Atmos Aqua clean Operating Instructions Manual

Hide thumbs Also See for Aqua clean:

Advertisement

Quick Links

English
ATMOS Aqua clean
Water Supply System
530.2120.B
502.1200.B
2014-08 Index: 07

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Aqua clean and is the answer not in the manual?

Questions and answers

Summary of Contents for Atmos Aqua clean

  • Page 1 English ATMOS Aqua clean Water Supply System 530.2120.B 502.1200.B 2014-08 Index: 07...
  • Page 2: Table Of Contents

    Retrofi tting of a delay PCB ........... 18 6.1.3 Delayed switching circuit for Aquastop ......19 6.1.4 Water connections ............20 Retrofi tting at ATMOS Servant 5 / 5 C..... 21-26 6.2.1 Drilling ................21 6.2.2 Tubing ................22 6.2.3 Add-on kit ..............
  • Page 3: Introduction

    ● Prior to start-up please peruse chapter 2.0 „For your safety“, in order to be prepared for any possible dangerous situations. ● These operating instructions are based on the version of the ATMOS Aqua clean available at the print date. The same applies to the applied safety-related standards.
  • Page 4: Function

    In the mixing container, the water inlet is in the form of a cascade , ensuring that the ATMOS Aqua clean is separated from the water supply, as required by the DVGW (The German Gas and Water Association) (DIN 1988, Part 4).
  • Page 5: Scope Of Supply

    1.0 Introduction Scope of supply Prior to dispatch, this ATMOS Aqua clean was subjected to an extensive functional test and has been carefully packed. Nevertheless, please compare the contents of the shipment on completeness immediately upon receipt (see delivery note).
  • Page 6: Explanation Of Pictures And Symbols

    Pictures contained in this manual Proper disposal  Warning, Important notes especial diligent notice! Symbols of the ATMOS Aqua clean The CE sign shows that this Serial number product meets the appropriate requirements of the EC Directive Order number 93/42 EEC.
  • Page 7: For Your Safety

    For your Safety • • ATMOS cannot guarantee perfect functioning Prior to starting up the ATMOS Aqua clean, read these operating instructions carefully. neither will it be liable for damage to people or • The safety standard of the ATMOS Aqua clean...
  • Page 8: Setting Up And Starting

    Initial connection of the chemical bottle Fig. 4 Open the housing of the ATMOS Aqua clean and feed the hose (Fig. 4) through the housing to the outside. Plug the provided seal onto the bottle and insert the hose. Fix...
  • Page 9: Explanation Of The Controls

    Control light 3: Disinfectant sensor (yellow) Alarm RESET button Switch On/Checking routine: The green control signal (Fig. 6) jitters: During the system is switched on the ATMOS Aqua clean automatically performs a checking routine. While this Fig. 5 procedure the control signal (Fig.
  • Page 10: Connection Conditions For Atmos S 61 Servant

    Fig. 10 Starting Sequence: 1. Open main water tap 2. Switch on ATMOS S 61 Servant 3. Switch on ATMOS Aqua clean If error “F1” appears, switch the ATMOS S 61 Servant off and on, until the error disappears. Fig. 11...
  • Page 11: Connection Conditions For Irrigation And Stimulation Devices

    Fig. 12 1. Open main water tap 2. Switch on irrigation and stimulation device 3. Switch on ATMOS Aqua clean If error “F1” appears, switch the irrigation and stimu- lation device off and on, until the error disappears. Chemical bottle connection PVC tubing Ø...
  • Page 12: Cleaning And Disinfection

    The empty bottle should be detached from the fi tting tube on the ATMOS Aqua clean and replaced with a new one. Please note that only ATMOS Green & Clean WK provides the concentration required for broadspectrum disinfection.
  • Page 13: Care And Cleaning

    1 jitters slowly. Supply container and pressure container are empty if no more water comes out of the unit. Now, the ATMOS Aqua clean can either be switched off or the normal operation can be activated by holding down the RESET button for another 8 sec.
  • Page 14: Maintenance

    Maintenance • Annual service inspection for the ATMOS Aqua clean with service kit (service interval is also dependent on water hardness!) This service contains a complete function control of the device, exchange of the input fi lter (Fig. 21), test (if required exchange) of the water dosing valves (Fig.
  • Page 15: Retrofi Tting

    Retrofi tting Retrofi tting at ATMOS S 61 Servant For retrofi tting at the ATMOS S 61 Servant the product, REF 530.2120.N is required. Scope of supply: Article description ATMOS Aqua clean 530.2121.0 1 piece Hose clip 061.0060.0 2 pieces Hose clip 061.0061.0...
  • Page 16 2. Switch on ATMOS S 61 Servant 3. Switch on ATMOS Aqua clean A Assemble the mounting plate and build it in. B Exchange existing plate against mounting plate If error “F1” appears, switch the ATMOS S 61 Servant off and on, until the error disappears.
  • Page 17: Tubing

