Orton autofire express Instruction Manual

Ramp/hold controller

Advertisement

rA1 = 0250
°F 1 = 0750
HLd1 = 00.00
rA2 = 0900
°F 2 = 1425
HLd2 = 00.30
rA3 = FULL
°F 3 = 1050
HLd3 = 00.00
rA4 = 0150
°F 4 = 0750
HLd4 = 00.00
rA5 = 0000
4. The display shows Strt after the final entry above.
5. Press (Program) button one last time to start the firing. Display will show -
On-

Heat Treating Program Example

Below is a typical 3 step firing schedule for Heat Treating Steel.
Step 1. Heat as fast as possible from starting temperature to 1800°F, Hold
at 1800°F for 20 minutes
Step 2. Cool as fast as possible from 1800°F to 400°F, with no hold time at
400°F
Step 3. Heat as fast as possible from 400°F to 500°F, Hold at 500°F for 5
hours, and then Shut-off
To program this schedule from IdLE mode;
1. Press (Program) button. The controller will display the last active
program that was used. PrO1 – PrO9
2. Press (Program) button again if you want to use the same program that
is shown or press (Increase) button to change to another program and
then press (Program) button again.
3. Program the steps below and press (Program) button after each segment.
rA1 = FULL
°F 1 = 1800
HLd1 = 00.20
rA2 = FULL
°F 2 = 0400
HLd2 = 00.00
rA3 = FULL
°F 3 = 0500
HLd3 = 05.00
rA4 = 0000
4. The display shows Strt after the final entry above.
5. Press (Program) button one last time to start the firing. Display will show -
On-
12
®
Autofire
Express
Ramp/Hold Controller
Instruction Guide
This button is for selecting a firing program and advancing
through the programming steps. After programming is
complete, use this button to Start and Stop the firing.
This button is used to change the firing program during
programming and to change the display values for specific
program settings. During a firing, use this button for special
firing options (including Skip Step).
This button is used to change the display values for specific
program settings. It is also used to activate the Program
Review feature.
When using the Increase and Decrease buttons to
change number settings, the values will change
more rapidly if the button is held in.
1

Advertisement

loading

Summary of Contents for Orton autofire express

  • Page 1 rA1 = 0250 °F 1 = 0750 HLd1 = 00.00 rA2 = 0900 °F 2 = 1425 HLd2 = 00.30 rA3 = FULL °F 3 = 1050 HLd3 = 00.00 rA4 = 0150 °F 4 = 0750 HLd4 = 00.00 ®...
  • Page 2 1. Press (Program) button. The controller will display the last active program that was used. PrO1 – PrO9 Orton controllers require 3 variables for each heating or cooling step of a firing schedule. These variables are; 2. Press (Program) button again if you want to use the same program that...
  • Page 3 FE # - A complete firing schedule can be multiple heating and/or cooling Fatal software Errors, FE Alarms indicate a hardware failure or steps. However, for many applications a single step is all that is software problem with the controller. These alarms will disable the normal controller operation and require corrective action.
  • Page 4 the wiring for the thermocouple and the physical condition of the probe Programming Hold Time inside the firing chamber. Each step of a firing program can have an optional Hold time. Hold time is the amount of time you want to stay at the previously determined heating or FtL - This alarm indicates that the firing was taking too much time to complete cooling temperature.
  • Page 5 Status Display Codes An active delay time can be canceled by pressing the Program/Start button any time during the delay count-down to begin the actual firing. Below is a list of normal display codes which indicate the controller mode of operation.
  • Page 6 options allow you to make adjustments to the firing program without The alarm is disabled (turned off) when the alarm value is set to 32°F (0ºC). The alarm value can be reset or changed many times during a single firing. stopping the firing.