HUVITZ HPE-810 Introduction Main Functions 1.1. This product is for processing lenses fit into eyeglasses frame, which automatically covers from tracing to edging. Edging Flat Safety Beveling Grooving Polishing Drilling Material Edging Beveling ○ ○ ○ ○ ○ ○...
HUVITZ HPE-810 NOTE HPE-810 (ND) type does not support drilling function. HPE-810 (ND) type is able to perform drilling by connecting an external drill machine (HDM-8000) in the above system configuration. Edger (HPE-810 (ND) type) Driller (HDM-8000) Classification 1.3.
Safety Information Introduction 2.1. Safety is everyone’s responsibility. The safe use of this machine is largely dependent upon the installers, users, operators, and managers. It is prerequisite to read and understand these specifications before installing, using, cleaning, fixing or revising. Fully understanding the whole instructions must be the first priority.
HUVITZ HPE-810 Safety Symbols 2.2. The International Electrotechnical Commission (IEC) has established a set of symbols, which are listed below. This applies only to the instrument that has the certification symbol printed explicitly on the product label or sticker. I and O on power switch represent ON and OFF respectively.
Environmental Considerations 2.3. Avoid the following environments for operation or storage: Where the machine is exposed to water vapor. Don’t operate the machine with wet hands Indoor use only Where the machine is exposed to direct sunlight. Where there are big changes in temperature. Optimal temperature range for normal operation is from 10C to 40C.
Page 12
HUVITZ HPE-810 Don’t disassemble or open the product. HUVITZ does not take responsibility for the possible problems Be careful not to block the fan located on the backside of the machine. Don’t plug the AC power cord into the outlet unless all parts of the machine are completely connected.
Such tampering will forfeit any rights to receive services during the term of guarantee. This machine must be connected with the accessories supplied by HUVITZ. If you are to use other accessories, their safety or usability must be checked and proved by their manufacturers or HUVITZ.
Page 14
HUVITZ HPE-810 Keep the User’s Manual and Service Manual in a place easily accessible at all times for persons operating and maintaining the equipment. Do not apply excessive force to cable connections. If the cable does not connect easily, make sure that the connector (plug) is appropriate for the receptacle (socket). If you caused any damage to a cable connector(s) or receptacle(s), let the damage(s) be repaired by an authorized service technician.
Page 15
stop the operation and ask for the service. If the edging wheel is broken during the operation, the broken pieces may cause a serious injury Be sure not to edge the materials other than the spectacle lenses. Otherwise, it may ...
Page 16
HUVITZ HPE-810 the maintenance. Otherwise, it may cause the malfunction of the product. When processing wheel dressing, be sure to use dressing sticks that fit each wheel. The dressing stick should be replaced when its length becomes so short that you cannot hold it easily.
System Overview Edger (HPE-810) 3.1. <Front View> Main Display (Touch Screen) Control Button Control Button Edging Room Edging Room Window (Auto) <Edging Room> (HPE-810) Drill Feeler Clamp Main Wheel Wheel for Safety Beveling & Grooving...
Page 18
NOTE 110V model rear view To connect external drill machine (HDM-8000) in the HPE-810 (ND) type, you need to use the Auto Blocker port or Frame Reader port and change the communication port setting in the preferences menu. (Refer to ‘Chapter 7.1.1 Preferences Comm. Port’)
HUVITZ HPE-810 Frame Reader (HFR-8000) 3.2. <Front View> Cover Work Space for Stylus Control Panel Display Window Gripper Gripping Bar Stylus <Rear View> Main Switch Main Power Input and Fuse Interface Terminal - EDGER 1 - EDGER 2 - Barcode Reader...
HUVITZ HPE-810 Standard Accessory (Tool Box) 3.4. Lens Adaptor Spanner(10-11) LEAP-III Tape Drill Bit (HPE-810) Lens Adaptor Remover Fuse(250V, 10A) Pattern Holder Standard Frame Standard Pattern Standard JIG for Edger Wrench Set Carriage-Fixing Bracket Feeler Tip Dressing Sticks NOTE ...
Installation Procedure Edger 4.1. Remove the shock-absorbing material from the packing box and take the Edger out of the ① box carefully. Loosen the screw at the Edger Front Cover and open the cover ② Open the edging room window with hands ③...
HUVITZ HPE-810 NOTE Keep the removed fixing materials in the tool box and use them again when you move the machine Fill up more than 2/3 of the tank with water. Make sure that the water supply hose is ...