    Retrofi tting 6.1.1 Tubing Article description Length Hose RD.I. 6,0 x 3,0 000.0091.0 1000 mm Hose RD.I. 8,0 x 3,0 005.0020.0 1000 mm Hose, Rilsan RD.A. 8,0 005.0025.0 240 mm Hose, Rilsan RD.A. 6,0 005.0173.0 200 mm Hose, Rilsan RD.A.4,0 005.0026.0 450 mm Hose, Rilsan RD.A.4,0...
  • Page 18: Retrofi Tting Of A Delay Pcb

    Retrofi tting 6.1.2 Retrofi tting of a delay PCB Drilling of the holes for the fi xing bolts. Place the PCB on the carrier ● plate and mark the drillings. Afterwards drill the holes with a 3.5 mm drill. Screw the 4 bolts M3x10 including washer and locknut. ●...
  • Page 19: Delayed Switching Circuit For Aquastop

    Retrofi tting 6.1.3 Delayed switching circuit for Aquastop...
  • Page 20: Water Connections

    Retrofi tting 6.1.4 Water connections...
  • Page 21: Retrofi Tting At Atmos Servant 5 / 5 C

    Note: When drilling the holes for the screwed connection (water in- / outlet into and out of the ATMOS Servant 5 housing) it is important to consider the equipment of the ATMOS Servant 5 which needs to be retrofi tted.
  • Page 22: Tubing

    Retrofi tting 6.2.2 Tubing Fig. 29 6.2.3 Add-on kit heating module existing Fig. 30...
  • Page 23: Delayed Switching Circuit For Aquastop Until 2005

    Retrofi tting 6.2.4 Delayed switching circuit for Aquastop until 2005...
  • Page 24: Delayed Switching Circuit For Aquastop From 2006

    Retrofi tting 6.2.4.1 Delayed switching circuit for Aquastop from 2006...
  • Page 25: Water Connections Until 2005

    Retrofi tting 6.2.5 Water connections until 2005...
  • Page 26: Water Connections From 2006

    Retrofi tting 6.2.5.1 Water connections from 2006...
  • Page 27: Troubleshooting

    Troubleshooting Prior to dispatch, the ATMOS Aqua clean was subjected to an extensive functional test. If, nevertheless, a failure should appear, you may possibly clear it yourself if you follow these notes: Error Possible reason Troubleshooting Device/pump does not – Loose power plug –...
  • Page 28: Technical Information

    Technical information Structure To ensure that it meets the most stringent in-service requirements, the ATMOS Aqua clean consists of several modules: Module 1 - Electronic control system. Module 2 - Pressure container; this generates the water pressure. In addition to the container, this module contains a magnetic valve, a pressure-sensitive switch and a non-return valve.
  • Page 29: Electrical Connections, Control Pcb

    Technical 9.2 Electrical connections, circuit board Overfl ow sensor (yellow) Pressure-sensitive switch (red) Mixing container probe (yellow/red/black) Disinfectant sensor (green) Level monitor in the pressure container (blue) Air cushion valve on the pressure container (black) MV4 External display Chemical valve (black) MV3 Water safety valve 5 (brown) Pump motor connection Membrane breaking (yellow)
  • Page 30: Technical Specifi Cations

    Permitted water pressure: 2-6 bar Operating pressure (water): 2.2 bar Operating pressure (air): 4 bar Maximum water fl ow rate: 550 ml/min Water quality: drinking water Disinfectant: ATMOS Green & Clean WK Hydrogen peroxide, 2% solution Mixing ratio: 1:85 standard setting 1:42 intensive decontamination...
  • Page 31: Disposal

    11.0 Disposal The ATMOS Aqua clean does not contain any hazardous materials. ● The housing is recyclable. ● Device and accessories must be decontaminated prior to disposal. ● Please take care on a careful separation of the different materials. ●...
  • Page 32: Notes On Emc

    12.1 Guidelines and Manufacturer‘s Declaration - Emissions The ATMOS Aqua clean is intended for use in the electromagnetic environment specifi ed below. The customer or user of the ATMOS Aqua clean should ensure that it is used in such an environment. Emissions Test...
  • Page 33 12.3 Guidelines and Manufacturer´s Declaration - Immunity The ATMOS Aqua clean is intended for use in the electromagnetic environment specifi ed below. The customer or user of the ATMOS Aqua clean should ensure that it is used in such an environment.
  • Page 34: Notes On Emc

    To determine the electromagnetic environment in regard to stationary transmitters, a study of the location is to be considered. If the measured fi eld strength at the location where the ATMOS Aqua clean is used exceeds the above compliance level, the ATMOS Aqua clean is to be observed to verify the intended use.
  • Page 35 For your notes...
  • Page 36: Atmos General Terms And Conditions

    ATMOS MedizinTechnik GmbH & Co. KG in accepting such goods or payment arrears. Acts of God or stoppages upon in writing by us.

Table of Contents