Tsuruga 4021 Instruction Manual

Digital large display meter

Advertisement

Quick Links

1. Preface
Please take care that this instruction manual is certainly delivered to the person in charge of operating this instrument.
Unpack the product and confirm that the following items are included.
(1) 4021 itself (2) Instruction manual (3)CJS (for thermocouple only. It is installed in the terminals)
For safety and proper use of this product, please observe the following caution and also read the instruction manuals to follow before
the initial operation.
● The 4021 becomes active as soon as it is connected to a power source.
It has no On or Off switch. But warm-up time is necessary at least 15 minutes.
● Do not install under the following conditions.
-Where it is exposed to direct sunlight, dust, corrosive gases, rain, etc.
-Where ambient temperature or humidity is high.
-Where it is exposed to excessive noise or static electricity.
-Where there is constant vibration or shock
2. Standard Specifications
2.1 Model Name
4021 – □ – □ – □ – □ – □
( 1 )
( 2 ) ( 3 ) ( 4 ) ( 5 )
(1) Measuring Input
[ Thermocouple input ]
Model
Sensor
4021-02
K
4021-04
J
4021-05
T
< NOTE > Accuracy: Defined at 23
Temperature coefficient: ±300ppm/
Compensation of standard contact: ±1
Calibration: In conjunction with JIS C 1602-1995, by each standard thermoelectric power mV input.
[ RTD input ]
Model
Sensor
4021-11
Pt100Ω
4021-12
Pt100Ω
< NOTE > Accuracy: Defined at 23
Temperature coefficient: ±200ppm/
Calibration: In conjunction with JIS C 1604-1997, by the resistive value of standard resistive elements.
(2) Power Supply
No.
Power
3
AC100 ∼ 120V
5
AC200 ∼ 240V
9
DC 24V
(4) Color
No
Type
(Null)
Red LED
G
Green LED
Instruction Manual
Digital Large Display Meter, Model 4021
CAUTION
Measuring range
o
-200 ∼ 999
C
o
-200 ∼ 999
C
o
-200 ∼ 400
C
o
o
C±5
C, 45 to 75%RH.
o
C within the 0 to 50
o
C within the 0 to 50
Measuring range
o
-200 ∼ 850
C
o
-99.9 ∼ 99.9
C
o
o
C±5
C, 45 to 75%RH.
o
C within the 0 to 50
(3) Mounting
No.
51
52
53
61
62
63
(5) Option
No.
(Null)
A01
A02
Display range
o
-270 ∼ 999
C
±(0.3% of rdg +1
o
-210 ∼ 999
C
±(0.3% of rdg +1
o
-270 ∼ 420
C
±(0.3% of rdg +1
o
C temperature range.
o
C temperature range.
Display range
o
-200 ∼ 870
C
o
-99.9 ∼ 99.9
C
±(0.2% of rdg +0.3
o
C temperature range.
Type
Wall-mount
Hanging-mount
Sticking-mount
Wall-mount
Integral Pt sensor
Hanging-mount
(for RTD input only)
Sticking-mount
Function
Peak Hold
Bottom Hold
Change of the burnout direction <NOTE>
<NOTE>: Thermocouple input only.
- 1 -
Accuracy <Note>
o
C)
o
C)
o
C)
Accuracy <Note>
o
±(0.4% of rdg +1
C)
o
C)
I-01950
Overload
DC±10 V
DC±10 V
DC±10 V
Overload
DC±10 V
DC±10 V

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the 4021 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Tsuruga 4021

  • Page 1 Unpack the product and confirm that the following items are included. ● (1) 4021 itself (2) Instruction manual (3)CJS (for thermocouple only. It is installed in the terminals) For safety and proper use of this product, please observe the following caution and also read the instruction manuals to follow before the initial operation.
  • Page 2: Optional Specifications

    Model 4021 2.2 General Specifications Display : Red or Green LED (character height 56mm) with zero-suppress function. Over-range indication : Flashes at minimum or maximum value of the display range when exceeding. Burnout : Blinking with maximum value of display range.
  • Page 3 Model 4021 2.4 Mounting Wall-mount (model code-51) ● NOTE: Cut the case bottom or side to pull out lead wires. W al l Hanging-mount (model code-52) ● NOTE: Fixing points should be locating 10mm away from the wall. Mount to the top roof Sticking-mount (model code-53) ●...
  • Page 4 Model 4021 3. Setting of Each Function ● Name of each part ● Display Lit-up display Polarity LED PH LED (BH LED) Blinking display ℃ Turned-off display Unit Setting switch※ ● Function of each switch ※ F M Functions switch : Change between the measuring mode and the setting mode.
  • Page 5: Terminal Arrangement

    Model 4021 4. Wiring CAUTION  ● Do not miswiring. Otherwise, the meter is broken. ● To avoid an electrical shock, turn the power off when wiring. ● Do not wire with moistened hands. Locate away from the wet place.
  • Page 6: Maintenance

