Omron HEM-4030 Instruction Manual

Omron HEM-4030 Instruction Manual

Manual inflation blood pressure monitor

Advertisement

Quick Links

Display:
K.
Heartbeat symbol
1. Flashes during measurement
Manual Inflation Blood
2. If flashing after measurement
Pressure Monitor
complete indicates blood
pressure out of recommended
range
Model HEM-4030
L.
Reinflation symbol
Instruction Manual
M.
Battery low symbol
N.
Memory symbol
5325569-2B
Introduction
2. Preparation
Thank you for purchasing the OMRON HEM-4030 Manual Inflation Blood
2.1 Installing/Replacing the Batteries
Pressure Monitor.
1.
The OMRON HEM-4030 is a compact manual inflation blood pressure monitor,
Remove the battery cover.
operating on the oscillometric principle. It measures your blood pressure and
pulse rate simply and quickly.
2.
Insert two "AAA" batteries
Please read this instruction manual thoroughly before using the unit.
as indicated in the battery
Please keep for future reference.
compartment and then
For specific information about your own blood pressure, CONSULT YOUR
replace the battery cover.
DOCTOR.
Notes:
Important Safety Information
• If the battery low symbol (
then replace both batteries at the same time.
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia
• The measurement values continue to be stored in memory even after
or arteriosclerosis.
the batteries are replaced.
Please read this section carefully before using the unit.
Disposal of used batteries should be carried out in accordance with the
Warning:
national regulations for the disposal of batteries.
• Indicates a potentially hazardous situation which, if not avoided, could result
in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and
3. Using the Unit
self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a
3.1 Applying the Arm Cuff
doctor before using the unit, as cuff inflation can cause internal bleeding.
(Battery Usage)
Remove tight-fitting clothing or tight rolled up sleeve from your
• If battery fluid should get in your eyes, immediately rinse with plenty of clean
upper arm.
water. Consult a doctor immediately.
Caution:
Do not place the cuff over thick clothes.
• Indicates a potentially hazardous situation which, if not avoided, may result in
minor or moderate injury to the user or patient or damage to the equipment or
1.
Insert the air tubes into the air jacks.
other property.
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express
their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result
in incorrect measurement results.
• Make sure that the air tube is not wrapped around other parts of your body
when taking measurements at night. This could result in injury when the air
pressure in the air tube is increased.
• Do not leave the cuff wrapped on the arm if taking measurements during the
2.
Put your arm through the cuff loop.
night. This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields,
near the unit. This may result in incorrect operation of the unit.
• Do not operate the unit in a moving vehicle (car, airplane).
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with
plenty of clean water.
• Use only two "AAA" alkaline or manganese batteries with this unit. Do not use
other types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
3.
• Replace old batteries with new ones immediately. Replace both batteries at
Position the arm correctly.
the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
• Do not take measurements after bathing, drinking alcohol, smoking,
exercising or eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull the air tube with holding near the air jack.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the "Important information regarding Electro Magnetic
The bottom edge of the cuff should be 1 to 2 cm above the elbow.
Compatibility (EMC)" in the Technical Data Section.
Marker (arrow under tube) is centred on the middle of your inner arm.
• Read and follow the "Correct Disposal of This Product" in the Technical Data
Close the fabric fastener FIRMLY.
Section when disposing of the device and any used accessories or optional
parts.
1. Overview
Main unit:
A
D
Notes:
• When you take a measurement on the right
B
arm, air tube will be at the side of your
E
elbow. Be careful not to rest your arm on
C
the air tube.
• The blood pressure can differ between the right arm and the left arm,
A. Display
C. ON/OFF (Power switch) button
and therefore also the measured blood pressure values can be
B. M (Memory) button
D. Air jacks for cuff and inflation bulb
different. Omron recommends to always use the same arm for
E. Battery compartment
measurement. If the values between the two arms differ substantially,
Arm cuff:
please check with your doctor which arm to use for your
measurement.
