DIGITAL OUTSIDE MICROMETER
Thank you for purchasing the Digital Outside Micrometer.
Please read this manual thoroughly before use for proper operation.
FEATURES
Easy to read digital display.
Carbide measuring surfaces.
Constant pressure ratchet mechanism.
Auto-on feature.
Auto-display-off function ( approx. 6 minutes )
Insulating cover
( Excluding: MCD130-125,
MCD130-150 )
Spherical Anvil Attachment ( SR2.5 ) included.
SR44 ( silver oxide battery ) included for testing.
PART IDENTIFICATION
0-Point
adjustment hole
[ ZERO ] Button
Battery Cover
TROUBLESHOOTING
Error Condition
LCD displays "inc" and
incorrect reading
LCD display reads "E 1"
LCD display reads "E 3"
Measured value is wrong
LCD does not display
Displayed value is unstable
Display is not clear
※If problem persists, or if you have any questions, please contact distributor or place of purchase.
※Please note, manufacturer is unable to respond to inquires or provide service directly.
 Please contact distributor or place of purchase.
Anvil ( Fixed side )
Measuring
Surfaces
( Carbide )
【Back】
MCD130-125
MCD130-150
※Does not include
  Insulating Cover
Insulating Plate
Press and hold [ ZERO ] Button
on rear for 2 seconds.
Measurement range exceeded.
Return spindle to measurement
range, press [ ZERO ] Button.
Measurement range exceeded.
Cycle power by removing and replacing
battery.
Clean measuring surfaces with cloth.
Confirm 0-Point setting and press
[ ZERO ] Button.
Cycle power by removing and
replacing battery.
Replace battery with new one.
SPECIFICATIONS
MCD130-25
MCD130-50
MCD130-75
MCD130-100
MCD130-125
MCD130-150
※Maximum Instrument Error does not include quantization error ( ±1count )
Spindle
Frame
【Front】
【ACCESSORIES】
5
0
25
30
3
45
Corrective Action
INSTRUCTION MANUAL
Model No. : MCD130- *** <combined>
Measurement
Resolution
Range (mm)
(mm)
0
25
0.001
-
25
50
0.001
-
50
75
0.001
-
75
100
0.001
-
100
125
0.001
-
125
150
0.001
-
Sleeve
Spindle Clamp
0
5
1
LCD Display
Insulating Cover
Spherical Attachment
Spanner
INSERTING THE BATTERY
①Rotate the battery cover on rear
using coin or similar tool.
②Insert battery with "+" side up.
③Replace battery cover.
CAUTION
Only use SR44 type Battery.
See Reverse for Operation→
Maximum
Weight
Instrument
(g)
Error (µm)
4
260
4
380
6
470
6
640
6
490
6
570
Thimble
5
0
45
Ratchet Stop
Calibration Standard
( Not included for MCD130-25 )
SR44
SR44
( Silver Oxide Battery )
※for test
Open
Battery
Cover

Advertisement

loading

Summary of Contents for Niigata seiki MCD130-25

  • Page 1 LCD Display [ ZERO ] Button Insulating Cover Battery Cover 【Back】 【Front】 【ACCESSORIES】 Spherical Attachment Calibration Standard ( Not included for MCD130-25 ) MCD130-125 MCD130-150 SR44 ※Does not include Spanner SR44   Insulating Cover ( Silver Oxide Battery ) ※for test...
  • Page 2 Keep instrument away from water and oil. ● Niigata seiki Co., Ltd. Please do not disassemble or modify. ● Remove dust and cutting chips after use and apply rust preventative oil to metallic 6-15-22, Tsukanome, Sanjo, Niigata, Japan, 955-0055 ●...

This manual is also suitable for:

Mcd130-50Mcd130-75Mcd130-100Mcd130-125Mcd130-150