Operation Frame Reader 5.1. Control Panel 5.1.1. MENU FRAME STOP <Material Selection - FRAME Button> You must press FRAME button and select one of the following options (frame materials) before the frame is traced. METAL PL-HARD (Hard Plastic) ...
HUVITZ HPE-810 Frame Tracing 5.1.2. ① Place the frame between the upper grippers and the lower grippers. ② Move the frame in order that the bridge of frame is located at the central position between the left and right side grippers.
Blocker 5.2. Scale Plate 5.2.1. Vertical Scale (unit:mm) Horizontal Scale (unit:mm) Near Vision guide for Bi- Focal Lens Lens Adapter Setter 5.2.2. Lens Adapter should be inserted into the Lens Adapter Setter as shown like (a). Blocking Mode 5.2.3. <Active Blocking vs Passive Blocking> There are two kinds of blockings, Active blocking and Passive blocking.
Page 28
HUVITZ HPE-810 The edger shows movement amount of the optical center for blocking on the screen. Active blocking is much easier than Passive blocking for blocking. However, active blocking has some limitations, which may result in unexpected outputs. The following table compares active blocking and passive blocking and summarizes the advantages and disadvantages of them.
Operation 5.2.4. Plug the Power Cable into the socket at the rear side of the machine. ① Put the Lens Adapter into the Lens Adapter Setter (Pay attention to the direction). ② Place the marked Lens on the Lens Holder. ③...
Page 30
HUVITZ HPE-810 <Blockings according to the lens types> Single Vision Progressive Lens Far Vision Eye Point Adjust the marking point of the lens to the Adjust the Far Vision Eye Point of the center of scale plate progressive lens to the center of scale...
HUVITZ HPE-810 Edger User Interface Main Screen 6.1. Side Button Frame Information Job Management Layout Option Function Button Edging Option Function Button Side Button 6.1.1. ① Press the R button to select right side. ② Press the L button to select left side.
Page 33
NOTE Each side of frame shape is exchanged by pressing the R button for a while. The function of exchange R/L is useful when you selected wrong side while reading demo lens or patterns. (Message pops up when press the R button for a while) (Left and Right sides are exchanged.
HUVITZ HPE-810 Job Management 6.1.2. ① ② ③ ① New : Clear current job screen and wait new job. ② Job Number : Display current job number (Press the button to change job number) ③ Job Manager : Press the job manager button to load a new job or move on to another job NOTE ...
Page 35
NOTE Following additional functions are available by pressing ‘Job Management’ button for a while. If you press ‘New’ button for a while, you can delete the current job on display immediately. You can save current job on SD card by pressing ‘Job Number’ for a while. ...
Page 36
HUVITZ HPE-810 You can start jobs with auto-blocker or remote edging host pc by pressing ‘Job Manager’ button for a while. The function of job request works only when communication option on configuration menu is set properly.
Frame Information 6.1.3. ① Frame Curve ② Circumference ③ Minimum Lens Diameter ④ Width/Height Layout Option 6.1.4. ① Binocular PD: Input prescribed binocular PD value. ② Monocular PD : Input PD values for each side (L/R) ③ Bridge Size (DBL): Input value of bridge size. ④...
Page 38
HUVITZ HPE-810 OH value is recalculated when you change OH type. The exact meaning of the mix height and the box height is as following. Heigh Height To enter OH value R/L differently, long press the OH input area.
Page 39
The box which is not changed from Default value is colored yellow. It changes into white when you input value. If frame reader traces shape of frame for both side, FPD value is calculated automatically. Therefore, input boxes of bridge size and FPD are displayed white. ...
Page 40
HUVITZ HPE-810 NOTE Numeric Keypad Click the value you want to change then this numeric keypad will pop up. ① Name of the value ② Range of the value ③ Value ④ Increase/Decrease ⑤ Backspace ⑥ Reset (Cancel) – If you press it, the value goes back to original value.
Edging Option 6.1.5. ① Edging Type Bevel Flat Groove Mini Bevel NOTE You can input additional parameters according to the selected edging type. Grooving ① Groove Depth ② Groove Width Mini Bevel ① Bevel Height...
Page 42
HUVITZ HPE-810 NOTE After you’ve selected grooving as the edging type, you can select partial grooving by pressing ‘Edging type’ button for a while. There are four ways of partial grooving. Partial Grooving Hybrid Grooving Partial Bevel Dual...
