Quantum Data 700A Series Quick Start Manual page 10

Portable video test generators
Hide thumbs Also See for 700A Series:
Table of Contents

Advertisement

Calibrating
Calibrating Outputs
1.
Select Options > Calibrate > Reset to set the generator's
calibration factors to default nominal values.
2.
Press Options > Calibrate > Full Scale.
3.
Set the voltmeter's scale factor to be able to measure a nominal
1000 millivolts DC.
4.
Using a VGA-to-BNC cable, connect the generator's red output to
the positive input on the voltmeter, with a 75 ohm (+/- 1%) input
terminator. Connect the negative lead of the meter to any ground
pin.
5.
Select R_FS+ or -R_FS to set the DC output level to 1000 mV
+/- 3 mV.
6.
Repeat steps 4 and 5 for the green and blue outputs.
7.
Select Save.
8.
Select Options > Calibrate > Zero.
9.
Connect the red output to the positive input of the voltmeter.
10. Select R_Zero+ or -R_Zero to set the DC output level to 0 mV +/- 3 mV.
11. Repeat steps 9 and 10 for the green and blue components.
12. Select Save.
13. Select Options > NTSC_CAL.
14. Adjust the oscilloscope's timebase and sweep trigger to view the entire waveform. Adjust the
G_NTSC so that its chrominance is 100 IRE or 714mV.
15. Using a vector scope, use the R_NTSC and B_NTSC options to adjust the red and green outputs
using so that their signals align with the 2% boxes.
16. Press Save.
Note: The modulated RF output does not have any calibration settings.
10
Model 700A, 701A Quick Start Guide