Model DFG-RS3 Digital Force/Torque Indicator
CALIBRATION
COMPLETE
Save & exit
Exit w/o saving
To save the calibration information, select "Save & exit". To exit without saving the data select
"Exit without saving".
12. Any errors are reported by the following screens:
CALIBRATION
Units must be gF.
Please try again
Press ENTER.
Displayed at the start of calibration if a disallowed unit is selected.
CALIBRATION
Load not stable.
Please try again.
Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again.
CALIBRATION
COMPRESSION
Load too low.
Please try again.
Causes:
1. The calibration weight does not match the set value.
2. If using a DFG-RSA configurable adapter, ensure that the output signal leads (SG+ and SG-)
have been installed into the appropriate terminal blocks. Some sensor manufacturers
consider SG+ to be a compression value, while others consider it to be a tension value. If the
indicator is expecting a compression load but it is receiving a tension signal, calibration
cannot continue. Verify that the tension/compression indicator on the home screen properly
corresponds to the load direction, and switch the signal leads, if required.