Milltronics miniranger plus Instruction Manual

Milltronics miniranger plus Instruction Manual

Hide thumbs Also See for miniranger plus:

Advertisement

Milltronics_LevelMonitor_MiniRanger_im_D103
Instruction Manual January 2003
miniranger
PLUS

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the miniranger plus and is the answer not in the manual?

Questions and answers

Summary of Contents for Milltronics miniranger plus

  • Page 1 Milltronics_LevelMonitor_MiniRanger_im_D103 Instruction Manual January 2003 miniranger PLUS...
  • Page 2 Contact SMPI Technical Publications at the following address: Technical Publications Siemens Milltronics Process Instruments Inc. 1954 Technology Drive, P.O. Box 4225 Peterborough, Ontario, Canada, K9J 7B1 Email: techpubs@siemens-milltronics.com For the library of SMPI instruction manuals, visit our Web site: www.siemens-milltronics.com © Siemens Milltronics Process Instruments Inc. 2003...
  • Page 3: Table Of Contents

    Table of Contents General Information................5 About this Manual..............5 About MiniRanger Plus............6 Specifications ..................7 Installation...................11 Dimensions..............11 Mounting................13 Interconnection..............18 Synchronization...............19 Communication ...............21 Start Up....................25 General ................25 Display and Keypad ............26 Run Mode................27 Program Mode..............28 Setting Up................29 Display Messages ............31 Functional....................33 Transceiver..............33 Transducer ..............33 Temperature..............33...
  • Page 4 Page 4 MiniRanger Plus 7ML19981CJ01...
  • Page 5: General Information

    About this Manual It is essential that this manual be referred to for proper installation and operation of your MiniRanger Plus. As MiniRanger Plus requires an ultrasonic transducer to make a working system, refer to the associated transducer manual as well.
  • Page 6: About Miniranger Plus

    About MiniRanger Plus Note: The MiniRanger Plus is to be used only in the manner outlined in this instruction manual. MiniRanger Plus is a versatile material level monitoring instrument. Material level measurement is achieved using advanced sonic echo ranging techniques. The unit is complete with an integral multifunction liquid crystal display, a four button keypad and an integral junction box.
  • Page 7: Specifications

    Specifications MiniRanger Plus Power • 9-250 Vac, 40-70 Hz, 38 VA, 12W or 9-250 Vdc, 11W • leakage current to ground: 3 mA at 120 Vac, 3.5 mA at 240 Vac Fuse • F1 2AG, FastBlow, 1.5 amp, 250 V •...
  • Page 8 • installation category: II • pollution degree: 4 Display • 38 x 100 mm (1.5 x 4") multi-field backlit liquid display Communications • Dolphin Plus compatible via optional Milltronics ComVerter Enclosure • Type 4X/NEMA 4X/IP65 Weight • 1 kg (2.2 lb.)
  • Page 9 • Belden 8760, shielded / twisted pair, 18 AWG (0.75mm ) or equivalent • Maximum cable length 10 m (50 ft) RJ-11 Link (Rack & Panel Only): • No shielded cable necessary. • Maximum cable length 3 m (10 ft) 7ML19981CJ01 MiniRanger Plus Page 9...
  • Page 10 Page 10 MiniRanger Plus 7ML19981CJ01...
  • Page 11: Installation

    This product is susceptible to electrostatic shock. Follow proper grounding procedures. Dimensions Wall Mount 270mm (10.6”) 254mm (10”) 138mm (5.4”) 69mm (2.7”) sealed enclosure integral junction box 74mm (2.9”) mounting hole 6.4mm (1/4) ∅ (0.2”) 7ML19981CJ01 MiniRanger Plus Page 11...
  • Page 12 Rack Mount The MiniRanger Plus rack mount version is a standard DIN 3U/14 HP, 4 rail plug-in unit, for a standard 84 HP deep sub-rack. Panel Mount 208mm 72mm (8.2”) (2.9”) 16mm (0.6”) 144mm (5.7”) 150mm (5.9”) slip on mounting bracket top and bottom screws to be tightened to no more than 1 inch / lb.
  • Page 13: Mounting