Page 43
② Bevel/Groove Position Front % Front mm Rear mm Base Curve Auto Manual ③ Size Adjustment (-2.0 ~ 2.0mm) ④ Polishing Polishing Off Polishing On High Polishing ⑤ Front Safety Beveling Safety Beveling Off Safety Beveling On ⑥ Rear Safety Beveling Safety Beveling Off Safety Beveling Small Safety Beveling Large...
Page 44
HUVITZ HPE-810 NOTE Polishing / Safety Beveling touch screen area ② Polishing On/Off ③ Front Safety Beveling On/Off ④ Rear Safety Beveling Off/Small/Large When you selected ‘Hybrid Grooving’, Safety Beveling is not available. ⑦ Lens Material Plastic Polycarbonate...
Page 45
⑨ Blocking Mode Optical Center Boxing(Frame) Center ⑩ Roughing Mode Normal Spiral Axial NOTE Normal mode is a mode where a lens contacts the wheel directly while processing. Spiral mode is a mode where a lens contacts the wheel gradually with spiral rotation. ...
Page 46
HUVITZ HPE-810 shape and check the interference. According to the Edging Options, you can control the Beveling and Grooving position. Ratio(%) from the front side of the lens Distance(mm) from the front side of the lens ...
Page 47
Min Max Max Min (-) Lens (+) Lens [Base curve mode] In the base curve mode, base curves of (+) lenses and (-) lenses exist the opposite side. In this mode, the bevel or grooving line always passes the mid of the thinnest parts, one of which is located at the midst zone of (-) lens and the other is located at the edged zone of (+) lens.
HUVITZ HPE-810 Detailed Edging Option 6.1.6. ① Feeling after roughing ② Pre-Roughing ③ Feeling Position Normal Frame Change Outside of Bevel ④ Clamp Pressure Normal Strong Weak ⑤ Wrap Removal (HPE-810)
Page 49
You can input clamp pressure value by pressing modification icon of clamp pressure for a while. (50~150) When the wrap removal mode is turned on, the edging process paused before drilling step to allow user to manually remove the wrap scraps on the lens (Only available in HPE-810).
The expert user can work more quickly by using ‘Expert Job Editor’. ‘Retouch Mode’ button is available after edging process is finished. HPE-810 (ND) type provides a drilling assignment button that can be used when the drill machine (HDM-8000) connected.
Manual Bevel/Groove Position Modification 6.2. The manual position adjustment screen will be displayed after feeling process is finished when edging position mode is manual. Stop Processing Base Position Type Min/Max Thickness Region Selection Position Side View Position Modification Reset & Apply Button Base Position 6.2.1.
HUVITZ HPE-810 Region Selection 6.2.2. ① Red Dot – Current Location ② Yellow Line – Manual Position ③ Move region clockwise (fine adjustment) ④ Move region counterclockwise (fine adjustment) ⑤ Move region clockwise ⑥ Move region counterclockwise ⑦ View Type ⑧...
Page 53
To see the shape in actual size, touch the upper right icon. Touch Area (Actual measurement) (Auto – Shape is automatically resized fitting the screen.) Angle guideline function is turned on or off by touching the center of the shape. ...
HUVITZ HPE-810 Position Modification 6.2.3. ① Front curve of the lens ② Rear curve of the lens ③ Yellow Line – Manual Position ④ Gray Line – Base Position ⑤ Bevel/Groove Preview – Distance between current position and front surface (mm) ⑥...
Page 55
NOTE Global modification function is to move the entire edging position to the lens front surface or rear. Move front Move Rear ...
Page 56
HUVITZ HPE-810 NOTE Partial modification function is to move the current edging position to the front surface or rear.. Move front Move Rear...
Page 57
NOTE Tilting function is to fix a reference point and tilt the edging position to the lens front surface or rear. Move front Move Rear...
HUVITZ HPE-810 Reset & Apply Button 6.2.4. ① Reset Button – Cancel the changes ② Apply button – Apply modified position and continue to processing NOTE If you press the base position button while changing base position, previously selected base position is initialized to selected base position.
Partial Grooving Screen 6.3. When you select Partial Grooving mode, partial grooving screen pops up after the feeling process. Region Selection Processing Type Safety Bevel Button Reset & Apply Button...
