INSTRUCTION MANUAL
MULTI-PARAMETER TESTR 35 SERIES
pH / Conductivity / TDS / Salinity / Temperature
Part of Thermo Fisher Scientific
68X441601 Rev. 1 March 2010
Thank you for selecting our Multi-Parameter Testr. This manual serves the following (3) models:
• PCTestr 35 (Eutech PCTEST35-01X441504 / Oakton 35425-00)
pH / Conductivity / Temperature
• PTTestr 35 (Eutech PCTEST35-01X441505 / Oakton 35425-05)
pH / Total Dissolved Solids / Temperature
• PCSTestr 35 (Eutech PCSTEST35-01X441506 / Oakton 35425-10)
pH / Conductivity / Total Dissolved Solids / Salinity / Temperature

Getting Started:

Your instrument has been factory calibrated and usually works well out of the box. However, after
extended periods of non-use, it's best to remove the sensor cap and soak the
sensor (pictured here) in warm tap water or pH buffer for 10 minutes or so. A
brief rinse with deionized (DI) water is OK, but avoid soaking or storing in
deionized water as this will shorten the pH electrode life. Prior to taking
measurements, periodic calibration with certified standards is recommended for
best accuracy.
Your Testr begins in the measuring mode that was previously used. Just prior to measurement or
when switching modes, you will see the setting associated with each parameter i.e.) pH (buffer
group selected), Conductivity (Auto), TDS (factor), Salinity (unit of measure).
Setup:
Your Testr allows customization of various settings. To access the setup mode:
1.
With the Testr off, keep the
pressed down while you press and release
(Setup) will appear, then as you release
2.
Press
or
to choose
(Parameter Setup) or
menu.
3.
Press
to enter the selected setup menu.
*IMPORTANT* It is necessary to save your Parameter and System
changes in order for them to take effect.
1.
When you are finished making your desired changes, press both
time and keep them pressed until you see "SA" (Save) on the display.
2.
With the primary display "SA" and secondary display "YES", press
changes. The instrument will resume measurement mode with new setting(s).
Note: If auto-shut off is used, changes will be automatically saved 8.5 minutes after the last
change was made.
Parameter Setup:
parameters – pH, Conductivity TDS, Salinity. Note: only the PCS Testr will have all of
these options. See below for menus available from each parameter. To Navigate the
menus:
Press
Press
pH Options:
USA or NIST Buffer Group for calibration buffer option.
5-pt calibration (all points) or 3-pt calibration (middle three points only).
Salinity (SALt) Option (PCS Testr only)
Choose PPt (parts per thousand) or Per (percentage %) as unit of measure.
Total Dissolved Solids (tDS) Option (PT and PCS Tester only)
FACt factor the instrument uses to convert from conductivity to TDS value.
Adjustable from 0.40 to 1.00 (default factor is 0.71).
Conductivity Options (PC and PCS Tester only)
A.Cal (Automatic Calibration) Choose YES or NO (manual).
TIP:
The PC Testr 35 and PCS Testr 35 are capable of automatic or manual
conductivity calibration. In automatic calibration mode, the meter will automatically
choose one of (3) conductivity calibration standards depending on the ranges listed
below. If you will only use 84
automatic calibration is a time saving option. If you intend to calibrate with one or more
standards that are not listed below, choose "NO" which will disable auto calibration and
allow you to enter your desired value manually.
Conductivity Range
0.0 – 200.0
201 – 2000
.
2.01 – 20.00 mS
SPC (Single-Point Calibration) Choose YES or NO (multi-point calibration).
,
(Parameter) will appear.
TIP:
The PC Testr 35 and PCS Testr 35 are capable of single or multi-point conductivity
calibration. Use Single-Point Calibration to apply a single calibration value across all