    250 V. Hazardous voltage present on transducer terminals during operation. Relay contact terminals are for use with equipment having no accessible live parts and wiring having insulation suitable for at least 250 V. 7ML19981CJ01 MiniRanger Plus Page 13...
  • Page 14 Rack Mount Slide the MiniRanger Plus into the sub-rack, aligning the connector with the rail mounted terminal block. connector, 2 row form D male 32 way, per DIN 41612 terminal block, 2 row form D female 32 way, per DIN 41612 Push the MiniRanger Plus into the sub-rack until the MiniRanger Plus front cover is snug to the front rails.
  • Page 15 Terminal Block Layout (Rack Mount) RS-232 ground only Warning: All field wiring must have insulation suitable for at least 250V. Hazardous voltage present on transducer terminals during operation. 7ML19981CJ01 MiniRanger Plus Page 15...
  • Page 16 Panel Mount Slide the MiniRanger Plus through the panel cut-out. Assemble mounting brackets to the MiniRanger Plus chassis by hooking them into the top and bottom. mounting bracket (top and bottom) cut-out dimension 68 + 0.7 mm x 138 + 1.0 mm (2.68 + 0.03”...
  • Page 17 Terminal Block Layout (Panel Mount) RS-232 ground only Warning: All field wiring must have insulation suitable for at least 250V. Hazardous voltage present on transducer terminals during operation. 7ML19981CJ01 MiniRanger Plus Page 17...
  • Page 18: Interconnection

    Interconnection System Diagram MiniRanger Plus *Milltronics transducer, see Specifications mA output customer device isolated customer alarm, pump or control device bipolar current * Milltronics BIC-2, buffered interface (Milltronics communications) converter *rack and panel only Transducer Wall Mount Rack and Panel Mount...
  • Page 19: Synchronization

    Synchronization In applications where more than one MiniRanger Plus, to a maximum of 8, are going to be used or where their transducers are sharing a common conduit, synchronization is required. When synchronized, no device transmits within 180 ms of the prior one(s).
  • Page 20 Note: All relays are certified for use in equipment where the short circuit capacity of the circuits in which they are connected is limited by fuses having ratings not exceeding the rating of the relays. Page 20 MiniRanger Plus 7ML19981CJ01...
  • Page 21: Communication

    Bipolar and RS-232 communication are both achieved through a common `TX' port terminal C4 and C6, without any hardware or software intervention. IBM PC Computer Connection / RS-232 MiniRanger Plus (Rack and Panel mount (only) *100 Ω computer serial port...
  • Page 22 BIC-2 Connection / Bipolar Current MiniRanger Plus Rack and Panel mount BIC-2 ♦ ♦ typical of BIC-2 ports 1-6 C8 is direct to ground Note: Ground shield at one end only Power 9-250 V , 40 – 70 Hz, 38 VA, 12W...
  • Page 23 Dolphin Optional Dolphin Interface Wall Mount ComVerter MiniRanger Plus computer RS-232 link via infrared ComVerter Rack and Panel Mount computer RS-232 link direct Note: Refer to Dolphin instruction manual for interconnection details. 7ML19981CJ01 MiniRanger Plus Page 23...
  • Page 24 Page 24 MiniRanger Plus 7ML19981CJ01...
  • Page 25: Start Up

    Start Up General The MiniRanger Plus has two modes of operation: run and program. When the unit is powered, after installation procedures have been completed, it is programmed to start up in the run mode, to detect the distance from the transducer face to the target in meters.
  • Page 26: Display And Keypad

    Display and Keypad Wall Mount Rack and Panel Mount communication jack keypad keypad Page 26 MiniRanger Plus 7ML19981CJ01...
  • Page 27: Run Mode

    = low, = low-low reading questionable, appears during fail-safe operation auxiliary reading relay status = normal operation =fail-safe operation Keypad access display toggle display mA program temperature % and (auxiliary) mode (auxiliary) units 7ML19981CJ01 MiniRanger Plus Page 27...
  • Page 28: Program Mode

    Rack and Panel Mount parameter type ( transducer, relay, or mA output ) reading units `change value' accessed breakpoint accessed Keypad alternates program mode between access scroll ‘select parameter’ scroll up run mode down and ‘change value’ functions Page 28 MiniRanger Plus 7ML19981CJ01...
  • Page 29: Setting Up

    To Access Program: run display program starts at Press P001 To Select a Parameter : Press To Change a Parameter Value : select parameter, e.g. P001 =3 initiate change Press function Note: Security must be disabled! 7ML19981CJ01 MiniRanger Plus Page 29...
  • Page 30 Press ‘select parameter’ function, Note: e.g. p001 =4 Must be pressed to Save change! Express: Parameter: to jump back to P001, or Value: to jump to the factory setting. e.g. P999 jump back to P001 Page 30 MiniRanger Plus 7ML19981CJ01...
  • Page 31: Display Messages