HUVITZ HPE-810 Region Selection 6.3.1. ① Orange Circle – Start Point ② Blue Circle – End Point ③ Red Line – Bevel / Flat edging Area ④ White Line – Grooving Area ⑤ Move start point clockwise ⑥ Move start point counterclockwise ⑦...
Page 61
NOTE To see the shape in actual size, press the upper right icon. Touch Area (Actual measurement) (Auto – Shape resized to fit to the screen) Angle guideline pops up when you touch the center of the shape. (Guideline spacing - ...
HUVITZ HPE-810 Processing Type Selection 6.3.2. ① Angle of the start point ② Angle of the end point ③ Grooving (Cancel Partial Grooving) ④ Flat Edging + Grooving ⑤ Beveling + Grooving ⑥ Beveling + Flat Edging ⑦ Grooving + Grooving (Dual) ⑧...
Safety Bevel Button 6.3.3. ① Front Safety Bevel OFF ② Front Safety Bevel ON ③ Rear Safety Bevel OFF ④ Rear Safety Bevel Small ⑤ Rear Safety Bevel Large. NOTE Safety bevel is not available when the processing type is Beveling + Grooving or Beveling + Flat Edging.
HUVITZ HPE-810 Job Manager 6.4. Exit Job selection tab Job list Function button Job selection tab 6.4.1. ① Waiting Job: Newly received Job data from Frame Reader or Auto Blocker. ② Working Job: Edging job or Editing job. ③ Completed Job : Finished Job ④...
Function Button 6.4.2. ① Open : Load the selected job ② Save : Save the selected job to the SD memory card ③ Search : Search Job number ④ Delete : Delete selected job NOTE Save function is only available on Waiting/Working/Completed list. ...
HUVITZ HPE-810 Working Job 6.4.4. ① Edging: Displays the Edging Processing state. ② Drilling: Displays the Drilling Processing state. ③ Time: Displays the job start time. Completed Job 6.4.5. ① Time: Displays the job finish time. NOTE You can check current processing state with the following icons.
Job Search 6.4.6. You can search for a job on the job list. ① Start ‘Job Search Mode’ by pressing the search button. ② Enter the job number you want with the virtual keyboard. ③ Check the list. Enter Job Number Search Button...
Page 68
You can search with only few letters, not the whole words. For example, you can search the Job Number ‘HUVITZ_123’ only with ‘HUVITZ’ or ‘123’. You can look at the other research results with ‘Next’ and ‘Prev’ button You can finish the Job Search mode by pressing the search button one more time.
Digital Pattern 6.5. The “Digital Pattern” modifies the lens shape in left/right, top/bottom as well as rotation and optimizes the fitting for Rimless and Semi-Rimless. To use Digital Pattern, press the Digital Designer button on the bottom right of the main screen and then select Digital Pattern tab at top left of the screen.
Page 70
HUVITZ HPE-810 <Touch Interface> To modify lens shape, make use of the 8 touch points around the lens shape. Horizontal + Vertical Horizontal Vertical Horizontal + Vertical Horizontal Vertical <Value Input> You can modify the lens shape by entering the values directly in the Input Area.
Near Vision Area Enlargement/Reduction 6.5.2. Lens shape can be changed in four directions (Bottom - Near Vision Area, Top, Left, Right). <Touch Interface> To modify lens shape, make use of 4 touch points around the lens shape. Left or Right Top or Bottom Near Vision Area (Bottom) Left...
HUVITZ HPE-810 <Value Input> You can modify the lens shape by entering the values directly in the Input Area. ① Left ② Right ③ Near Vision Area (Bottom) ④ After selecting the item to change, enter the value using the numeric keypad.
Page 73
Clockwise Counterclockwise <Value Input> You can modify the lens shape by entering value directly in the Input Area. ① Width ② Height ③ Angle (Clockwise, 0 ~ 360) Function of ① and ② are same with that Enlargement/ Reduction mode.
HUVITZ HPE-810 Partial Modification 6.5.4. Lens shape can be modified partially. <Touch Interface> Set modification area by touching any point on lens shape. The point becomes the center of the modification area. To adjust modification range, touch and drag one of...
<Modification range/ Lens shape change> You can modify the lens shape with Input button. ① Position change (counter clockwise, clockwise) ② Range change (enlargement, reduction) ③ Wide Partial Modification (concave, convex) ④ Partial Modification (concave, convex) During partial modification, where is not able to process physically is colored red.
