Troubleshooting And Error Modes - Revolutionary Science Saniclave 200 Instruction Manual

Table of Contents

Advertisement

Troubleshooting
Error Conditions and Display Readings
Note: If a failure occurs, the load must be considered non-sterile. The sterilizer and
load should then be investigated. Contact Revolutionary Science if problem persists.
1. Manual Cycle Abort:
2. Preheat too long (Door Ajar):
3. Over Temperature Alarm:
4. Under Temperature Alarm:
5. Low Water:
If 'START' is pressed, display will read 'Err 1' for
cycle abort. User may abort the cycle anytime,
during, preheat, sterilize or dry modes. If cycle is
aborted, the contents of the load shall be
considered non-sterile. Press SET to acknowledge
the error.
If door is not closed properly and the cycle is
initiated, the cycle will fail when the autoclave
cannot pressurize. Display will read 'Err 2.' Press
SET to acknowledge the error. Take off the door.
Remove door gasket. Inspect the door gasket for
cracks. Replace with a new gasket if necessary.
Reinstall gasket into door. Properly close door by
lining up arrows on door and on body and twisting
the door closed. Restart the cycle.
If temperature exceeds the maximum of 124°C the
display will read 'Err 3.' Allow unit to drain the
reservoir fully into the exhaust tank or sink. Press
SET to acknowledge the error. Open door when
chamber has cooled. Remove load from chamber.
Refill reservoir. Reinsert load. Close the door.
Restart cycle. If an over-temperature condition
occurs again, contact Revolutionary Science to
investigate the problem further. Discontinue use
until problem is resolved.
If the temperature or pressure of the
chamber falls below sterilization temperature of
121°C during the sterilization cycle, the digital
display will read 'Err 4.' Press SET to acknowledge
the error. Open door when chamber has cooled.
Remove load from chamber. Refill reservoir.
Reinsert load. Close the door. Restart cycle. If an
under-temperature condition occurs again, contact
Revolutionary Science to investigate the problem
further. Discontinue use until problem is resolved.
The display will read 'Err 5' if there is an insufficient
amount of water in the chamber. Press SET to
acknowledge the error. Fill reservoir up to the
temperature sensing probe. Restart cycle.