    To Return to Run Mode from the program mode e.g. P001 Press exit and return to run mode. Display Messages ‘OPEn’ in transducer circuit. Refer to Troubleshooting. ‘Short’ in transducer circuit or wiring reversed. Refer to Troubleshooting 7ML19981CJ01 MiniRanger Plus Page 31...
  • Page 32 Express Viewing To view the following readings while in run mode, simultaneously press the depicted keys. Echo Confidence (P805) Echo Strength (P806) Noise (P807) Distance (P923) Page 32 MiniRanger Plus 7ML19981CJ01...
  • Page 33: Functional

    Functional Transceiver The MiniRanger Plus transceiver operates under 1 of 5 sets of preset conditions (P003), summarized as follows: parameter measurement echo f-s timer* long value response verification filter (P070) pulses* 0.1 m/min (slow) 1 m/min 10 m/min 102 m/min / 1.7 m/sec 1020 m/min / 17 m/sec (fast) *each measurement consists of one short pulse plus the set number of long pulses.
  • Page 34: Sound Velocity

    The basis is to physically measure the level (measuring tape or sight glass) and enter the value via P651. The MiniRanger Plus then calculates the sound velocity by comparing the entered physical measurement to its own ultrasonic measurement.
  • Page 35: Loss Of Echo

    Loss of Echo A loss of echo occurs when the MiniRanger Plus deems that the calculated measurement is unreliable, i.e. the echo confidence (P805) is less than the threshold (P804). This can be due to such circumstances as high level of electrical noise, poor grounding or poor transducer aiming.
  • Page 36: Relays

    Relays General Two onboard relays are provided in the MiniRanger Plus. Each relay can be assigned one of two functions: Alarm or Pump. Relay two can also be used to control a totalizer when the unit is in OCM mode.
  • Page 37 The MiniRanger Plus can convert a head measurement into a flow rate using a nine breakpoint head vs. flow characterisation curve; this chart is usually available from the manufacturer of the weir or flume or other primary measuring device (PMD).
  • Page 38 1 function e.g. 0 = off relay 2 function Press e.g. 0 = off Press relay 1 on setpoint Press relay 2 on setpoint Press relay 1 off setpoint Press relay 2 off setpoint Page 38 MiniRanger Plus 7ML19981CJ01...
  • Page 39 * not applicable to high level switch operation, P001 = 4. In such cases, the timer is bypassed and the mode is always ‘low’. 7ML19981CJ01 MiniRanger Plus Page 39...
  • Page 40: Ma Output

    Output The MiniRanger Plus can be programmed to provide an mA output (P200) of 0 to 20 or 4 to 20 mA, and for proportional or inverse span. Programming Upon entering the Program mode, the mA output level holds its prior value.
  • Page 41: Run / Program

    Run / Program When the MiniRanger Plus changes from run to program, the transceiver stops operating and the last measurement is stored; the associated reading, alarm status and mA output are held (pump control relays are turned off). As a courtesy, the unit reverts to the parameter last addressed during the previous program session.
  • Page 42 Page 42 MiniRanger Plus 7ML19981CJ01...
  • Page 43: Applications

    Applications This section highlights the most common applications to which the MiniRanger Plus is likely to be applied. Other applications not described here, such as position monitoring of a piston on a wood pulverizer, are basically improvisational and imaginative ways of applying the MiniRanger Plus.
  • Page 44: Simple Level Application

    Simple Level Application One of the most common applications of a Milltronics process material level monitor is simple level measurement, whereby the material level or space between the transducer and reflecting surface is measured and displayed. This may or may not include alarms and mA output.
  • Page 45 `3.9' value, 0.1 m deadband) relay 2, off setpoint = 1.1 m (nominal P113-2 enter `1.1' value, 0.1 m deadband) P200 enter `2' mA output = 4 - 20 press run to start normal operation 7ML19981CJ01 MiniRanger Plus Page 45...
  • Page 46: Pump Control Application

    6 sec is required to protect the pumps from running dry. P006 3.6m relay 1, pump on P007 P112-1, 2m 3.0m relay 2, pump on P112-2, 1m relay 1 & 2, pumps off P113-1&2, 0.25m Page 46 MiniRanger Plus 7ML19981CJ01...
  • Page 47 1, off setpoint = 0.25 m (nominal P113-1 enter `.25' value, 0.1 m deadband) relay 2, off setpoint = 0.25 m (nominal P113-2 enter `.25' value, 0.1 m deadband) P802 enter `1' submergence = on press run to start normal operation 7ML19981CJ01 MiniRanger Plus Page 47...
  • Page 48: High Level Switch