HUVITZ HPE-810 Toggle R/L Sync 6.5.6. Function for R/L Sync. R/L Sync Off R/L Sync On NOTE “Copy” button ( ) is available when R/L sync option is turned off. Shape of the current side is copied to the other side by pressing this button.
Page 77
NOTE You can modify half-framed lens or rimless lens conveniently by using the digital pattern. The digital pattern is useful for the bi-focal lens or the progressive. Sometime too small frame shape might cut out the near vision area. In this case, the digital pattern provides the enlargement of the frame shape.
If the processed lens is too large to be to the frame, you can process the lens again with the Retouch Mode. ① Retouch Type (HPE-810) - Shape Retouch, Hole Retouch, Retouch All (Shape & Hole) ② Retouch Size(minus value ~ 0.0) - Input retouch size ③...
Hole Editor 6.7. This screen is for the edition of hole position for drilling. To use the hole editor, click the button on the main screen right below, and then select the hole editor button in the upper left. Top of hole editor Working Area Preview Function Buttons...
HUVITZ HPE-810 6.7.1. Upper Side of Hole Editor ① Hole Editor: Starts the hole editor. If the button is green, hole editor is now available. ② Digital Pattern: Starts the digital pattern. ③ Job Number: Displays the job number of currently working job. You can modify the number by touching here.
NOTE Elements required for drilling are displayed on working area. The figure below describes each element. Hole Frame Center Horz. Coordinate Notch or Slot Undrillable area Vert. Coordinate 6.7.3. Preview ① Select Right/Left: You can change the selection of the left/right. ②...
HUVITZ HPE-810 6.7.4. Function Buttons ① Select / deselect all: You can select or deselect all of the holes/slots at once. ② Retouch hole / slot: You can register the selected holes/slots for retouching. ③ Hole / Slot Easy Moving: You can move the selected holes/slots.
Page 83
The slot icon on the buttons of moving type represents the start( ) and end( position like below figure. The details of how to move the hole / slot by moving type are like below. This provides the function of parallel movement. The movement of hole / slot is like below.
Page 84
HUVITZ HPE-810 (Slot only) This provides the function to change the length of a slot. (Slot only) This provides the function to change a slot to a notch and to move the notch. The slot will be moved to the edge of a frame close to current position, and become a notch.
Page 85
List of hole / slot data: This displays a list of hole / slot data in the SD card. File name: You can set the file name of the hole / slot data to be saved. Save: You can save the hole / slot data by pressing this button.
Page 86
HUVITZ HPE-810 Select preset button: Select the preset button to apply the hole / slot data. And then choose hole / slot data you want. List of hole / slot data: This displays the list of hole / slot data in SD card.
Page 87
[Mirroring the slot] [Mirroring the group of hole / slot]...
HUVITZ HPE-810 6.7.5. List of Hole/Slot It displays the hole / slot list of the current job. You can change the position of the hole / slot by pressing the coordinates displayed on the list. Select the hole / slot: You can select the holes / slots to make a group by pressing here. The green circle is the selected one.
Page 89
NOTE Display options for X coordinate Center of frame Frame edge Display options for Y coordinate Center of frame Bottom of box...
HUVITZ HPE-810 6.7.6. Property of Hole/Slot It displays the properties of the selected hole / slot. ① Group: If selected hole / slot belongs to a group, it displays the group id. If selected hole / slot does not belongs to a group, it displays with the ‘-’.
Feeling mode & position Clamp pressure Drilling quality Size Wrap removal (HPE-810) NOTE The job editor options are applied to right/left both side concurrently. But if the options distinguish the two sides, those are applied only to the selected side.
Edging Process 6.9. Edging Start 6.9.1. Start process with following procedures after all edging option settings are completed. ① Open the edging room window by pressing ‘WINDOW’ button. ② Release the clamp by pressing ‘CLAMP’ button, and then set blocked lens. ③...
HUVITZ HPE-810 Edging Screen 6.9.2. NOTE Some buttons are unlocked after a while. Then, you can handle multi-tasking with ‘Job Manager’. According to edging state, following icons are displayed on screen. Edging Icons Processing Processed Drilling Icons ...
Edging Stop 6.9.3. Edging process stops by pressing ‘STOP’ button at any time. Edging process stops when an unexpected situation happens. CAUTION In case of emergency, stop the machine immediately by pressing the ‘STOP’ button. Motor needs some time to stop its working when you stop the machine while processing. ...
