  • Page 2: Release Information

    PHABRIX Ltd. The information in this document has been produced by PHABRIX Ltd with care and is believed to be accurate. PHABRIX Ltd does not assume responsibility for loss or damage resulting from errors, omissions or inaccuracies herein. This document is subject to change and revisions may be made and issued to include such changes.
  • Page 3 Jitter (Optional extra) ............. 8 System wide Settings ............9 Memories ................ 9 Logging ................9 Log Eye Jitter ..............9 Jitter Thresholds ............10 Eye Timings ............... 10 Eye Amplitude ............10 PHABRIX Sx Manual August 21, 2009 Page 3...
  • Page 4: Operation

    TURNING THE UNIT ON OR OFF To turn on your PHABRIX Sx press the red button at the top right hand side of the unit. Once the system has started, pressing the button for a few seconds will turn it off. If for some reason the unit stops responding, pressing and holding the button for a few seconds will turn it off.
  • Page 5: Analyser

    Note that the eye display is of the signal pre-equaliser, the equalizer in a system being used to compensate for long cable lengths. There is no method for the SxE to view the signal post-equaliser.
  • Page 6 0.34UI 0.34UI Full-Scale Alignment 0.3UI 0.2UI 0.2UI (Red Threshold) Alignment 0.2UI 0.14UI 0.14UI (Yellow Threshold) Timing 3.4 UI 1.7UI 0.34UI Full Scale Timing 0.2UI (Red Threshold) Timing 1.4UI 0.7UI 0.14UI (Yellow Threshold) PHABRIX Sx Manual August 21, 2009 Page 6...
  • Page 7: Advanced Analysis Features (Optional Extra)

    Clear: This button clears the eye display. It is only enabled when the update mode is set to infinite update mode. PHABRIX Sx Manual August 21, 2009 Page 7...
  • Page 8: Jitter (Optional Extra)

    Clear: This button clears the eye display. It is only enabled when the update mode is set to infinite update mode. PHABRIX Sx Manual August 21, 2009 Page 8...
  • Page 9: System Wide Settings

    Logging allows events to be detected and recorded for future examination. The events to be logged can be specified so that unwanted events do not appear in the event log. If a system has a problem with intermittent signals, the PHABRIX Sx can be connected to that source and can be left for several days to log any errors.
  • Page 10: Jitter Thresholds

    If checked, these controls will enable logging of invalid amplitude or over-shoot/undershoot values. An event will be logged if the value exceeds the valid SMPTE range and an event will be logged when the signal becomes valid again. PHABRIX Sx Manual August 21, 2009 Page 10...