Range & Accuracy
Range
ohms Read (low)
ohms Read (high)
Range
Ohms Source (low)
Ohms Source (high)
Thermocouple Read and Source (All errors included)
Range in °C
J Thermocouple
K Thermocouple
T Thermocouple
E Thermocouple
R Thermocouple
S Thermocouple
B Thermocouple
L Thermocouple
U Thermocouple
N Thermocouple
mV Read/Source
Min
0.00 Ω
400.0 Ω
1500.0 Ω
Min
Max
5.00 Ω
400.00 Ω
5.00 Ω
400.0 Ω
400.0 Ω
1500.0 Ω
1500.0 Ω
3200.0 Ω
Max
400.00 Ω
1500.0 Ω
3200.0 Ω
Excitation Current
0.1 to 0.5 mA
0 0.5 to 3 mA
0.05 to 0.8 mA
0.05 to 0.4 mA
Min
-200.0
0.0
1200.0
-200.0
0.0
1370.0
-200.0
0.0
400.0
-200.0
0.0
950.0
-20
0
500
-20
0
500
600
800
1000
-200
0.0
900.0
-200
0.0
400.0
-200
0.0
1300.0
-10.00
75.00
Accuracy
± 0.1 Ω
± 0.5 Ω
± 1.0 Ω
Accuracy
0.15 Ω
0.1 Ω
0.5 Ω
1.0 Ω
Max
Accuracy
0.0
0.6°C
0.4°C
0.0
0.8°C
0.5°C
0.0
0.8°C
0.5°C
0.0
0.5°C
0.4°C
0.0
2.4°C
500
1.7°C
1750
1.3°C
0
2.4°C
500
1.7°C
1750
1.4°C
800
2.1°C
1000
1.7°C
1800
1.3°C
0.0
0.45°C
0.4°C
0.0
0.7°C
0.45°C
0.0
1.1°C
0.6°C
0.015 % ± 2 dig.
15