    Model 4021 5. Maintenance Store the meter within the rated storage temperature (from -20 to +70 C). When the front panel or the case is cleaned, use soft cloth dipped with cleaner liquid. Do not use organic solvent like benzene or paint thinner as they may deform or discolor the case.
  • Page 7 Model 4021 6.2 Calibration of RTD input Prepare the standard variable resistor. To enter the calibration mode, turn the power on with pushing M switch. Connect the standard variable resistor as shown in the Figure-3. Power OFF Turn the power on with pushing the M switch.
  • Page 8: Basic Type

    Model 4021 7. Dimensions 7.1 Basic type ● Wall-mount (-51) ℃ Split-able Provided at both side and bottom R2.25 2-φ4.5 ● Hanging-mount (-52) 81211 4-φ4.5 ℃ Split-able Provided at both side and bottom 10max ● Sticking-mount (-53) ℃ R2.25 2-φ4.5...
  • Page 9: Contact Information

    4-φ4.5 ● Hanging-mount (-62) ℃ Split-able Provided at both side side and bottom 10max ● Sticking-mount (-63) ℃ R2.25 2-φ4.5 Unit: mm Contact Information Name : Tsuruga Electric Corporation Address : 1-3-23 Minami-Sumiyoshi, Sumiyoshi-ku, Osaka-shi 558-0041 Japan - 9 -...
  • Page 10: Standard Specifications

    Instruction Manual Digital Large Display Meter, Model 4022 I-01951 1. Preface ● Please take care that this instruction manual is certainly delivered to the person in charge of operating this instrument. ● Unpack the product and confirm that the following items are included. (1) 4022 itself (2) Instruction manual For safety and proper use of this product, please observe the following caution and also read the instruction manuals to follow before the initial operation.
  • Page 11: General Specifications

    Model 4022 2.2 General Specifications Display : 0 to 999, Red or Green LED (character height 56mm) with zero-suppress function. Scaling function : Full scale display -999 to +999, programmable. Offset display -999 to +999, programmable. Hold function: : Measured data is held (Not isolated from input). Peak hold: : Maximum value of the measured data is held (Not isolated from input).
  • Page 12 Model 4022 2.4 Mounting Wall-mount (model code-51) ● NOTE: Cut the case bottom or side to pull out lead wires. W al l Hanging-mount (model code-52) ● NOTE: Fixing points should be locating 10mm away from the wall. Mount to the ceiling. Sticking-mount (model code-53) ●...
  • Page 13: Shift Switch

    Model 4022 3. Setting of Each Function ● Name of each part ● Display Lit-up display Polarity LED PH LED (BH LED) Blinking display ℃ Turned-off display Unit Setting switch※ ● Function of each switch ※ F M Functions switch : Change between the measuring mode and the setting mode.
  • Page 14: Setting Method

    Model 4022 3.1 Setting method 3.1.1 Display scaling - Offset setting Offset display can be set to an arbitrary value. Example: With the rated input 1to 5V, the display at the input 1V is adjusted from 0 to 100. Measuring mode Push switch for 3seconds.
  • Page 15 Model 4022 3.1.4 Average operation Average operates the interval average or the moving average. Example: Changes the non-average operation to the moving average of 8-data. Setting mode Display shows setting condition > Push the switch to move to the setting mode. (blinking ) Push the Λ...
  • Page 16: Adjustment Function

    Model 4022 3.2 Adjustment function 3.2.1 Zero adjustment Fine calibration of the offset display is available to meet the real input value. Example: your meter is 1-5V input, and when inputting 1V, displayed value is 5, not 0. Measuring mode Push switch for 6seconds.
  • Page 17 Model 4022 4. Wiring CAUTION ● Do not miswiring. Otherwise, the meter is broken. ● To avoid an electrical shock, turn the power off when wiring. ● Do not wire with moistened hands. Locate away from the wet place. ● Do not touch terminals when turning the power on. 4.1 Terminal Arrangement INHi INLo...
  • Page 18 Model 4022 Common (COM) ● “COM” terminal is used for Hold and Peak (or Bottom) Hold, Zero set, and Decimal point terminals. NOTE: Those terminals, COM, DP1, DP2, HOLD, ZS, PH, are not isolated to the input. Use a photo-coupler or switch to insulate. It is essential when using the input floating.
  • Page 19 ● Hanging-mount (-52) 81211 4-φ4.5 t o n Split-able Provided at both side and bottom 10max ● Sticking-mount (-53) V R2.25 2-φ4.5 Unit: mm Contact Information Name : Tsuruga Electric Corporation Address : 1-3-23 Minami-Sumiyoshi, Sumiyoshi-ku, Osaka-shi - 10 - 558-0041 Japan...