Out of Range Control Solution Results

Repeat the test if control solution results are outside the range printed on
the test strip vial .
Repeat the test to confirm the results .
The correct test procedures may not have been followed .
Stop using the meter if control solution results are consistently outside the
range printed on the test strip vial . Contact Customer Care .
Causes of out-of-range results include:
Expired or bad control solution
Expired or bad test strip
Testing error
Watered-down control solution
Meter malfunction
Test strip deterioration
Control solution test done outside 59° – 104° F (15° – 40° C)
Correct code is not set in the meter
25
Contact Customer Care at 1-888-963-8471

Performing an AlphaTRAK Control Solution Test

1.
Check control solution expiry date .
IMPORTANT: Do not use control solution past expiry date . Discard
control solution 3 months after opening or on the expiry date printed
on the bottle whichever comes first . (Example: open April 15, discard
July 15; write the discard date on the side of the bottle .)
2.
Remove test strip from vial .
3.
Insert test strip to turn meter on .
Note: The meter turns off after 2 minutes of inactivity .
Remove and reinsert the unused test strip to restart
the meter .
If the meter does not turn on go to Troubleshooting section .
Top
Printed
side up
Sample
areas