718 Static Sensor
Instruction Manual
Safety Information
Intended Use
The 3M™ 718 Static Sensor is a portable,
handheld instrument designed for measuring
voltages associated with electrostatic charge.
Its intended use is for measuring the amount
of voltage, in a range from 0-20 kilovolts,
associated with an electrostatic charge
buildup on a surface. Any deviation from
this intended use could impair the
instrument's effectiveness and possibly lead
to an unsafe operating condition.
The 3M™ 718A Air Ionizer Test Kit is a
set of accessories for use in conjunction
with the 3M 718 Static Sensor. Their
intended use is for verifying the operation of
air ionizers by measuring the neutralization
time for a static charge placed on a
stationary metal plate. Any deviation
from this intended use could impair the
instrument's effectiveness and possibly lead
to an unsafe operating condition.
Caution Statements
• The 718 Static Sensor and 718A Charger
use DC 9V power supplied by a 9V
alkaline battery. Usage of any other
power source may cause damage to the
instruments.
• The 718 Static Sensor and 718A Air
Ionizer Test Kit have no user-serviceable
parts. Do not disassemble the products for
any reason. UNAUTHORIZED SERVICE WILL VOID THE WARRANTY.
• The 718 and 718A are NOT designed for usage in hazardous environments where the
possibility of explosion or fire exists.
Read and understand all safety information before installing and operating
this equipment.

Advertisement

loading
Need help?

Need help?

Do you have a question about the 718 and is the answer not in the manual?

Questions and answers

Summary of Contents for 3M 718

  • Page 1 Do not disassemble the products for any reason. UNAUTHORIZED SERVICE WILL VOID THE WARRANTY. • The 718 and 718A are NOT designed for usage in hazardous environments where the possibility of explosion or fire exists. Read and understand all safety information before installing and operating...
  • Page 3 1.0 Description 2.2.3 Connect the two, and place the connected battery into the housing The 3M™ 718 Static Sensor is a portable by inserting the connector end first, then following up with the other end handheld instrument used for locating and of the battery.
  • Page 4 3.3 Measurement Range: all measurements are in kilovolts (kV) as stated on the front label of the unit. The 718 has two measurement ranges: 0-2 kV and 0-20 kV. The unit’s current measurement range mode can be verified by checking the display.
  • Page 5 HOLD switch once again. Please note that during HOLD condition the distance- indicating LED’s are turned off. 3.5 Zero adjustment: the 3M™ 718 Static Sensor has a zero adjustment function, which sets a zero reference point for all subsequent measurements.
  • Page 6 ESD DSP3.3 if the voltages. Please use a 3/32 inch (2.5 3M™ 718A Air Ionizer Test Kit is used to mm) mono-phone plug to connect perform verification testing on ionizer into the output jack. The output is performance.
  • Page 7 5.0 Service/Calibration 5.1 Service and Repair: in the event that you believe the 718 Static Sensor or 4.3 Testing ionizer discharge time: after the 3M™ 718A Air Ionizer Test Kit charging the plate, hold the 718 is in need of repair, please contact approximately one foot (30.5 cm)
  • Page 8 718 is now calibrated. 5.5 Procedure: 5.5.7 Repeat procedures 5.5.2 - 5.5.6 for 5.5.1 Place the 3M™ 718 Static Sensor on the 20 kV range of the meter, using the metal stand. Verify that it is a test voltage of 5,000 volts.
  • Page 9 5% of measured voltage Measurement Stability ± 10 counts Certifications UL, C-UL, CE, CB-scheme, NOM 3M™ 718A Air Ionizer Test Kit Charge Plate Assembly Per ESD Association Draft Standard D3.3 Aluminum bracket Bare stainless steel plate Teflon spacers isolate plate from bracket Charge Plate Bracket Dimensions 1"...
  • Page 10 7.0 Parts Listing 8.0 Contact Information: Customer and Technical Service 3M™ 718 Static Sensor 1 ea. 718 Static Sensor Within the U.S. : Customer service and 1 ea. 718/718A Operator’s Manual technical support can be obtained by 1 ea. Certificate of Conformance calling the 3M Electronic Handling &...
  • Page 12 Before product. Except where prohibited by using this product, you must evaluate it and law, 3M will not be liable for any indirect, determine if it is suitable for your intended special, incidental or consequential loss application.You assume all risks and...