INSTRUMENT OVERVIEW Display Push Buttons Battery Charge Port Transducer port Measuring Tip Transducer Locking Extension Bar 10 mm Ring Extension Bar 20 mm Extension Bar 40 mm Extension Bar 80 mm Fine Adjustment Extension Bar 160 mm Thread SPARE PART LIST ITEM NO DESCRIPTION ITEM NO DESCRIPTION 412-2005 Transducer Standard, min 89 mm...
12. Repeat the procedure from step 6. FACTS ABOUT PRISMA DI-5 TOLERANCE • Prisma DI-5 has a tolerance of max. 1% in the range of +/- 1,000 mm and max 2% in the range of 1,000-2,000 mm. HOW TO ACHIEVE OPTIMUM RESULTS • An accurate measurement result can be achieved if both transducer and the measured object (Crankshaft) are having the same temperature so that the transducer temperature is not changed during the measurement.
Page 4
ITEM NO 488-8100 Cylinder liner maintenance. The Ovality Kit is an accessory to the Prisma DI-5C and Prisma DI-5. The method is simply giving 5 measuring points at each level of the liner. To compare the levels you will also see how much the wear of the liner is in the cylinder top.