HUVITZ HPE-810 Edging Error 6.9.4. In case of an unexpected situation, error message will pop up and the process will stop immediately. NOTE Please check the error message and take a proper step. And then, restart the process.
Edging Finish 6.9.5. After edging process is finished, beep sound and finish message will be provided with activated main screen. Follow these steps to take out the processed lens. ① After edging finishes, window is open automatically. ② Press the ‘CLAMP’ button. Take out the processed lens from edging room ③...
HUVITZ HPE-810 6.10. Frame Change Frame Change Lens Registration 6.10.1. This function is for registration of the frame change lens for frame change processing. To register the frame change lens, double-click (or long click) the job number area after selecting the frame change lens on waiting tab.
Frame Change Mode 6.10.2. You can activate the frame change mode by touching the feeler icon on detail edging options at the right bottom of main screen. In this mode, feeler reads the lens 1mm inner than its actual size. So even when the size of the frame change lens and the edging size are similar, it helps to process without problems.
HUVITZ HPE-810 Frame Change Procedure 6.10.3. ① Frame reader reads the lens which will be used in frame change. (when you hold the lens at pattern holder, you must block the lens adaptor at optical center of lens) ② The data will be transmitted to the edger after the reading process is completed.
Configuration Configuration of Frame Reader 7.1. <How to start the configuration> ① Press these buttons sequentially : ‘MENU’ >> ‘BOTH’ ② Press ‘FRAME’ button, then you can navigate all setting options. ③ Press the ‘BOTH’ button to execute a specific option. ④...
HUVITZ HPE-810 Configuration of Edger 7.2. Press the MENU button at the bottom left of the Main screen. <MENU Screen> ① Preferences ② Adjustments ③ Software Version ④ Exit ⑤ Shutdown NOTE Refer to ‘Chapter 8. System Maintenance’ for the other functions.
Preferences 7.2.1. User Interface can be customized to the user’s preference. <General> Default settings for user interface Language – Sets default language Beep Sound – Turns on or off the beep sound Frame Fitting Mode – Applies frame fitting values R/L separately or R/L equally Sleep Mode –...
Page 104
HUVITZ HPE-810 NOTE Setting Screen Tab Menu ① Radio Button – Choose only one of a predefined sets of options ② List Box – Choose a list from existing options ③ Input Box – Input value to be used by the program ④...
Page 105
<Layout> Default settings for main screen layout option Default Side – Set default editing side Lens Type – Set default layout mode of main screen Blocking Mode – Set default blocking method OH Type – Set default OH display method Default Value –...
Page 106
HUVITZ HPE-810 <Edging> Default settings for main screen edging option Edging Start Side –Select left or right to start edging. Warning message will pop up if the processing is not started from the selected side. Default Lens Material – Set default lens material ...
Page 107
<Digital Pattern> Default settings for digital pattern screen Digital Pattern Apply Method Fix FPD – Calculates bridge size value based on FPD Fix Bridge Size – Calculates FPD value based on bridge size Zoom Ratio – Set default zoom ratio of touch input area ...
Page 108
HUVITZ HPE-810 <Hole Editor> Default settings for hole editor screen Hole Depth – Set default hole/slot depth value Hole Diameter – Set default hole/slot diameter value Angle Default Options – Set default tilting angle mode and manual tilting value ...
Page 109
<Frame Fitting> Frame Fitting values are used to adjust the size variation caused by frame material difference. Master Size – Input size adjustment value for all frame material Metal – Input size adjustment value for metal frame Hard Plastic – Input size adjustment value for hard plastic frame ...
Page 110
Blocker Port – Use blocker port Protocol – Set communication protocol Baud Rate – Set communication speed External Driller Connection – Set external driller communication port.(HPE-810 (ND) type) Job Request Port – Set communication port for job downloading ...
Adjustment settings are closely related to the edging quality, so keep the system default values as much as possible. If you want to change adjustment values, ask to the service technician of HUVITZ or the technician authorized by HUVITZ.
Page 112
HUVITZ HPE-810 OFF – Use as it is ON – Substitute Pump 2 for Pump 1. Change GL Roughing Rotation – Each time the glass lens rotates, direction of rotation is changed. Frame Change Mode : Feeler Offset – Set the feeling offset for the frame change mode ...
Page 113
<Clamp> You can set default clamping pressure value. Normal – value for Normal Pressure Mode High – value for High Pressure Mode Low – value for Low Pressure Mode NOTE You can input the value from 50% to 150%. 100% means 55kgf±3...
