4.
Confirm System Check Screen .
This display appears when you turn on the meter . Do not use the meter if
the display check screen does not exactly match the example. Contact
Customer Care.
See The AlphaTRAK 2 Meter and How It Works section for more information .
If the system check screen does not appear, go to Troubleshooting section .
5.
Code the meter
When
and code number appear
c
Press the
to scroll forwards or
m
Press
to scroll backwards until the number matches the
test strip vial for dogs or cats . The AlphaTRAK 2 meter has a total of 99 codes .
Note: To ensure accurate test results, make sure the code number on the meter
always matches the number in the outlined box on the test strip vial for either dog
or cat .
27
Contact Customer Care at 1-888-963-8471
6.
Gently invert the control solution bottle . Wipe tip with tissue and discard the
first two drops . Apply and hold control solution to 1 sample area of the test
strip until:
You see short lines moving clockwise on the display or
You hear a beep
This indicates the test strip obtained enough control solution and the meter is
processing . If the test does not start after applying the control solution, go to
Troubleshooting section .
IMPORTANT: If a short line does not appear after 5 seconds, the
sample may be too small . Do not add control solution to both sides
of the strip . The test strips are for single use only . Do not reuse test
strips . Discard used test strips .
on the
7.
View result .
The result appears on the display when the test is complete .