Page 1
Instruction Manual Model 1823A 2.4 GHz Universal Frequency Counter GlobalTestSupply www. .com Find Quality Products Online at: sales@GlobalTestSupply.com...
Page 2
Introduction Thank you for purchasing our product. Electronic measuring instruments produced by us are high technology products made under strict quality control. We guarantee their exceptional precision and utmost reliability.
: Depending on input frequency • GATE TIME < 10 mS------------Somewhere between 0.9 and 9 mS < 0.1 S--------------Somewhere between 9 and 90 mS ...
3. OPERATION 3-1. Controls, indicators and connectors FIG(2) FRONT PANEL GATE INDICATOR : The gate light, when lit, indicates the main gate is open and measurement in progress.
Page 11
INPUT C, BNC : Input for frequency measurements above 80 MHz Female BNC connector terminated in 50 Ω. LOW PASS FILTER : With this switch pushed in, the input B is routed through a SWITCH(LPF) low-pass filter with -3 dB ...
Page 12
displays that count. f. T.l (A→B). When this mode is selected, the unit measures the time interval from an edge of the input A signal to an edge of the input B signal. Positive going or Negative going edge of each signal is ...
Page 13
VOLTAGE FUSE POWER MAX 115V ~ 0.5A F 230V ~ 0.2A F S/N : Made in Korea FIG (3) REAR PANEL ...
3-2. Operating Instruction Below is the basic operating information needed for frequency counter. a. Connect the unit to AC power cord into receptacle on rear panel and ...
3-6. Time Interval Measurements(A → B) In time interval mode, unit measures the elapsed time from a selected edge of the input A waveform to a selected edge of the input B waveform.
Frequency Ratio Measurements (A/B) 3-7. In this mode of operation, the counter displays the ratio of the frequency applied to input A to the frequency applied to input B.
4. MAINTENANCE CAUTION IT IS ESSENTIAL FOR SAFETY TO PROPERLY MAINTAIN AND SERVICE THIS INSTRUMENT WARNING VOLTAGES WITHIN THIS INSTRUMENT ARE SUFFICIENTLY HIGH TO ENDANGER LIFE. COVERS MUST NOT BE REMOVED EXCEPT BY PERSONS QUALIFIED AND AUTHORIZED TO DO SO AND THESE ...
5-3. Line Frequency Measurements Use of the attenuator, low pass filter, and/or x10 probe is advisable when measuring line frequency because the high amplitude signal and noise can cause ...
Need help?
Do you have a question about the 1823a and is the answer not in the manual?
Questions and answers