Table of Contents
loading

Summary of Contents for Revolutionary Science Saniclave 200

  • Page 1 The Saniclave 200 Instruction Manual RS-SC-200...
  • Page 2: Table Of Contents

    Read this manual before use! This manual contains important information, safeguards and operating instructions for the operation of the REVOLUTIONARY SCIENCE model RS- SC-200 Saniclave 200 by Alternative Pioneering Research & Development, Inc. (APR&D, Inc.) © Alternative Pioneering Research and Development, Inc.
  • Page 3: Introduction

    Introduction Dear Valued Customer, Thank you for purchasing the Revolutionary Science RS-SC-200 Saniclave 200. We are confident that when used correctly, the Saniclave 200 will provide years of reliable service. The high quality of construction and durable parts deliver a performance equivalent to much costlier equipment.
  • Page 5: Installation, Instructions And Intended Use

    Installation Instructions Remove Saniclave 200 from packaging and place it on a dry level surface. Plug it into an accessible 120 VAC power receptacle. For printer setup, see page 6. The Saniclave 200 is now ready for use. See operating instructions on page 7.
  • Page 6: Safeguards And Cautions

    6. NEVER use accessory attachments not recommended by Revolutionary Science. Their use may cause fire, electrical shock or injury. 7. NEVER place the Saniclave 200 or any part of the Saniclave 200 on or near a hot gas or electric burner or other heating source.
  • Page 7: Operating The Saniclave 200

    The printer must be powered and connected to the Saniclave 200 via a serial cable prior to use. Sufficient paper must be installed in the printer. At the end of the cycle (whether it is a complete cycle or a failure), the SET button, found on the control panel may be pressed to obtain the printout.
  • Page 8 18. Cycle Complete: “CC” will display if the cycle was successful. 19. Print Mode: If the Saniclave 200 is connected to a printer, a cycle report will print when SET button is pressed, showing the date/time, preset and achieved parameters.
  • Page 9 Recommended Load Configuration Load including the trays must not exceed 4.5 lbs. Cycle Phases Fill Display reads “Fill” if the 'START' button is pressed and the chamber has not yet been filled with sufficient water. Preheat Mode. Display will read 'Ph' for Preheat. This mode will usually last for about 30 minutes, depending on the load size.
  • Page 10 If set correctly, the time and date will be printed on the printout. Print – To print, the printer must be connected to mains and plugged into the Saniclave 200 prior to use. After a cycle is complete, when display reads 'CC,' press the SET button once.
  • Page 11: Troubleshooting And Error Modes

    Error Conditions and Display Readings Note: If a failure occurs, the load must be considered non-sterile. The sterilizer and load should then be investigated. Contact Revolutionary Science if problem persists. 1. Manual Cycle Abort: If 'START' is pressed, display will read 'Err 1' for cycle abort.
  • Page 12 Water may leak from the unit if the drain valve is left open. Check drain valve and move it into the 'closed' position. If leaking persists, unplug from mains and contact Revolutionary Science. If any of the above occurs, turn off Saniclave 200 and investigate. If problem persists, contact Revolutionary Science...
  • Page 13: Warranty, Registration And Policy

    If your RS-SC-200 Saniclave 200 proves to be defective within this warranty period, APR&D, Inc. will repair or replace your unit at our option. This warranty will be void if...
  • Page 14: Specifications

    Automatic Sterilization time: 30 minutes Dry time: 30 minutes Capacity: 12.5 Liters Further Questions? Contact Information Revolutionary Science 17319 Lake Blvd Shafer, Minnesota 55074, USA Customer Service: 1(800) 775-7942 Technical Assistance: (651) 674-3112 Fax: (775) 218-7015 Web site: www.revsci.com E-mail:...
  • Page 15: Checklist And Notes Of Service Record

    Serial Number: __________________________________________ Model Number: __________________________________________ Purchased From: _________________________________________ Your Notes: ____________________________________________________________________________ ____________________________________________________________________________ ____________________________________________________________________________ ____________________________________________________________________________ ____________________________________________________________________________ ____________________________________________________________________________ ________________________________________________ Revolutionary Science recommends the weekly use of G. Bacillus Stearothermophilus biological indicator ampules. Place indicator in center of full load and test weekly.
  • Page 16: Care, Storage And Maintenance

    Caution: Water from the exhaust water tank or chamber may be warm or hot! 3. When your Saniclave 200 is cool, wipe both the exterior and the interior of the autoclave with a clean, dry paper towel or terry cloth. If needed, clean plastic parts with warm water or isopropyl alcohol.
  • Page 17: Sterilization Monitoring Guidelines

    Also, check the BI's instructions for proper use, storage and disposal. Only BI's that are cleared by the FDA and labeled for use with the same physical parameters (of exposure time and temperature) of the Saniclave 200 shall be used to monitor sterility of loads processed in Saniclave 200.