(System Setup)

ranges. Use Multi-Point Calibration for individual calibration in each range. This will
restrict an individual calibration so that it is applied to one range only. When using
multi-point calibration, perform a calibration in each range that you expect to use for
best results.
System Setup:
and
at the same
Select
to make changes relating to the system. See below for available menus.
Note: other than changing Temperature units, it is advised to keep the factory default
to save the
settings for best results. To Navigate the System menus:
Press
Find Quality Products Online at:
Select
to make changes relating to the
to select or confirm the displayed option.
or
to scroll thru options or change values.
S, 1413
S, or 12.88 mS calibration standards,
µ
µ
Automatic Calibration Value
Available with
S
84
S
PCS only
µ
µ
S
1413
S
PC or PCS
µ
µ
12.88 mS
PC or PCS
to select or confirm the displayed option.
GlobalTestSupply
www.
.com
Press
or
to scroll thru menu options or change values.
Unit rSt (Instrument reset)
PH (pH) or EC (electrical conductivity / TDS / Salinity)
CAL (calibration reset) or FCt (Reset to factory default settings)
Set A.Off (Automatic shut off after 8.5 minutes) Choose YES or NO.
Set t.C (Temperature Coefficient) 0.0-10.0% (2.1% is default)
Set AtC (Auto Temperature Compensation) Choose YES or NO (25°C is used).
Set °C °F (select temperature units) Choose °Celsius or °Fahrenheit.
Temperature Calibration:
The factory temperature calibration should last for the life of the original sensor since it
doesn't normally drift. Temperature calibration is always recommended upon sensor
replacement. It may also be desirable to adjust the temperature to match a certified
accurate thermometer or another Testr. The temperature value is common to all
parameters so only one calibration is needed. To perform temperature calibration:
1.
Press
Place the reference thermometer and your Testr
to turn on meter.
into the same sample. Allow enough time for both to stabilize.
2.
Press
as needed to select the pH measuring mode. Press
calibration mode.
3.
Press
for 5 seconds to begin temperature calibration mode. The current
temperature will be displayed on top while the factory default temperature is
below.
4.
Press
or
to manually adjust to the desired temperature—up to ± 5°
C or ± 9° F of the factory default value.
5.
Press
to confirm and return to the pH measuring mode.
pH Calibration:
For best results, calibrate with certified accurate pH calibration standards (buffers).
You may calibrate up to five points with the USA (1.68, 4.01, 7.00, 10.01, 12.45) or
the NIST (1.68, 4.01, 6.86, 9.18, 12.45) buffer group.
1.
Press
to turn meter on and
2.
Rinse the sensor with clean water. Immerse the sensor into your pH buffer and
press
. The primary display will show the un-calibrated pH value, while the
secondary display should search for and lock on the closest automatic calibration
value.
3.
Allow the primary display to stabilize, then press
value. The primary value will blink briefly before the secondary value automatically
scrolls thru the remaining pH buffers available for calibration.
Repeat steps 2 & 3 with additional buffers or press
4.
measurement mode.
sales@GlobalTestSupply.com
to begin pH
to select pH mode as needed.
to confirm the calibration
to return to

Advertisement

loading

Summary of Contents for Thermo Scientific PCTestr 35

  • Page 1: Getting Started

     USA or NIST Buffer Group for calibration buffer option. Temperature Calibration: • PCTestr 35 (Eutech PCTEST35-01X441504 / Oakton 35425-00)  5-pt calibration (all points) or 3-pt calibration (middle three points only). The factory temperature calibration should last for the life of the original sensor since it pH / Conductivity / Temperature doesn’t normally drift.
  • Page 2: Sensor Replacement

    Conductivity Calibration (Automatic): Sensor Replacement: Message Indicates For best results, calibrate with certified accurate conductivity calibration standards. Your instrument includes a replaceable sensor (Eutech PCSENSOR - 01X097108 / Selection of multi-point calibration will allow up to three of the following values, while Oakton 35425-50).

This manual is also suitable for:

Eutech pctest35-01x441504Eutech pctest35-01x441505Oakton 35425-00Eutech pcstest35-01x441506Oakton 35425-05Oakton 35425-10