Page 114
HUVITZ HPE-810 <Size> You can adjust the edging size. Finishing Size – Size adjustment value for finishing step. Polishing Size – Size adjustment value for polishing step NOTE How to adjust finishing size ① Load internal circle pattern(Ø 50).
Page 115
circle (Ø50) 52mm Edged lens Remember that the adjustment value is based on radius. (Do not confuse with the Frame Fitting value. Frame Fitting value is based on diameter) Adjust the value within the tolerance ±0.1mm.
Page 116
HUVITZ HPE-810 <Margin> You can set margin for the following edging steps. Roughing Margin – Margin value for roughing Polishing Margin – Margin value for polishing CAUTION Margin values have influence on the lens material or edging quality. So it is recommended not to...
Page 117
<Axis> You can adjust the distortion of axis. Main – Compensation value for main axis Polishing – Compensation value for polishing axis Grooving – Compensation value for grooving axis Safety Beveling – Compensation value for safety beveling...
Page 118
HUVITZ HPE-810 NOTE How to adjust axis ① Load internal square pattern (40mm x 40mm). ② Draw a horizontal line on the lens and locate the lens adaptor at the center and parallel to the line. ③ Execute flat edging, check the main axis, and input the adjustment value.
Page 119
How to adjust grooving axis value Add (-) value Add (+) value How to adjust safety beveling axis value Front Front Rear Add (-) value Add (+) value Add (+) value In main axis adjustment, use the digital pattern function to rotate square pattern 45 degrees before edging.
Page 120
HUVITZ HPE-810 <Beve> You can adjust bevel / groove position Bevel – Set adjustment value for bevel position. Groove – Set adjustment value for groove position and depth. Mini Bevel – Set adjustment value for mini bevel polishing position.
Page 121
How to adjust polished bevel position value Add (-) value NOTE How to adjust groove ① Load internal circle pattern (Ø 50). ② Set groove position 50% and execute grooving. ③ Check the groove position and depth. ④ The tolerance must be within ±0.1mm. ...
Page 122
HUVITZ HPE-810 <Safety Bevel> You can adjust safety bevel position Front – Adjust front safety bevel position Rear – Adjust rear safety bevel position NOTE How to adjust safety bevel ① Load internal circle pattern (Ø 50).
Page 123
How to adjust rear safety bevel value Add (-) value How to adjust front safety bevel value Add (+) value...
Page 124
HUVITZ HPE-810 <Wheel> You can set the options related to edging wheels. Polishing Wheel Size – Set size compensation value for polishing wheel Max Rimless Thickness – Set maximum width of lens for flat edging NOTE If the polishing wheel is worn out, you can improve polishing quality by adjusting the Polish Wheel Size value but if you adjust the value inappropriately, it may cause size problem or poor polishing quality.
Page 125
If there is a problem with the edging room window movement, window motor can be adjusted with these values. But it is recommended not to change the default values as much as possible. If you need to change, ask advice of the service technician of HUVITZ or the technician authorized by HUVITZ.
HUVITZ HPE-810 System Maintenance Automatic Calibration of Frame Reader 8.1. Automatic Calibration of Frame Reader is required with the following cases, When the Size or Axis value of the traced data seems to have a problem, When the Frame Reader seems to have a problem in operating normally.
Pattern Calibration 8.1.3. Procedures for pattern calibration ① Press MENU button and select CALIB. ② Press FRAME button and select PATTERN. ③ Press BOTH button. ④ Load the Standard Pattern. ⑤ Press BOTH button to start calibration. Next, follow the instructions on displayed. NOTE The result of Automatic Calibration will be automatically saved.
You can execute size calibration and feeler calibration with standard edging Jig. CAUTION Calibration function has influence on the edging quality so ask advice of the service technician of HUVITZ or the technician authorized by HUVITZ if necessary. Size Calibration 8.2.1.
Test Mode 8.3. Test Mode is efficient to find out whether the machine works properly or not. Sensor 8.3.1. It displays all sensor values of the edger. ① All sensor information ② Refresh sensor status ③ Initialize edger ④ Exit Test Mode...
Page 130
Drill Y – Drill Y axis PI sensor value Drill Tilting – Drill tilting PI sensor value Drill Current – Drill motor current value Drill sensor values (Drill Y, Drill Tilting, Drill Current) are displayed only in HPE-810 model.
