Model 923 Data Logger – User Manual
6.2.4.5 Time Synchronization
The 923-GUI allows the user to synchronize the 923-EDL real time clock (RTC) with the user's computer time.
If time synchronization is enabled and a time discrepancy of greater than or equal to one (1) minute is measured,
a prompt will be raised to ask the user if the 923-EDL RTC should be synchronized with the local computer (see
figure below). By pressing "synchronize", the 923-EDL RTC will be synchronized with the PC clock and by
pressing "Cancel" the time synchronization feature will be disabled. This window will timeout in 10 seconds and
assume time synchronization is not desired.
Figure 6-7: 923-GUI – Time Synchronization of 923-EDL RTC's with PC
Focal Technologies Corporation
Page 6-10
A Moog Inc. Company
Document No: 923-0603-01, Rev. 2