Description Switching off Press the OFF key, or auto- matic switching off 1,10 or 60 The „LOVE CONTROLS“ handheld pressure min after the last keystroke. gauge HM28 is a digital pressure measuring In case of a temperature change the instrument instrument with integrated pressure sensor requires at least 30 minutes to adjust to the for the measurement of differential, relative or...
Page 5
Display example Notes Switching on Max. Range: Measuring range 200 mbar displayed in desired unit Full scale error Accuracy class as a % 0.05 % of full scale reading Normal print zero 123.45 mbar Only with differential (differential) hold display ====== or or relative pressure 123.4 mbar...
Configuration Select mode: Press >2 seconds on the display appears CONFIGURATION Store and exit: Press >2 seconds on the display appears SAVE CONFIGURATION Display example Notes set unit mbar, kPa, bar, UNIT O, mH mbar mmHg, psi, inH inHg, hPa, Pa, MPa RESOLUTION Display resolution, high...
RS 232 mode The answer is terminated with a CR - The checksum is formed from the least The HM28 can be remotely operated from a significant byte of the command string personal computer via an RS 232 interface. The automatic switching off (chapter 3.1) is inactive.
Battery change Open battery receptacle Insert 9 V-alkali battery (IEC 6LR61) or accumulator Ensure correct polarity! Correct disposel of the used batteries according to environment regulations! Recalibration Recalibration to be carried out by specially skilled staff only. Relevant instructions see in section 8. We recommend to recalibrate the instrument at least once a year.
Page 11
Measuring rate in RS232- mode Units Measuring ranges -Class 0,2 20 measurements/s up to 7 bar 10 to 30 bar 70 bar -Class 0,1 and 0,05 10 measurements/s mbar Measuring rate, 2 ½ or normal mode 5 measurements/s Memory capacity max.
Warning messages and faults Fault/indication Possible cause Remedy 10% exceeding or under- Apply permissible measuring ERROR running of measuring range pressure OUT OF RANGE CHANGE BATTERY Battery voltage too low Insert new battery No change in Over pressure applied to Dispatch instrument for repair measurement pressuresensor...
acceptable values acquired. In the Calibration instructions event of a maloperation, the calibration process is interrupted and the previous Recalibration to be carried out by calibration data are retained. specially skilled staff only! Actuate the following key combinations: • First press MODE and keep it An incorrect internal instrument tempera- ture can lead to accuracy fluctuations! pressed...