Reboot Udl2; Configuration Group; Timing; Figure 19 Detailed Status Dialog - Cortalk uDL2 Configuration & Operation Manual

Data logger for cp coupons
Table of Contents

Advertisement

Figure 19 Detailed Status Dialog

4.3.1.11

Reboot uDL2

The "Reboot uDL2" is used to restart the uDL2 application.

4.3.2 Configuration Group

All items in the configuration group may be reset back to factory defaults by clicking on the "Defaults" button.

4.3.2.1 Timing

The "Timing" section allows you to select various timing parameters. The "Sampling interval" (1 second to 12 hours)
determines how often the uDL2 takes a sample for each enabled measurement type. Between samples, the uDL2 enters
a low power sleep mode to conserve battery power.
The "Disconnect Cycle Period" determines how often the coupon interruption relay will be opened for instant-off and
de-polarization measurements. The "Disconnect Duration" value is how long the relay is left open for each disconnect
event. After each relay transition, the uDL2 will wait before taking the next DC potential measurement; this wait time
is determined by one of the two "Sampling Delay" parameters.
"Disconnect Holdoff" sets how long the uDL2 will wait after the analog acquisition starts before the first coupon
disconnect cycle will occur. After the initial disconnect event, disconnect events will repeat at the "Disconnect Cycle
Period" setting.
When the uDL2 is operated in dual-potential channel mode, only the "Sampling interval" timing parameters apply. All
other timing parameters only apply to coupon mode.