HP 81101A Reference Manual page 43

50 mhz
Hide thumbs Also See for 81101A:
Table of Contents

Advertisement

Command
:ARM:SOUR
Long
:ARM[:SEQuence[1]|STARt][:LAYer]:SOURce
Form
Set & Query
Parameter
IMMediate|INTernal[1]|INTernal2|EXTernal[1]|MANual
*RST value
IMM
Description
Use this command to select the triggering mode of the instrument by
selecting the source of the arming signal:
Triggering Source :ARM:SOURce
Internal Osc.
PLL
EXT INPUT
MAN key
Use :ARM:SENSe EDGE | LEVel to choose between Triggered and
Gated.
Command
:DISP
Long
:DISPlay[:WINDow][:STATe]
Form
Set & Query
Parameter
ON | OFF | 1 | 0
*RST value
ON
Description
Use this command to turn the front panel display on and off. Switching
off the display improves the programming speed of the instrument.
*RST switches the display back on. Use :SYSTem:PRESet to perform an
*RST without switching the display back on.
Example
To switch off the front panel display:
DISP OFF
IMMediate|INTernal[1]
INTernal2
EXTernal1
MANual
Programming Reference
SCPI Instrument Command List
Mode
Continuous
Triggered | Gated by PLL
Triggered | Gated by EXT IN
Triggered | Gated by MANKey
43

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents