6.
Locate the fuse slot shown in Figure 29.
1
Fuse
7.
Pry the fuse loose from the slot.
8.
Verify the fuse is blown by measuring discontinuity across the fuse using another meter.
9.
Insert a new fuse into the fuse slot as shown in Figure 30.
1
Fuse
10. Gently press the fuse with the screwdriver until both fuse clamps snap the fuse into place.
© National Instruments Corporation
Figure 29. Removing the Fuse from the NI PXI-407x
Figure 30. Replacing the Blown Fuse in the NI PXI-407x
33
1
2
2
Fuse Slot
1
2
2
Fuse Slot
NI Digital Multimeters Getting Started Guide