7.5.3 Text Format Log Files (Audit and Alarm Log)

Alarm Event logs for each process group and the Audit log are
archived into individual files.
The filenames are formatted as shown in Table 7.6.
Log File
<hour min> <day, mm, yy>
Alarm Event
<process group tag>.e00
<hour min> <day, mm, yy>
Audit Log
<instrument tag>.a00
Table 7.6 Text Format Log File Format
If one of the archive log files becomes full (>64000 entries) a new
file is created with an extension incremented by 1, for example:
a01, e01.
New text format log data files are also created when the daylight
saving period starts.

7.5.4 Daylight Saving

Files containing data generated during the daylight saving period
have '~DS' appended to the filename.
Start of daylight saving period
A daily file is started at 00:00:00 on 30
30Mar12AW633.D00
Summertime starts at 2:00am on 30
changes automatically to 3:00am.
The existing file is closed and a new file is created filename:
30Mar12AW633~DS.D00
The file '30Mar12AW633.D00' contains data generated from
00:00:00 to 01:59:59.
The file '30Mar12AW633~DS.D00' contains data generated
from 03:00:00.
End of daylight saving period
A daily file is started at 00:00:00 on 26
26Oct12AW633~DS.D00
Summertime ends at 3:00am on 26
clock changes automatically to 2:00am.
The existing file is closed and a new file is created filename:
26Oct12AW633.D00
The file '26Oct12AW633~DS.D00' contains data generated
from 00:00:00 to 02:59:59.
The file '26Oct12AW633.D00' contains data generated from
02:00:00.
OI/AXM630–EN Rev. I
Filename
th
March 2012 filename:
th
March 2012 and the clock
th
October 2012 filename:
th
October 2012 and the

7.5.5 Text Format Data Verification and Integrity

When text format data is saved to the archive media it is
checked automatically to verify that the data stored on the
media matches exactly what is stored in the internal memory.

7.6 Binary Format Data Files

Binary format archived data is stored in a secure binary encoded
format. A separate file is created for each recording channel.
The log data is stored in an encrypted text format.
The files can be read on a PC using ABB's DataManager Pro
data analysis software package.

7.6.1 Binary Format Data Filenames

When the 'Archive file format' parameter is set to 'Binary
format', the 'New File Interval' parameters (see Section 6.6.4,
page 40) are disabled and binary format filenames are created
with the content – see Table 7.7.
Data Type
<Start Time HHMMSS>
<Start Date DDMMMYY>
Stream Data
Ch<Group><Channel><monitor tag>
e.g.: 14322719Mar12Ch1_2Final Water3
<Start Time HH_MM>
<Start Date DDMMMYY>
Alarm Event
<Process Group Tag>
Log Data
e.g.: 14_3219Mar12Final Water5
<Start Time HH_MM>
<Start Date DDMMMYY>
Audit log
<Instrument Tag>
Data
e.g.: 14_3219Mar12Final Water3
Table 7.7 Binary Format Data Filenames
Filename extensions are assigned according to the data type
archived – see Table 7.8.
Data
Stream data
Alarm Event Log data files
– the historical record of the alarm events
related to the stream plus the history of any
operator messages
Audit Log data files
– the historical entries from the audit log
Table 7.8 Binary Format Data Filename Extensions
7 Logging
Filename Content
Binary
Format File
Extension
*.B**
*.EE*
*.AE*
49