Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer's declaration - electromagnetic immunity
The PO3M is intended for use in the electromagnetic environment specified below.
The customer or the user of the PO3M should assure that it is used in such an environment.
IMMUNITY
test
Radiated RF
IEC/EN 61000-4-3
IEC/EN 60601
test level
3 V/m 80 MHz to
2.5 GHz
Compliance
Electromagnetic environment
level
- guidance
Portable and mobile RF
3 V/m
communications equipment
should be used no closer to any
part of the [PO3M], including
cables, than the recommended
separation distance calculated from
the equation applicable to the
frequency of the transmitter.
Recommended separation
distance:
d=1.2 p
d=2.3 p
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the
transmitter manufacturer and d is
17
80 MHz to 800 MHz
800 MHz to 2,5 GHz