F. Arm cuff
3.2 How to Sit Correctly
(Medium cuff: arm
circumference 22-32 cm)
To take a measurement, you need to be
G. Air tube
relaxed and comfortably seated, under
F
G
comfortable room temperature. No eating,
smoking or exercising 30 minutes before
Inflation bulb:
Storage case:
taking a measurement.
H
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight.
• The cuff should be at the same level as your
I
heart.
H. Air release button
J
I. Air tube
J. Air inflation bulb
3.3 Taking a Reading
O
K
Note: Remain still while taking measurement.
P
L
1.
Press the air release button to release any air in the arm
Q
M
cuff, then press the ON/OFF button to turn the unit on.
R
N
S
O.
Systolic blood pressure
P.
Diastolic blood pressure
Q.
Pulse display
R. Memory No. symbol
S. Deflation symbol
Note:
If the deflation symbol does not disappear soon, press the air release
button to release any air in the arm cuff.
2.
Pump the inflation bulb to inflate the arm cuff.
1) Inflate the cuff until it is 30 to 40 mmHg
above your expected systolic blood
pressure value.
ex) If your expected blood pressure is
around 140mmHg, inflate the arm cuff to
between 170 and 180 mmHg. Inflate the
cuff rapidly so that the pressure is reached in about five seconds.
2) When the desired pressure has been achieved, release the inflation
bulb.
Notes:
• If the reinflation symbol (
) appears, squeeze the inflation bulb to
re-inflate the arm cuff.
• Do not inflate the arm cuff more than necessary.
) appears on the display, turn the unit off
3.
Measurement starts.
Measurement starts automatically after you stop inflating
the arm cuff.
START
DEFLATE
Press the air release button
to release the air in the arm
cuff until the deflation symbol
is no longer displayed.
Notes:
• To cancel a measurement, press the ON/OFF button to turn off the
unit and press the air release button to release the air in the arm cuff.
• Wait 2-3 minutes before taking another blood pressure measurement.
Waiting between readings allows the arteries to return to the condition
prior to taking the blood pressure measurement.
Warning:
Self-diagnosis of measured results and treatment are dangerous.
Please follow the instructions of your doctor.
4.
Remove the arm cuff.
5.
Press the ON/OFF button to turn off the monitor.
The monitor automatically stores the measurement in its
memory.
It will automatically turn off after five minutes.
Important:
• If your systolic or diastolic pressure is outside
the standard range, the heartbeat symbol will
blink when the measurement result is
displayed.
Recent research suggests that the following
values can be used as a guide to high blood
pressure for measurements taken at home.
Systolic Blood Pressure
Diastolic Blood Pressure
This criteria is for home blood pressure measurement.
1 to 2 cm
3.4 Using the Memory Function
The monitor automatically stores the result up to 21 sets.
Note: If the memory is full, the monitor will delete the oldest readings.
Previous Reading
Turn the Unit on
View Previous
Reading
• The Memory number
appears for a second
before the pulse rate
is displayed. The
newest set is
numbered "1".
To delete all the values stored in memory
When the memory symbol (
) appears, first press the M button.
Then while holding it down, press the ON/OFF button simultane-
ously for about 2-3 seconds.
Note: You cannot partially delete the stored readings.
4. Troubleshooting and Maintenance
4.1 The Icons and Error Messages
Error Display
e
s
u
a
C
Recent research suggests
The systolic blood pressure
these values can be used
is above 135 mmHg or
as a guide to high blood
diastolic blood pressure is
pressure for measurements
above 85 mmHg.
taken at home.
You should replace them
with new ones ahead of
Blink
The batteries are low.
time.
Refer to section 2.1.
You should replace them
The batteries are
Lit
with new ones at once.
exhausted.
Refer to section 2.1.
Cuff is under inflated.
Movement during
Carefully read and repeat
measurement.
the steps listed under
section 3.3.
Cuff over inflated.
Contact your OMRON retail
Device error.
outlet or distributor.
Refer to Chapter 6.
Press the inflation bulb to
inflate the arm cuff until the
reinflation symbol goes out.
END
COMPLETE
Or, deflate the arm cuff and
Pressure is too low.
repeat measurement after
checking that the heartbeat
symbol (
displayed.
Refer to section 3.3.