HUVITZ HPE-810 Pump 8.3.5. You can check pump and vacuum cleaner. ① Stop all pumps ② Pump 1 test ③ Pump 2 test ④ Vacuum test ⑤ Pump and vacuum state...
HUVITZ HPE-810 Drill 8.3.7. You can check drilling module. ① Drill Y test ② Drill Tilting test ③ Drill Motor test ④ Drill sensor state ⑤ Drill bit replacement NOTE Drill test mode is only available in HPE-810 model.
8.3.8. You can check all the PI sensors and motors. If the test fails, ask advice of the service technician of HUVITZ or the technician authorized by HUVITZ. ① X PI – X axis PI sensor test ② Y PI – Y axis PI sensor test ③...
HUVITZ HPE-810 Wheel Dressing 8.4. Wheel dressing is useful to maintain the wheel performance because it removes the lens debris in the wheel. Main Wheel – Glass roughing , finishing and polishing wheel manual dressing mode Safety Bevel – Safety beveling wheel manual dressing mode ...
It shows the number of lenses that have been processed since the reset point. NOTE The edging is counted only after the edging completes successfully. You can reset current statistics values whenever you need. Drilling values are displayed only in HPE-810 model ...
The wheel statistics values are for reference to determine replacement time, so it is not relevant to the actual lifetime of the wheel. Please refer to Wheel statistics values to execute wheel dressing or replace the wheel. Drilling value is displayed only in HPE-810 model.
It shows the total number of lenses that have been processed since the very first time of the installment. NOTE The edging is counted only after the edging completes successfully. Total statistics values cannot be initialized. Drilling values are displayed only in HPE-810 model.
HUVITZ HPE-810 Date & Time 8.6. You can change the system time. Please set the date and time first after you install this machine. Touch Test 8.7. You can check the normal operation of the touch screen NOTE You must use the emergency touch screen test function if the touch test is not possible.
Configuration Management 8.8. You can backup or restore system configuration data. ① Backup – Backup configuration data to SD card. ② Restore – Restore configuration data from SD card ③ Factory Reset – Reset to factory default setting NOTE Configuration Management function is available when SD card is inserted.
HUVITZ HPE-810 SW Upgrade 8.9. You can upgrade software. HPE-810 consists of two kinds of software. GUI software handles user input and MOTOR software controls mechanical parts of the system. ① GUI – Upgrade GUI software ② MOTOR – Upgrade MOTOR software ③...
③ Press the MOTOR button to start upgrade. ④ MOTOR upgrading process takes about 2 minutes. ⑤ After upgrading process finishes, motor program restarts automatically. 8.10. Maintenance Code Maintenance code is used for the engineer to diagnose the system.
HUVITZ HPE-810 8.11. Maintenance Tips Wheel Replacement 8.11.1. It is recommended to keep the replacement cycle of Wheel in order to maintain the Edging Performance. Replacement Cycle : 7,000 ~ 8,000 pieces of Lens edged NOTE Be sure to replace the Finishing Wheel and Polishing Wheel together.
Page 147
Replacement position Jaw cap ⑤ Loosen the jaw cap with a spanner (clockwise). 10mm spanner (in the Toolbox) ⑥ Remove the bit. Installing a new drill bit ⑦ Put a new bit and push it to the very end. ⑧...
8.11.3. It is recommended to keep the regular maintenance cycle in order to maintain the Edging Performance. Be sure to carry out the regular maintenance by the service technician of HUVITZ or the technician authorized by HUVITZ. Regular Maintenance Cycle : 1,500 ~ 2,000 pieces of Lens edged ...
Cleaning 8.11.5. Sweep the outer surface and LCD panel with soft fabric material. ① When you do not use this device, wrap it with dust cover. ② CAUTION Before cleaning your product, disconnect the power cord. Since the exterior of the product is easily scratched, be sure to use soft cleaning cloth. When cleaning the product, avoid spraying cleansing agent directly onto the product.
HUVITZ HPE-810 Troubleshooting Guide If problems occur, Please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list, please contact HUVITZ Problem Cause Solution LCD touch is Confirm the squashing following edge of Touch screen is squashed by LCD cover.
If you can’t contact with your local distributor, you can directly get in touch with the service department of the HUVITZ using the phone number and the address written in the below table. How to Contact HUVITZ Co., Ltd HUVITZ Co., Ltd.(Headquarter)
Need help?
Do you have a question about the HPE-810 and is the answer not in the manual?