    High Level Switch The MiniRanger Plus can be programmed as a high level switch. Significant differences between this and other modes of operation are: • the fail-safe function (P070 and P071) is effectively disabled • on loss of echo, the measurement immediately defaults to a low value (P006) •...
  • Page 49 1 P112-1 enter `1' relay 1, on setpoint = 1 m relay 1, off setpoint = 2 m (nominal P113-1 enter `2' value, 1 m deadband) press run to start normal operation 7ML19981CJ01 MiniRanger Plus Page 49...
  • Page 50: Applications With Standpipe

    Applications with Standpipe In many liquid applications, access to the vessel must be made via a standpipe. In such cases, Milltronics can provide optional flanged transducers, or split flanges that mate to a flanged standpipe. Another option is to hang the transducer from a blind flange.
  • Page 51: Volume Conversion

    • if the difficulty doesn't subside, the mA output must immediately default to high. • an XCT-8 transducer is mounted in a standpipe so that the face is 18 in. above the vessel top. 4.50 ft (P006 3.00 ft (P007) 12.00 ft (P053) 1.5ft (P052) 7ML19981CJ01 MiniRanger Plus Page 51...
  • Page 52 Tank Dimension L length, excluding parabolic ends). No entry required. (preset to “2”, P200 mA Range for 4-20 mA output). Now in the RUN mode, the Reading Value and mA output will represent volume in cubic feet. Page 52 MiniRanger Plus 7ML19981CJ01...
  • Page 53: Open Channel Monitoring (Ocm)

    P604 Maximum flow (associated with P007) P606 Time Units Then program the two following parameters to closely match the flow curve of your primary measuring device (PMD). Number Value P610[1..9] Head Breakpoints P611[1..9] Breakpoint Flowrates 7ML19981CJ01 MiniRanger Plus Page 53...
  • Page 54 Run the PMD at different head levels and verify the OCM reading given on the MiniRanger to double-check the programming. If this is not possible at least double-check the breakpoints for accuracy. Page 54 MiniRanger Plus 7ML19981CJ01...
  • Page 55: Parameter Description

