/m 30 A/m (50/60 ㎐) Magnetic Field FDA Home Use Guidance Note: UT is the a.c. main voltage prior to application of the test level.
  • Page 55: For More Information

    FOR MORE INFORMATION Visit us at www.pumpables.co Or, contact our expert customer service team: facebook.com/pumpables twitter.com/pumpables hello@pumpables.co We're pumping mamas too and we understand. Every one of our customer service team can assemble a breast pump blindfolded, and we love to help other pumping mamas.
  • Page 56 © 2017 Pumpables...