Caribratlion Procedure (Performance Check)-'DM 505
6. Check
Ohms
Accuracy
a.
Connect the resistance standard to the VOLTS/Rl
and LOW
connectors.
b. Press the kf2 pushbutton. Make each measurement
in
thefollowing
tablewith the HI
!2-LO R
button in the
LO
(in) and
the
HI (out) positions.
c. CHECK-that the display reads within the limits for
the ambient tern perature as listed in Table 4 4 .
d.
Remove all
connections
to
the DM 505.
7. Check PC Current Accuracy
a.
Connect the current calibrator
output
through a
coaxial cable and necessary adapters t o t he mA and LOW
connectors. Connect the
d c voltage
caPibrator output
through
a coaxial
cable and adapters to the current
calibrator input.
b. Press the rnA DC button and set the DM 505 range
buttons and current source output as listed
in
Table 4-5.
c. CHECK-that the display reads within the limitsfor
the ambient temperature as listed in Table
4-5.
d.
Set the dc current
source to
0
A and disconnect it
from the DM
505.
Table 4-4
OHMS ACCURACY
I
I
Table 4-5
Range Button
200
n
DC
CURRENT ACCURACY
-
-
Standard
180.000
n
DM
505
Range Button
200
p A
2
mA
20
mA
200 m
A
2000
mA
.
.
%lB°C
t o
+2a°C
178.8
to
181.2
Dc Current
Source
180.000
pd
1.80000
mA
18.0000
mA
180.000 mA
1800.00
mA
0 to S18"C, f 2 8 " C ta +50°C
177.9
to
182.1
Display Umlts
4-18' C to f 28" C
179.4 to 180.6
1.794 to 1.806
17.94 to 18.06
179.4 to 180.6
1794 to 1806
0
to +18OC, +2BDC
to +50°C
179.0
to 181.0
1.790
to
1.810
17.90 to 18.10
179.6 to 181.0
1700
to 1810