    Space (P001 = 2) Distance (P001 = 3) H.L.S (P001 = 4) P002 Material Optimizes measurement reliability for target type. entry : 1 = liquids, fluids or flat surfaces 2 = irregular surfaces and solids 7ML19981CJ01 MiniRanger Plus Page 55...
  • Page 56 If MiniRanger Plus cannot keep up with the rate of level change, select a faster rate. If the reading bounces around an average value, select a slower rate.
  • Page 57 P050 Tank Shape This parameter, in conjunction with parameters P051, P052 and P053, enable the MiniRanger Plus to show readings based on reservoir volume (rather than level). Note: Volume calculation is an optional feature. MiniRangers with the volume feature have this icon on the front: Enter the Tank Shape value that matches the monitored vessel or reservoir.
  • Page 58 3 = parabolic bottom 4 = spheric bottom 5 = angled bottom 6 = flat ends 7 = parabolic ends 8 = sphere Tank dimensions ‘A’ and ‘L’ are entered via parameters P052 and P053 respectively. Page 58 MiniRanger Plus 7ML19981CJ01...
  • Page 59 The operation of the device is not affected by the Offset Reading. This value is used for display purposes only. All control measurements are still referenced to Empty. The Offset Reading is ignored when open channel monitoring (OCM, P600 = 1) is used. 7ML19981CJ01 MiniRanger Plus Page 59...
  • Page 60 Selects the default measurement in the event that the fail-safe timer expires. entry : 1 = high; maximum span value 2 = low; minimum span value 3 = hold; hold current value Note: Disabled for h.I.s. operation (P001 = 4) Page 60 MiniRanger Plus 7ML19981CJ01...
  • Page 61 4 = 20 to 4 mA P300 Temperature, Transducer Max The maximum temperature encountered by the integral (transducer) temperature sensor. P341 Run Time View the accumulated number of days the MiniRanger Plus has been operating. 7ML19981CJ01 MiniRanger Plus Page 61...
  • Page 62 If flowrate is displayed in millions of gallons / day, and the maximum flowrate is 376,500,000 gallons / day, enter “376.5” for Maximum Flow (P604) and “4” for Time Units (P606). Values 1 = seconds 2 = minutes 3 = hours 4 = days (default) Page 62 MiniRanger Plus 7ML19981CJ01...
  • Page 63 Values 0 = 1 1 = 10 2 = 100 (default) 3 = 1000 4 = 10,000 5 = 100,000 6 = 1,000,000 7ML19981CJ01 MiniRanger Plus Page 63...
  • Page 64 • scroll value to the actual level, space or distance respective to the mode of measurement (P001). • Press (to validate entry). The MiniRanger Plus calculates the new sound velocity for measurement correction. P661 Temp Fixed For use with ST-25 Ultrason® transducers. As these transducers do not have an integral temperature sensor, a nominal value representing the ambient temperature of the transducer must be entered.
  • Page 65 2 = formatted commas are inserted between message fields The MiniRanger plus Rack and Panel units provide a proprietary 20 mA bipolar current loop for communication with the Milltronics BIC-2 peripheral device. The MiniRanger Plus uses simplex convention communication protocol.
  • Page 66 2 characters, $32 to $FA (convert to decimal and subtract 100, e.g. temperature = 20 °C) Percent 5 characters, 00.00 to 9999. (e.g. current level = 85.00% of span, EEEE. = field overflow) End Of Message CR ($0D) Page 66 MiniRanger Plus 7ML19981CJ01...
  • Page 67 (P070). short minimum echo confidence for short pulse echoes; those within the short pulse echo range. long minimum echo confidence for long pulse echoes. entry: ## : ## = short : long 7ML19981CJ01 MiniRanger Plus Page 67...
  • Page 68 Enter the value (in percent of echo height) to ensure the Echo Lock Window intersects the Echo Profile at the sharpest rising portion of the Echo Profile representing the true echo. This value is preset to 50%. Values 5 to 95% Page 68 MiniRanger Plus 7ML19981CJ01...
  • Page 69 Tests the memory. Test is initiated by scrolling to the parameter or repeated by pressing. display: PASS = normal FAIL = consult Milltronics P907 Programmer Interface Tests the infrared communication link. Test is initiated by scrolling to the parameter or repeated by pressing.
  • Page 70 P920 Reading Measurement The reading as viewed in the run mode. Note: Not accessible via application device keypad P999 Master Reset Resets parameters to their factory setting. Press Page 70 MiniRanger Plus 7ML19981CJ01...
  • Page 71: Troubleshooting

    P070 clean move transducer to a better location material build up on mount in standpipe transducer face *refer to associated transducer manual. 7ML19981CJ01 MiniRanger Plus Page 71...
  • Page 72 Reading does not MiniRanger Plus re-aim transducer* change, but the processing wrong echo, check standpipe for level does i.e. vessel wall, internal burrs or welds...
  • Page 73 High level reading material is within near decrease blanking lower than material blanking zone P800 (limit to level transducer min range*) echo multiple being or raise transducer processed *refer to associated transducer manual. 7ML19981CJ01 MiniRanger Plus Page 73...
  • Page 74: Maintenance

    The MiniRanger Plus requires no maintenance, however a program of periodic checks is advised. If the MiniRanger Plus wall unit is mounted in a dusty or oily environment, make sure that the programmer, optional Dolphin ComVerter, and front cover are kept clean using a vacuum cleaner and a clean dry brush;...
  • Page 75: Appendices

    P005 noise P807 offset calibration P650 offset reading P062 operation P001 P740 peripheral communication primary measuring device P600 P907 programmer interface not applicable to wall mount units not applicable to rack and panel units 7ML19981CJ01 MiniRanger Plus Page 75...
  • Page 76 Sound Velocity Calibration P661 Temperature, Fixed P740 Peripheral Communication* P800 Near Blanking P801 Range Extension P802 Submergence Transducer P804 Confidence Threshold P820 Algorithm P825 Echo Marker Trigger P830 TVT Type * rack and panel mounts only Page 76 MiniRanger Plus 7ML19981CJ01...
  • Page 77 Siemens Milltronics Process Instruments Inc. Siemens Milltronics Process Instruments Inc. 2003 1954Technology Drive, P .O. Box 4225 Subject to change without prior notice Peterborough, ON, Canada K9J 7B1 Rev. 1.3 Tel: (705) 745-2431 Fax: (705) 741-0466 *7ml19981cj01* Email: techpubs@siemens-milltronics.com...

Table of Contents

Save PDF