4.2 Troubleshooting
Problem
Cause
Apply the arm cuff
Arm cuff not applied
correctly.
correctly.
Refer to section 3.1.
Remain still and do not
The reading is
Movement or talking
talk during
extremely low (or
measurement.
during measurement.
high).
Refer to section 3.3.
Remove any clothing
interfering with the arm
Clothing is interfering
with the arm cuff.
cuff.
Refer to section 3.2.
Make sure that the air
The air tube is not
tube is connected
securely connected
securely.
Arm cuff pressure
into the main unit.
Refer to section 3.1.
does not rise.
Replace the arm cuff
Air is leaking from the
with a new one.
arm cuff.
Refer to Chapter 5.
The arm cuff has not
Inflate the cuff so that it
is 30 to 40 mmHg above
been inflated
sufficiently.
your previous
Cannot measure or
measurement result.
readings are too low
Refer to section 3.3.
or too high.
The air release button
Be careful not to press
being pressed during
the air release button
inflation.
during measurement.
Replace the batteries
The unit loses power
The batteries are
with new ones.
during measurement.
empty.
Above 135 mmHg
Refer to section 2.1.
Above 85 mmHg
Replace the batteries
The batteries are
with new ones.
empty.
Nothing happens
Refer to section 2.1.
when you press the
The batteries have
Insert the batteries with
buttons.
been inserted
the correct (+/-) polarity.
incorrectly.
Refer to section 2.1.
• Press the ON/OFF button and repeat measurement.
• If the problem continues, try replacing the
batteries with new ones.
Other problems.
If this still does not solve the problem, contact
your OMRON retail outlet or distributor.
View Readings
Stored in Memory
• Press repeatedly to
cycle through the
previous results.
• Hold down to cycle
rapidly.
4.3 Maintenance
To protect your unit from damage, please observe the following:
• Do not subject the main unit, cuff and inflation bulb to
y
d
e
m
e
R
extreme temperatures, humidity, moisture or direct
sunlight.
• Do not fold the cuff or tubing tightly.
• Do not inflate the arm cuff over 299 mmHg.
• Do not disassemble the unit.
• Do not subject the unit to strong shocks or vibrations
(for example, dropping the unit on the floor).
• Do not use volatile liquids to clean the main unit. The
unit should be cleaned with a soft, dry cloth.
• Use a soft, moistened cloth and soap to clean the arm
cuff.
• Do not wash the arm cuff or immerse it in water.
• Do not use petrol, thinners or similar solvents to clean
the arm cuff.
• Do not carry out repairs of any kind yourself. If a defect
occurs, consult your local authorised OMRON
distributor or dealer.
Calibration and Service
) has been
• The accuracy of this blood pressure monitor has been
carefully tested and is designed for a long service life.
• It is generally recommended to have the unit inspected
every two years to ensure correct functioning and
accuracy. Please consult your local authorised
Remedy
OMRON distributor or dealer.
4.4 Storage
Keep the unit in its storage case when not in use.
Do not keep the main unit being wound with the arm cuff.
1.
Unplug the air tube from the air connector.
2.
Gently fold the air tube into the arm cuff.
Note:
Do not bend the air tube excessively.
3.
Place the arm cuff, inflation bulb and
main unit in the storage case.
Do not store the unit in the following situations:
• If the unit is wet.
• Locations exposed to extreme temperatures, humidity, direct
sunlight, dust or corrosive vapours.
• Locations exposed to vibrations, shocks or where it will be at an
angle.
5. Optional Parts
Medium Arm Cuff
Large Arm Cuff
Small Arm Cuff
Arm circumference
Arm circumference
Arm circumference
22 - 32 cm
32 - 42 cm
17 - 22 cm
Small cuff and bulb combination
Regular bulb
Arm circumference 17 - 22 cm
Note:
Should you require a small cuff, please ensure that the small cuff and
small bulb are used together.
They may also be purchased as a combination set.
6. Technical Data
Product Description
Manual Inflation Blood Pressure Monitor
Model
HEM-4030
Display
LCD Digital Display
Measurement Method
Oscillometric method
Measurement Range
Pressure: 0 mmHg to 299 mmHg
Pulse: 40 to 180/min.
