Quick Reference Sheet for Basal and Bolus Options
Deliver a Bolus using ezBG.
1.
From the Home
2 : 27
PM
screen, select
"Menu."
Basal Rate
Press
0.025U/Hr
Insulin: 105U
Status
Menu
5.
ezBG
Press
mg/dL
Actual
220
Target
-
120
= +
100
ISF
50 mg/dL
Show Result
Main Menu
Deliver a Bolus using ezCarb.
1.
From the Home
2 : 27
PM
screen, select
"Menu."
Basal Rate
Press
0.025U/Hr
Insulin: 105U
Status
Menu
5.
Press
If not entering a
BG value, select
"Show Result"
and go to step 8.
9.
Press
to
Bolus Total
Carb
2.26U
deliver bolus.
BG +
2.00U
IOB -
1.00U
Total =
3.25U
3.25U
Go
Type
Normal
Main Menu
Insulin Pump and CGM System
®
2.
From the MAIN MENU,
MAIN MENU
select "Bolus."
Bolus
Suspnd/Resum
Press
History
Basal
Setup
Prime/Rewind
Status
Home
6.
/
Use
ezBG Total
to enter the
BG
+
2.00U
IOB
-
1.00U
suggested
Total =
1.00U
bolus amount.
0.00U
Press
Go
Main Menu
2.
From the MAIN MENU,
select "Bolus."
Press
6.
Use
/
to
BG CORRECT
mg/dL
enter BG value.*
Actual
220
Press
Target
-
120
= +
100
ISF
50 mg/dL
Show Result
Main Menu
10.
This screen will be
displayed only if
using Dexcom G4
PLATINUM Sensor
and Transmitter
with pump.
3.
From the BOLUS MENU,
BOLUS MENU
scroll to "ezBG."
Normal
ezCarb
ezBG
Press
Combo Bolus
Reminders
Main Menu
7.
ezBG Total
Press
BG
+
2.00U
deliver bolus.
IOB
-
1.00U
Total =
1.00U
1.00U
Go
Main Menu
3.
From the
BOLUS MENU
Normal
BOLUS
ezCarb
scroll to "ezCarb."
ezBG
Combo Bolus
Reminders
Press
Main Menu
7.
BG CORRECT
Press
mg/dL
Actual
220
Target
-
120
= +
100
ISF
50 mg/dL
Show Result
Main Menu
®
* DO NOT enter Sensor glucose readings as BG values in pump bolus calculations. Always use a BG value from a BG meter. Using a Sensor glucose
reading in bolus calculations may lead to inaccurate suggested bolus amounts, and may result in under-delivery or over-delivery of insulin if you
choose to bolus that amount.
ATTENTION CAREGIVERS: It is recommended for patients 7 and under to use a tamper-resistant case that shields the buttons from accidental
button pushing. Always lock the pump when not in use to avoid inadvertent button presses. Pumps for younger children will automatically lock
when the pump enters into sleep mode.
page 1 of 2
4.
ezBG
mg/dL
Actual
220
Target
-
120
= +
100
ISF
50 mg/dL
Show Result
Main Menu
8.
to
4.
ezCarb Home
MENU,
Carbs:
34g
I:C
1U:
15g
Add BG
Show Result
Main Menu
8.
Bolus Total
Carb
2.26U
BG +
2.00U
IOB -
1.00U
Total =
3.25U
0.00U
Go
Type
Normal
Main Menu
/
Use
to enter
BG value.*
Press
This screen will be
displayed only if
using Dexcom G4
®
PLATINUM Sensor
and Transmitter
with pump.
/
Use
to enter the
total number of carbs to
be eaten. Press
/
Use
to
enter the suggested
bolus amount.
Press

Advertisement

loading

Summary of Contents for Animas Vibe

  • Page 1 Vibe The Animas Insulin Pump and CGM System ® ® Quick Reference Sheet for Basal and Bolus Options Deliver a Bolus using ezBG. From the Home From the MAIN MENU, From the BOLUS MENU, to enter 2 : 27 ezBG...
  • Page 2 Any broken sensors or adverse events should be reported to Customer Service. Contact Customer serious injury or death. Younger children may inadvertently press the pump buttons and deliver Service at 1-877-937-7867 or visit www.animas.com/importantsafetyinformation for detailed CONTRAINDICATIONS: Insulin pump therapy is not recommended for people unwilling or unable insulin, which can lead to hypoglycemic events.