B+B Sensors Flashbeam 8869 Operation Manual page 2

Advertisement

OPERATION MANUAL
Infrared thermometer
Flashbeam 8869
Measurement conditions
Holding the meter by its handle, point the IR Sensor toward the object
whose temperature is to be measured.
The meter automatically compensates for temperature deviations
from ambient temperature. Keep in mind that it will take up to 30
minutes to adjust to wide ambient temperature changes.
When low temperatures are to be measured followed by high tempe-
rature measurements, some time (several minutes) is required after
the low (and before the high) temperature measurements are made.
This is a result of the cooling process which must take place for the
IR sensor.
Battery replacement
A flashing display is the indication that the battery voltage has
fallen into the critical region (6.5 to 7.5 V). Reliable readings can be
obtained for several hours after the first appearance of the low battery
indication. Open the Battery Compartment and remove the batte-
ry,then install a new battery and replace the cover.
Optical resolution
Technical changes reserved
0141 0315-20
02.12.2015
B+B Thermo-Technik GmbH | Heinrich-Hertz-Straße 4 | D-78166 Donaueschingen
Fon +49 771 83160 | Fax +49 771 8316-50 | info@bb-sensors.com | bb-sensors.com
Technical data
Features
Values
Measuring range
-20...+420 °C
Operating conditions
0...50 °C, max. 80% RH
Accuracy
±2 °C (-20...+100 °C),
±3% or ±3 °C (+1000...+420 °C)
Resolution
1 °C
Response time
1 s
Emissivity
0,95 fixed
Optical resolution
8:1
Battery
9 V
Dimensions
170x44x40 mm, 200 g
CE-conformance
2004/108/EG
Electromagnetic
EN 61326-1:2013
conductivity
Articlenumber
0560 8869
2 / 2