Accuracy
Pressure: ±3 mmHg
Pulse: ±5% of display reading
Inflation
Manual by inflation bulb
Deflation
Automatic pressure release valve
Memory
21 Measurements
Power Source
2 "AAA" batteries 1.5V
Battery Life
Capacity of new manganese batteries is approx.
1500 measurements
Applied Part
= Type B
Protection Against
Internally powered ME equipment
Electric Shock
Operating
+10°C to +40°C / Maximum: 30 to 90% RH
temperature/ Humidity
Storage temperature/
-20°C to +60°C / Maximum: 10 to 95% RH /
Humidity/ Air pressure
700-1060 hPa
Console Weight
Approximately 75g without batteries
Cuff Weight
Approximately 120g
Outer Dimensions
Approximately 64 (w) mm x 35 (h) mm x 105 (I) mm
Cuff Dimensions
Approximately 146 mm x 446 mm
(Medium cuff: arm circumference 22 to 32 cm)
Cuff Material
Nylon and polyester
Package Content
Main unit, cuff, inflation bulb, instruction manual,
storage case, battery set
Notes:
• Subject to technical modification without prior notice.
• Disposal of this product and used batteries should be carried out in accordance
with the national regulations for the disposal of electronic products.
Please read this instruction manual carefully before using the device.
• This device fulfils the provisions of EC directive 93/42/EEC (Medical Device
Directive).
• This blood pressure monitor is designed according to the European Standard
EN1060, Noninvasive sphygmomanometers Part 1: General Requirements
and Part 3: Supplementary requirements for electromechanical blood
pressure measuring systems.
Important information regarding Electro Magnetic Compatibility (EMC)
With the increased number of electronic devices such as PC's and mobile
(cellular) telephones, medical devices in use may be susceptible to
electromagnetic interference from other devices.
Electromagnetic interference may result in incorrect operation of the medical
device and create a potentially unsafe situation.
Medical devices should also not interfere with other devices.
In order to regulate the requirements for EMC (Electro Magnetic Compatibility)
with the aim to prevent unsafe product situations, the EN60601-1-2:2007
standard has been implemented. This standard defines the levels of immunity to
electromagnetic interferences as well as maximum levels of electromagnetic
emissions for medical devices.
This medical device manufactured by OMRON HEALTHCARE conforms to this
EN60601-1-2:2007 standard for both immunity and emissions.
Nevertheless, special precautions need to be observed:
• Do not use mobile (cellular) telephones and other devices, which generate
strong electrical or electromagnetic fields, near the medical device. This may
result in incorrect operation of the unit and create a potentially unsafe situation.
Recommendation is to keep a minimum distance of 7 m. Verify correct
operation of the device in case the distance is shorter.
Further documentation in accordance with EN60601-1-2:2007 is available at OMRON
HEALTHCARE EUROPE at the address mentioned in this instruction manual.
Documentation is also available at www.omron-healthcare.com.
Correct Disposal of This Product
(Waste Electrical & Electronic Equipment)
This marking shown on the product or its literature, indicates that it
should not be disposed of, with other household wastes at the end
of its working life. To prevent possible harm to the environment or
human health from uncontrolled waste disposal, please separate
this from other types of wastes and recycle it responsibly to
promote the sustainable reuse of material resources.
Household users should contact either the retailer where they purchased this
product, or their local government office, for details of where and how they can
take this item for environmentally safe recycling.
Business users should contact their supplier and check the terms and conditions
of the purchase contract. This product should not be mixed with other commercial
wastes for disposal.
This product does not contain any hazardous substances.
OMRON HEALTHCARE Co., Ltd.
24, Yamanouchi Yamanoshita-cho, Ukyo-ku, Kyoto
615-0084 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Kruisweg 577, 2132 NA Hoofddorp THE NETHERLANDS
OMRON HEALTHCARE SINGAPORE PTE LTD.
Asia Pacific HQ
438A Alexandra Road, #05-05/08, Alexandra Technopark,
Singapore 119967