Technical instructions
Lightly lubricate the plug sealing rings with
Klübersynth VR-252, Magnetic order No.
R50014.
The use of other lubricants can damage seal-
ing rings and the plastic housing.
Ensure that the groove on the plug is aligned
with the mark on the housing.
Otherwise, the plug cannot be connected
properly.
Fig. 7 – Inserting the actuator plug
Batteries must only be charged in well-venti-
lated rooms. The gases resulting from the
charging
process
Do not charge the batteries in damp environ-
ments!
If the charging cycle lasts more than 20
hours, the battery or control unit is faulty.
Remove the mains adapter from the socket.
(See also 5.1 Maintenance)
During charging, any lifter controlled by the
MOBILETTE must not be used.
Wait until the charging process is over before
using the MOBILETTE.
Fig. 8 – LED for battery charge status
531e2931_0101
February 2008
For your nearest Distributor, call 1-800-694-4525
Connecting the actuator(s)
Insert the actuator plug into the corresponding socket
Then proceed as follows:
Repeat steps
Otherwise, the actuator output which is not used is closed with a watertight blanking
plug at the factory. This plug must not be removed.

3.5 Startup

Charging the battery
The battery charging process is started when the mains adapter / mains cable is
are
dangerous.
plugged in, or the battery pack is fitted in place while the mains adapter / mains cable
is plugged in.
An LED (Fig. 8,
LED yellow
LED green
LED unlit
Current consumption at full load
During commissioning, measure the actuator's maximum current consumption at full
load. It must not exceed the value specified on the type plate of the linear actuator.
Higher current consumption means that the linear actuator is overloaded and may be
damaged.
An integrated overcurrent cut-off automatically deactivates the actuator if the cur-
rent consumption is too great.
Magnetic – The Linear Drives Company™
Insert the plug (the sealing rings must not be visible)
Ensure that the groove on the plug is aligned with the mark on the control unit.
Use the special plug disassembling tool No. 140375 to turn the plug approx. 30°
to the right up against the stop in order to lock it in position.
to
if you wish to connect an (optional) second actuator.
) indicates the battery's charge state
1
Batteries are being charged, mains voltage connected.
Batteries are fully charged, mains voltage connected.
Mains voltage not connected.
Control unit Mobilette
on the control unit (Fig. 5).
page 6/11