Check the drum position 6
Set the weight rod to 6 on the scale, the longer part points to the patient.
Check position 5.9
The check point is at 5.9. Move the measuring drum's calibration mark 6 down‑
wards, opposite the index, by a ½ interval.
The feeler arm should move towards the examiner.
Check position 6.1
The check point is at 6.1. Move the measuring drum's calibration mark 6 upwards,
opposite the index, by a ½ interval.
The feeler arm should move towards the patient.
A. Appendix
NOTE!
Order numbers are written in italics.
An asterisk (*) shows that you should contact your HAAG‑STREIT rep‑
resentative for further information.
A.1 Accessories – Original HAAG-STREIT measuring prisms
The original HAAG‑STREIT measuring prisms are made of
PMMA and guarantee the best optical and mechanical qual‑
ity. Reusable measuring prisms must be cleaned and disin‑
fected after every examination. See chapter 2 in the separate
instructions for use "Cleaning and disinfection", HS item no.
7220375
A.2 Accessories – Tonosafe
Tonosafe disposable prisms are a comfortable and effective
solution to reduce the risk of cross‑infection in patients. To‑
nosafe is available in sterile packages with 5 supports and 100
disposable prisms.
HS item no. 7220345
© HAAG‑STREIT AG, 3098 Koeniz, Switzerland – HS‑Doc. no. 1500.7006000.04240 – 24. Edition / 2018 – 04
01-IFU_AT900-7006000-04240_eng.indd 17
SVENSKA
NEDERLANDS
PORTUGUÊS
ESPAÑOL
ITALIANO
5.9
B. Legal regulations
HAAG‑STREIT maintains a quality management system in accordance with EN
ISO 13485.
The AT 900 tonometer is a Class I device with measuring function in accordance
with Appendix IX of Directive 93/42/EEC. By affixing the CE mark we confirm that
our device complies with the applicable standards and directives.
You can request a copy of the declaration of conformity for the appliance from
HAAG‑STREIT at any time.
C. Classification
CE Directive 93/42/EEC
Class Im (measuring function)
FDA
Class II
D. Disposal
For correct disposal, please contact your HAAG-STREIT representative. This will
guarantee that no hazardous substances enter the environment and that valuable
raw materials are recycled.
E. Standards
EN ISO 8612
FRANÇAIS
DEUTSCH
ENGLISH
6.1
EN ISO 17664
03.05.2018 10:11:21