4.0
Cleaning and disinfection
Fig. 15
Fig. 16
Fig. 17
Fig. 18
12
4.1
Disinfectant /
Changing the chemical bottle
4.1.1
Disinfectant
(REF 534.3110.0)
For disinfection a 2% hydrogen peroxide solution with
added silver, ATMOS Green & Clean WK, is used. Hydro-
gen peroxide-based concentrates are particularly suitable
for reducing micro-organism levels in the water supply
because of their potent bactericidal and fungicidal action.
Hydrogen peroxide also show an outstanding ratio of
potential risk to disinfectant effi cacy, and present no risk
for the patient.
During each disinfecting operation in the ATMOS Aqua
clean system, a specifi ed amount of disinfected solution is
added to the 150 ml of water in the mixing container. This
provides the concentration of 0.0235% required for broad-
spectrum disinfection.
If the ATMOS Green & Clean WK bottle is completely
empty, an error message appears on the external indi-
cator, stating that no more disinfectant solution can be
withdrawn. The empty bottle should be detached from the
fi tting tube on the ATMOS Aqua clean and replaced with a
new one.
Please note that only ATMOS Green & Clean WK
provides the concentration required for broadspectrum
disinfection.
The use of chemicals other than from ATMOS may damage
the ATMOS Aqua clean as well as other devices which are
connected.
Safety precautions: Keep out of the reach of children. In
case of contact with the eyes, rinse thoroughly with water.
In case of skin contact, wash with water. Store ATMOS
Green & Clean WK at room temperature (or +5°C to
max. +30°C) and preferably protected from light!
Composition: hydrogen peroxide, silver and less than 0.1%
stabilisers, water.
For any further information, please see the safety
data sheet!
4.1.2
Changing the chemical bottle
Remove the chemical bottle from the unit (Fig. 16) . Pull gently
to remove the screw top and the attached tube which
extends into the bottle (Fig. 17) .
Remove the screw top from the new bottle of chemical, insert
the tube into it and screw back onto the bottle (Fig. 18).
Replace the bottle into the holder.
Need help?